      "country": "United States",
      "city": "Parsippany",
      "address_1": "100 Interpace Pkwy",
      "reason_for_recall": "Polaris Translation Screw Failures were recently reported to Biomet Spine.  These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.",
      "address_2": "",
      "product_quantity": "7,092 units",
      "code_info": "Multiple Item and Lot numbers.  The firm expects all Translation Screw Systems (containing a cobalt chrome component) used and contacted with the   SpF Implantable Spine Fusion Stimulator (DC electrical current) to be at risk for this failure mode.",
      "center_classification_date": "20131119",
      "distribution_pattern": "Nationwide distribution including Puerto Rico.",
      "state": "NJ",
      "product_description": "BIOMET Spine,  5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only.  Non-Sterile.    Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system",
      "report_date": "20131127",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ebi, Llc",
      "recall_number": "Z-0359-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66300",
      "termination_date": "20160831",
      "more_code_info": "",
      "recall_initiation_date": "20130822",
      "postal_code": "07054-1149",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}