  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Santa Clarita",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83984",
      "recalling_firm": "TriMed Inc.",
      "address_1": "27533 Avenue Hopkins",
      "address_2": "N/A",
      "postal_code": "91355-3910",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of MD, UT.",
      "recall_number": "Z-0366-2020",
      "product_description": "TriMed Olecranon Hook Plate, Lt, 6-Holes, REF: OHOOKL-6, Non-Sterile, RX only,  (01) 00842188101281 - Product Usage: The OHOOKL is a semi-tubular design bone plate with two offset intramedullary tines, which are intended to penetrate the small terminal fragment to provide additional stabilization.",
      "product_quantity": "583 Bone Plates",
      "reason_for_recall": "The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and the modification may not have been fully qualified.",
      "recall_initiation_date": "20180815",
      "center_classification_date": "20191113",
      "report_date": "20191120",
      "code_info": "Lot # 110631,"
    }
  ]
}