{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Newton",
      "state": "KS",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90995",
      "recalling_firm": "Full Vision Inc",
      "address_1": "3017 Full Vision Dr",
      "address_2": "N/A",
      "postal_code": "67114-9750",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Wi, GA, NY, Mexico",
      "recall_number": "Z-0368-2023",
      "product_description": "FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE",
      "product_quantity": "69 units",
      "reason_for_recall": "This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.",
      "recall_initiation_date": "20220831",
      "center_classification_date": "20221202",
      "report_date": "20221214",
      "code_info": "UDI/DI 00860176000668, Serial Numbers:  GEDC-6632, GEDC-6633, GEDC-6634, GEDC-6635, GEDC-6636, GEDC-6637, GEDC-6638, GEDC-6639, GEDC-6640, GEDC-6641, GEDC-6642, GEDC-6643, GEDC-6644, GEDC-6645, GEDC-6646, GEDC-6647, GEDC-6648, GEDC-6649, GEDC-6650, GEDC-6651, GEDC-6652, GEDC-6653, GEDC-6654, GEDC-6655, GEDC-6656, GEDC-6657, GEDC-6658, GEDC-6659, GEDC-6660, GEDC-6661, GEDC-6662, GEDC-6663, GEDC-6664, GEDC-6665, GEDC-6666, GEDC-6667, GEDC-6668, GEDC-6669, GEDC-6670, GEDC-6671, GEDC-6672, GEDC-6673, GEDC-6674, GEDC-6675, GEDC-6676, GEDC-6677, GEDC-6678, GEDC-6679, GEDC-6680, GEDC-6681, GEDC-6755, GEDC-6756, GEDC-6757, GEDC-6758, GEDC-6759, GEDC-6760, GEDC-6761, GEDC-6762, GEDC-6763, GEDC-6764, GEDC-6839, GEDC-6840, GEDC-6841, GEDC-6842, GEDC-6843, GEDC-6844, GEDC-6845, GEDC-6846, GEDC-6847"
    }
  ]
}