{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88987",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide.",
      "recall_number": "Z-0353-2022",
      "product_description": "Puritan Bennett 980 Series Ventilator, 980X3JADIJJ",
      "product_quantity": "278 total",
      "reason_for_recall": "Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.",
      "recall_initiation_date": "20211104",
      "center_classification_date": "20211209",
      "termination_date": "20240709",
      "report_date": "20211215",
      "code_info": "GTIN/UDI - 10884521172531  Serial Numbers: 35B1700393 35B1700419 35B1700420 35B1700427 35B1700428 35B1700429 35B1700540 35B1700564 35B1700568 35B1700573 35B1800002",
      "more_code_info": ""
    }
  ]
}