  "results": [
    {
      "status": "Terminated",
      "city": "Covington",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75452",
      "recalling_firm": "C.R. Bard, Inc.",
      "address_1": "8195 Industrial Blvd NE",
      "address_2": "",
      "postal_code": "30014-1497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US distribution to: CA, KS, MO, NY, VA, WA, and to countries of: China, Italy, Sweden and United Kingdom.",
      "recall_number": "Z-0363-2017",
      "product_description": "X-Force Nephrostomy Balloon Dilation Catheter Kit with Inflation Device, PTFE Sheath, 8 mm x 15 cm, Product Code 996081    The product is a dual lumen catheter with a 24 (8mm) or 30Fr (10mm) balloon mounted on the distal tip. It has a radiopaque tip and radiopaque marker beneath the balloon. It is recommended for use in the dilation of the nephrostomy tact and for placement of the working sheath.",
      "product_quantity": "192 units",
      "reason_for_recall": "Contract manufacturer packaged an incorrect sheath with the product.  Packaged 30Fr instead of 24Fr.",
      "recall_initiation_date": "20161012",
      "center_classification_date": "20161104",
      "termination_date": "20210922",
      "report_date": "20161116",
      "code_info": "Lot BMZCE039, Expiration Date 2018-12-31",