  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Utica",
      "address_1": "525 French Rd",
      "reason_for_recall": "Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature.  This causes some devices to generate a \"Check Seal\" message with an audible alarm and to stop the heating cycle.",
      "address_2": "",
      "product_quantity": "9 devices",
      "code_info": "Serial Numbers: 1) 15090103 & 15150005, 15150006, 15150007, 15150008, 15150009, 15150010, 15150011, 15150012, 15130004, 15250013, 15250014, 15250015, 15250016, 15250017, 15300018 & 15300020, 2) 15050001, 15300005 & 15300006",
      "center_classification_date": "20151201",
      "distribution_pattern": "Distributed in the states of CA, and TX and in the country of   Mexico.",
      "state": "NY",
      "product_description": "Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.",
      "report_date": "20151209",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "ConMed Corporation",
      "recall_number": "Z-0359-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72586",
      "termination_date": "20180524",
      "more_code_info": "",
      "recall_initiation_date": "20151105",
      "postal_code": "13502-5945",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}