  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Foothill Ranch",
      "address_1": "27051 Towne Centre Dr",
      "reason_for_recall": "Ossur inititated a voluntary recall of the A-642100 Fixed Offset Adapter because it can develop fatigue cracks in normal use over an extended period of time and ultimately separate completely.",
      "address_2": "",
      "product_quantity": "66 units",
      "code_info": "Item Number:  A-642100       All Serial Numbers.",
      "center_classification_date": "20131119",
      "distribution_pattern": "US Distribution including the states of AL, TX, NY, CA, VA, GA, WI, NC, ID and NJ.",
      "state": "CA",
      "product_description": "Fixed Offset Adapter, A-642100.    The Offset Adapter is a fixed modular component which sets the tube 14.5 mm (0.57 inch) offset from the center line of the pyramid. It is made of a high-strength alloy with a pyramid receiver and is intended where there is a need to offset the tube horizontally in relation to other components. It is fully compatible with all standard 30mm prosthetic pylons. (3) Offset: 14,5mm (0.57inch).",
      "report_date": "20131127",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ossur North America Inc",
      "recall_number": "Z-0362-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66609",
      "termination_date": "20140617",
      "more_code_info": "",
      "recall_initiation_date": "20131024",
      "postal_code": "92610-2804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}