  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72649",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution -- To the states of WI, PA, OH, ME, IL, and NY.",
      "recall_number": "Z-0360-2016",
      "product_description": "Perseus A500 Anesthesia Workstation, MK06000.    Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing.",
      "product_quantity": "34",
      "reason_for_recall": "There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one of the following to occur: 1. The workstation has a fresh gas delivery failure and a ventilator failure with respective alarms. 2. The workstation shuts down completely with an audible alarm.",
      "recall_initiation_date": "20151112",
      "center_classification_date": "20151211",
      "termination_date": "20170405",
      "report_date": "20151223",
      "code_info": "ASHB-0140  ASHF-0039  ASHF-0055  ASHF-0056  ASHF-0057  ASHH-0008  ASHH-0009  ASHH-0010  ASHH-0011  ASHH-0012  ASHH-0013  ASHH-0014  ASHJ-0095  ASHJ-0096  ASHJ-0097  ASHJ-0098  ASHJ-0099  ASHK-0001  ASHK-0002  ASHK-0003  ASHK-0008  ASHK-0009  ASHK-0010  ASHK-0056  ASHK-0057  ASHK-0079  ASHK-0080  ASHK-0081  ASHK-0082  ASHK-oo83  ASHK-0084  ASHK-0085  ASHK-0086, and   ASHK-0087."
    }
  ]
}