  "results": [
    {
      "country": "United States",
      "city": "Plymouth",
      "address_1": "2300 Berkshire Ln N",
      "reason_for_recall": "On December 17, 2015, Greatbatch discovered that the Standard Offset Cup Impactor with a POM-C handle grip failed sterility testing when housed in a dedicated instrument case using the sterilization parameters identified in the Standard Offset Cup Impactor\u0019s Instructions for Use (IFU).",
      "address_2": "",
      "product_quantity": "4478 (2906 US, 1572 OUS)",
      "code_info": "Lot Number  30491  34897  2495656  2500095  2530283  2543053  2561976  2564518  2565790  2661673  2742720  2742725  2742726  2750445  2750447  2789308  2794360  2799474  2799475  2816992  2828788  2832952  2842658  2853782  2927152  2970762  2970798  3016939  3018991  3018992  3018993  3029352  3058553  3087813  3138763  3148504  3166474  3172918  3187998  3188004  3252224  3270048  3275306  5780001  2233370001  2233670001  2242190001  2259200001  2259210001  2271650001  2299480001  2299490001  2315880001  2315890001  2315900001  2315910001  2315920001  2322180001  2362520001  2363250001  2371900001  2371910001  2371920001  2371950001  2378420001  2459710001  2471740001  2487360001  2490460001  2491700001  2491710001  2504910001  2504930001  2504950001  2516110001  2518340001  2518350001  2518370001  2528690001  2553410001  2553440001  2553450001  2559210001  2566230001  2595380001  2627950001  2665630001  2724230001  2725870001  2725880001  2745160001  2766420001  2781250001  2782010001  3006450001  3007860001  3029980001  3037400001  3037470001  3038530001  3038560001  3038700001  3038720001  3038730001  3038740001  3038750001  3038760001  3038790001  3038800001  3038810001  3038820001  3038830001  3038840001  3038850001  3038860001  3039320001  3039330001  3046100001  3046200001  3060700001  3061130001  3061390001  3061930001  3062700001  3065930001  3066030001  3066060001  3068430001  3083790001  3086350001  3091750001  3092720001  3092730001  3093900001  4006680001  4007350001  4007560001  6342540001  6411410001  6411450001  7004290001  7008770001  7010540001  7011200001  7011240001  7014640001  7019390001  7022800001  7022810001  7023900001  7023900002  7032220001  7032230001  7033140001  7033830001  7036020001  7043510001  7046530001  7047580001  7054290001  7055180001  7064180001  7064340001  7067270001  7067280001  7068200001  7068300001  7068310001  7070520001  7072060001  7072070001  7072370001  7072370001  7072380001  7072610001  7075380001  7084500001  7085220001  7085620001  7085660001  7094580001  7094640001  7095080001  7095400001  7095410001  7097350001  7099830001  7100040001  7102740001  7102860001  7103040001  7103110001  7103350001  7103370001  7103380001  7103390001  7103560001  7104750001  7105030001  7105130001  7105350001  7105680001  7105720001  7106840001  7107610001  7108540001  7110290001  7112630001  7112770001  7113670001  7115160001  7116610001  7116620001  7116970001  7116980001  7117140001  7117150001  7117350001  7117490001  7117750001  7117760001  7117800001  7118530001  7118960001  7119760001  7119790001  7119800001  7124570001  7124580001  7124840001  7124940001  7126400001  7127770001  7128600001  7132050001  7132570001  7136610001  7139090001  7143650001  7143700001  7143710001  7145140001  7147570001  7148710001  7151430001  7152660001  7152930001  7153580001  7153610001  7153840001  7153850001  7153860001  7153870001  7155930001  7157020001  7157820001  7158050001  7161200001  7162790001  7165980001  7168000001  7168010001  7168270001  7170610001  7172520001  7175480001  7176040001  7176290001  7176330001  7177340001  7177350001  7177360001  7177370001  7179510001  7180460001  7180910001  7187030001  7187110001  7187650001  7187880001  7187890001  7188410001  7188420001  7188510001  7188520001  7190750001  7204710001  7206200001  7206930001  7212370001  7212490001  7215880001  7221210001  7228670001  7233210001  7233470001  7233490001  7234170001  7234260001  7234750001  7235840001  7236740001  7237910001  7241550001  7243420001  7244460001  7244900001  7252070001  7252330001  7253080001  7253090001  7253200001  7255070001  7255430001  7259520001  7261980001  7263100001  7264550001  7265500001  7266960001  7268590001  7269650001  7269660001  7273870001  7276290001  7276350001  7277140001  7284380001  7284420001  7284540001  7285000001  7285150001  7285170001  7285260001  7285660001  7285780001  7285820001  7288860001  7290770001  7290780001  7290900001  7307910001  7307920001  7307930001  7307940001  7307950001  7307960001  7307970001  7307980001  7307990001  8181080001    1003292  149467  2225930001  2233670001-62  2259190001  255441  2740689  2822781  2824203  2827399  2827401  2840353  2846027  2846037  3037410001 & 3037420001  3037430001 & 3037440001  3037450001 & 3037460001  385228  446559  468474  493607  502998  520684  565663  573358  600359  682102  701817  701838  711367  7120480001  7124590001  7124600001  7124610001  7132510001  7144330001  7144340001  7144350001  7148720001  7149430001  7153590001 & 7153600001  7177130001  7188440001  7212390001  7234560001  7234570001  7242420001  7242430001  7245010001  7245020001  7245030001  7255380001  7255390001  7255400001  7255410001  7263030001  7263040001  7267070001  7267950001  7267960001  7267970001  7267980001  7267990001  743422  773568  799699  841060  847862  877907  892823  968916  991034  CIDO00  DIAP00  MCT705-05A178  MCT705-07D851  MCT705-07D851  MCT705-09B220  MCT705-10F633  MS1063  No Lot Code  OF-002-29515-0001  OF-002-32530-0001  OF-002-41119-0001  OF-002-42944-0001  OF-002-44825-0001  OF-002-44826-0001  OF-002-45448-0001  OF-002-45665-0001  OF-002-46173-0001  OF-002-49050-0001  OF-002-50500-0001  OF-002-50501-0001  OF-002-53569-0001  OF-002-58260-0001  OF-002-61715-0001  OF-002-72372-0001  OF-002-72373-0001  OF-002-72420-0001  OF-002-75542-0001  OF-003-01045-0001",
      "center_classification_date": "20161129",
      "distribution_pattern": "Worldwide Distribution - US including CA, IN, KS,NJ, NC, TN, TX, and UT and Internationally to ITALY, JAPAN, FRANCE, SPAIN, GERMANY, NETHERLANDS, GREECE, LUXEMBOURG, SOUTH KOREA, IRELAND, UNITED KINGDOM, AUSTRALIA, AUSTRIA, BELGIUM, SWITZERLAND, NORWAY, SWEDEN, and TAIWAN.",
      "state": "MN",
      "product_description": "Greatbatch, Offset Cup Impactor with the following Parts and Catalogue #:   Tl0600 ~ MMP3100CHA04  T10491 MIMP3100CHA04  T12230 M2M1P3100CHA04  T15752 ~MMP3100CHA04  T15949 ~MMP3100CHA04  T17238   MMP3100CHA04  T9954     MMP3100CHA04  T9955     MMP3100CHA04  T8468     MMP3100CHA04  T11506   MMP3100CHA13  T10281   MMP3100CHA04  T8044   ~MMP3100CHA04  T17203   87-4 20-00980  T8177 ~ MMP3100CHA04  T16934  MMP3100CHA62  Tll340 SHP9109.00010  T8043 ~ MMP3100CHA04  T17321 MMPP3100CHA04  T15311 GG920010029  T1 7650 255000110  T8143 803-t 1-103  Tl4384 EK-00200690  T1666t XEB227242  T9316 MIMP3100CHA04  Tl3642 500325  T10753 ~MMP3100CH04  T9348 ~ MMP3100CHA04  T13480  MMP3100CHA042  T10243 ~MIMP3001CHA08  Tll209 ~  MMP300CHA12  Tl0861    MMP3100CHA04  T16427   MMP3100CHA04  T9894 :-. MMP3100CHA04  T12861  MMP3100CHA\\041  T8088 2002CUPI  T1661 1 T16611  Standard Offset Cup Impactors are handheld, reusable surgical instruments used to position and seat acetabular cups into the reamed acetabulum during hip joint replacement surgery. They are equipped with a suitable connection system to the acetabular cup (e.g. a thread) and can be supplied with a fixed or removable shock absorber. A hammer is used to deliver the necessary impact to an anvil (or impactor disc) at the rear end of the instrument.",
      "report_date": "20161207",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Greatbatch Medical",
      "recall_number": "Z-0367-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75043",
      "termination_date": "20170727",
      "more_code_info": "",
      "recall_initiation_date": "20160802",
      "postal_code": "55441-4575",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}