  "results": [
    {
      "status": "Terminated",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84135",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of  Netherlands.",
      "recall_number": "Z-0360-2020",
      "product_description": "Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.",
      "product_quantity": "9 systems",
      "reason_for_recall": "Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.",
      "recall_initiation_date": "20190208",
      "center_classification_date": "20191112",
      "termination_date": "20220607",
      "report_date": "20191120",
      "code_info": "Model CAT-870B  D053991    Serial No.       Manufacture Date  99D08X2101  October 2008  99D0582039  August 2005  99D0762072  June 2007  99D0852090  May 2008  99D07Y2078  November 2007  99D0812081  January 2008  99D0582042  August 2005  99D0552033  May 2005  D551206        May 2005"
    }
  ]
}