  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Kabelsketal",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83707",
      "recalling_firm": "Tec Com Gmbh",
      "address_1": "Eingang B",
      "address_2": "Am Krummling 1",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the state of UT.",
      "recall_number": "Z-0362-2020",
      "product_description": "TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System TECOTHERM NEO is designed for controlled cold &#38; heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective.",
      "product_quantity": "5 units",
      "reason_for_recall": "The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element  would report a system failure and no longer be able to guarantee temperature control at full power.",
      "recall_initiation_date": "20180814",
      "center_classification_date": "20191112",
      "termination_date": "20191220",
      "report_date": "20191120",
      "code_info": "Serial Numbers: 2018/09/07, 2018/10/01, 2018/10/02, 2018/10/03, 2018/10/05"
    }
  ]
}