  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64832",
      "recalling_firm": "Hospira Inc.",
      "address_1": "275 N Field Dr",
      "address_2": "",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.   Internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.",
      "recall_number": "Z-0349-2014",
      "product_description": "The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.",
      "product_quantity": "1) List Number 13000: 2,162 pumps; 2) List Number 13100: 15 pumps; 3) List Number 13150: 1,955 pumps",
      "reason_for_recall": "The proximal and distal pressure sensor calibration can drift resulting in the pump failing the Proximal or Distal Occlusion Operational Test, as described in the GemStar Technical Service Manual, or reporting one of the following errors during device setup or infusion:     1) Cassette Check - D;   2) Cassette Check - P;   3) Proximal Occlusion;   4) Distal Occlusion;   5) Pressure Calibration Error;   6) Bad Pressure Sensor Event;   7) Bad Pressure State;   8) Distal Pressure is Out of Range;   9) Proximal Sensor is Out of Range.     A pump with this issue may, instead of reporting an error, not detected occlusions or issue false occlusion alarms, which will stop the infusion and invoke visual and audible warnings to the user.",
      "recall_initiation_date": "20130315",
      "center_classification_date": "20131121",
      "termination_date": "20170307",
      "report_date": "20131127",
      "code_info": "*** 1) List Number 13000, Serial Numbers:   98121336 ,   98121173 ,   98120835 ,   98120730 ,   98120768 ,   98120782 ,   98120788 ,   98120822 ,   98120824 ,   98120839 ,   98120856 ,   98120880 ,   98120885 ,   98120895 ,   98120915 ,   98120749 ,   98120769 ,   98120774 ,   98120784 ,   98120859 ,   98120884 ,   98121019 ,   98121066 ,   98121079 ,   98121083 ,   98121135 ,   98121153 ,   98121163 ,   98121197 ,   98120522 ,   98120653 ,   13431222 ,   98120635 ,   98120405 ,   98120508 ,   98120634 ,   98121147 ,   98120543 ,   12470380 ,   98120512 ,   98120815 ,   98120681 ,   97971490 ,   97970893 ,   97970240 ,   98070264 ,   98070295 ,   99170068 ,   99170032 ,   98070345 ,   98070476 ,   98070905 ,   98070995 ,   97972107 ,   97972609 ,   97971538 ,   98021653 ,   98021821 ,   99400401 ,   99400435 ,   97970456 ,   13431303 ,   98021816 ,   97972825 ,   98021121 ,   97971767 ,   99107521 ,   98020395 ,   14180727 ,   98120459 ,   98120495 ,   98120777 ,   98020964 ,   98070286 ,   98121157 ,   97972571 ,   97970541 ,   97971254 ,   98120535 ,   98120640 ,   98120840 ,   97970194 ,   98021636 ,   98020575 ,   97972639 ,   99222573 ,   98021550 ,   98021603 ,   98021682 ,   98370030 ,   98020523 ,   98020446 ,   98020441 ,   99107526 ,   97971754 ,   98020929 ,   98020932 ,   98022223 ,   13434785 ,   98020072 ,   98020317 ,   98021476 ,   98021478 ,   98021732 ,   98021891 ,   98021912 ,   98021252 ,   99170080 ,   98020623 ,   98020018 ,   98020293 ,   98020574 ,   98021031 ,   98021423 ,   98021449 ,   98020013 ,   97972018 ,   98021177 ,   98020264 ,   99221974 ,   98020817 ,   98021687 ,   98070125 ,   98021304 ,   98021546 ,   98022039 ,   98020281 ,   97972802 ,   97972086 ,   97971122 ,   12470492 ,   98121130 ,   97972659 ,   97972489 ,   98120869 ,   13433080 ,   97972047 ,   97972824 ,   98020549 ,   98021584 ,   97970313 ,   97970073 ,   98021364 ,   98020312 ,   98021961 ,   98020127 ,   98020606 ,   98021194 ,   99222658 ,   98020490 ,   98021141 ,   98021276 ,   98021171 ,   98020460 ,   98021969 ,   99107535 ,   99221815 ,   98021977 ,   98021120 ,   98021412 ,   99222620 ,   14180122 ,   99400456 ,   98071220 ,   98070607 ,   98070371 ,   98070781 ,   98070815 ,   98070696 ,   98070812 ,   97972981 ,   97970766 ,   98020074 ,   98020902 ,   14180174 ,   99219028 ,   98070190 ,   13433667 ,   97970613 ,   98021101 ,   98021099 ,   97970666 ,   97970308 ,   97970071 ,   98021367 ,   98021508 ,   98021308 ,   98021309 ,   98021208 ,   97972842 ,   97972841 ,   99222510 ,   98020727 ,   98020235 ,   98021264 ,   98021708 ,   98021794 ,   97971750 ,   99170302 ,   98070212 ,   98070006 ,   97972808 ,   98021554 ,   98021900 ,   99400482 ,   98020054 ,   98021195 ,   98021496 ,   98120534 ,   98020866 ,   98021179 ,   98070850 ,   98070448 ,   98070794 ,   97972040 ,   97971807 ,   98350110 ,   97971723 ,   98350105 ,   98350089 ,   98021587 ,   98021999 ,   98021207 ,   98021968 ,   98020855 ,   98021551 ,   99400235 ,   98070699 ,   97972950 ,   99222016 ,   98070351 ,   97971659 ,   98020901 ,   98020949 ,   98020474 ,   98021592 ,   98021932 ,   97971790 ,   97971766 ,   97972867 ,   97972283 ,   13432084 ,   99170140 ,   98070868 ,   98070028 ,   98070717 ,   98070296 ,   98070688 ,   99220393 ,   98121080 ,   12474733 ,   99222601 ,   97971952 ,   99222578 ,   98121160 ,   97970769 ,   97970839 ,   99222563 ,   99222322 ,   97970124 ,   97972801 ,   13430060 ,   14180300 ,   99221044 ,   97972664 ,   97970952 ,   97970873 ,   13430773 ,   97972475 ,   97972824 ,   13431359 ,   12471354 ,   99221849 ,   99221798 ,   98120493 ,   98120726 ,   98120696 ,   99222537 ,   99220861 ,   13432293 ,   13434708 ,   97970948 ,   14180791 ,   13433680 ,   13431566 ,   13431924 ,   99222636 ,   99222656 ,   12474581 ,   13433821 ,   12474354 ,   12471791 ,   13433340 ,   13432931 ,   13431284 ,   13440709 ,   13433382 ,   99221243 ,   13433983 ,   97971063 ,   97971874 ,   99222356 ,   99221852 ,   14180397 ,   13430576 ,   12474729 ,   13432281 ,   12473137 ,   13432245 ,   12470955 ,   13431610 ,   13436245 ,   99222612 ,   13434287 ,   12470279 ,   13433445 ,   13431619 ,   12473303 ,   12470403 ,   13431307 ,   98120382 ,   99222532 ,   13435380 ,   98020906 ,   12474665 ,   13430152 ,   99221958 ,   14180515 ,   99222578 ,   12473180 ,   13434080 ,   13439991 ,   97970460 ,   97970303 ,   13430198 ,   13430743 ,   13435220 ,   13438795 ,   13430039 ,   13432326 ,   13431725 ,   13431485 ,   13431224 ,   13433164 ,   99222614 ,   99222331 ,   13431989 ,   13432293 ,   13434726 ,   13431827 ,   13430627 ,   13431950 ,   13430182 ,   13434556 ,   12473423 ,   13436180 ,   13432449 ,   99222375 ,   13431069 ,   13430881 ,   97972801 ,   99222384 ,   13431420 ,   13432862 ,   13432588 ,   12474249 ,   13439388 ,   98021812 ,   99222471 ,   99222379 ,   13437386 ,   99222608 ,   99222423 ,   99221856 ,   13434243 ,   13433174 ,   13430135 ,   13433314 ,   12474465 ,   12473145 ,   13433684 ,   12471384 ,   14180622 ,   13440159 ,   99222418 ,   13431720 ,   12474332 ,   12471844 ,   13431992 ,   13439898 ,   13433072 ,   99222523 ,   13434042 ,   13432775 ,   99222578 ,   99220435 ,   13430344 ,   14180036 ,   13439887 ,   97970496 ,   97970510 ,   13431002 ,   13432432 ,   14180638 ,   13438445 ,   12471365 ,   98120641 ,   13432828 ,   99222539 ,   99220428 ,   12470727 ,   13431121 ,   13434646 ,   12472866 ,   13434339 ,   13433351 ,   13433021 ,   13433746 ,   13433597 ,   14180301 ,   13431965 ,   13433449 ,   99222633 ,   12470294 ,   99222348 ,   98021598 ,   12472916 ,   12472926 ,   12472854 ,   13431168 ,   98020252 ,   13432809 ,   12474227 ,   97970117 ,   13430007 ,   99222505 ,   13440463 ,   12471872 ,   13432893 ,   13430543 ,   14180874 ,   99222047 ,   14181058 ,   12470501 ,   12471513 ,   12474272 ,   13431727 ,   12474057 ,   99220921 ,   12471545 ,   98020372 ,   12470387 ,   13433411 ,   13433763 ,   12472231 ,   13434621 ,   13434369 ,   12471484 ,   13431471 ,   97972256 ,   12471430 ,   13431425 ,   14180319 ,   13431269 ,   13430443 ,   13430263 ,   13433150 ,   13430109 ,   99401327 ,   97971904 ,   13439881 ,   97970664 ,   97971885 ,   97970638 ,   13431396 ,   13430382 ,   12471101 ,   13430160 ,   13435527 ,   13436516 ,   97972950 ,   97972981 ,   97970743 ,   97970687 ,   97970844 ,   97970447 ,   97970796 ,   97972983 ,   99220427 ,   97970914 ,   13437108 ,   12472089 ,   13431393 ,   13431918 ,   99221429 ,   13440255 ,   14180725 ,   99220597 ,   99220574 ,   97970397 ,   14180479 ,   13440421 ,   99220925 ,   98022207 ,   13431186 ,   13431809 ,   13432210 ,   97971939 ,   97971936 ,   12474716 ,   13430489 ,   99222353 ,   99222678 ,   98120538 ,   13439829 ,   13431923 ,   12473802 ,   14180289 ,   13430070 ,   97970291 ,   99222501 ,   99222273 ,   12474114 ,   13432728 ,   13440049 ,   13431059 ,   13431830 ,   13432534 ,   13432513 ,   13433799 ,   13430665 ,   13432915 ,   13431361 ,   13431739 ,   98120478 ,   98120395 ,   13438672 ,   98022167 ,   13438727 ,   13432289 ,   14180895 ,   13432091 ,   13431252 ,   13432255 ,   13433270 ,   13430279 ,   13430538 ,   13430844 ,   13431479 ,   14180264 ,   14180898 ,   13434337 ,   13440017 ,   12474538 ,   97971882 ,   13431358 ,   13433089 ,   13431063 ,   99221063 ,   99222367 ,   12470447 ,   14180278 ,   12471371 ,   12470672 ,   13430539 ,   99221119 ,   13431568 ,   13432824 ,   13431741 ,   13431617 ,   13432250 ,   14181041 ,   13437404 ,   13437428 ,   12473076 ,   12473089 ,   13430016 ,   13430020 ,   13431344 ,   13430258 ,   12471929 ,   13431184 ,   13431443 ,   13431774 ,   13433246 ,   13431631 ,   13431697 ,   13431116 ,   13431119 ,   13430192 ,   13430604 ,   13431519 ,   13433865 ,   12471630 ,   13440771 ,   13436018 ,   13439275 ,   13439683 ,   13435913 ,   13431418 ,   13430093 ,   13437435 ,   13434268 ,   13432887 ,   13437573 ,   99222578 ,   13431642 ,   97971605 ,   13436546 ,   12471697 ,   12471390 ,   13431701 ,   13434523 ,   13432514 ,   13431728 ,   13432122 ,   13433114 ,   13430480 ,   13432090 ,   13430955 ,   14180208 ,   13439760 ,   99222337 ,   13430273 ,   97971904 ,   12473444 ,   14180765 ,   13430148 ,   12472880 ,   13434655 ,   97972588 ,   13432324 ,   13440380 ,   13433045 ,   13431994 ,   13438524 ,   99220352 ,   99220854 ,   14180184 ,   13431384 ,   13431127 ,   13438450 ,   99220882 ,   13431162 ,   13433002 ,   13432723 ,   13431908 ,   99222759 ,   13432874 ,   13435219 ,   13430262 ,   12472143 ,   98120435 ,   97972490 ,   97970390 ,   13438345 ,   97970470 ,   13434431 ,   13434442 ,   99220353 ,   13433454 ,   13431991 ,   12471555 ,   14180342 ,   97970621 ,   13431543 ,   12471517 ,   13438440 ,   12472157 ,   12473611 ,   13432276 ,   13440636 ,   99221054 ,   99222361 ,   99221310 ,   13430256 ,   13431360 ,   13433041 ,   12471186 ,   13430195 ,   13430815 ,   13436356 ,   13430201 ,   12471405 ,   12474235 ,   97970854 ,   99222615 ,   13430562 ,   99222479 ,   12473981 ,   12470581 ,   13432710 ,   12471922 ,   13438017 ,   97970362 ,   13431921 ,   97970326 ,   97971803 ,   13433335 ,   99222377 ,   13435064 ,   13435427 ,   99401385 ,   99401617 ,   99222126 ,   13432525 ,   13436949 ,   13430681 ,   13438671 ,   98020979 ,   12473779 ,   12471513 ,   13438805 ,   13438767 ,   12471749 ,   13433432 ,   99220613 ,   99222507 ,   14180029 ,   12471778 ,   13437662 ,   13431798 ,   13430736 ,   12471804 ,   12471355 ,   13430130 ,   13433176 ,   13433596 ,   13439582 ,   97970578 ,   13440172 ,   13430868 ,   13430816 ,   13432788 ,   13431480 ,   12472460 ,   12474554 ,   13434676 ,   13435905 ,   99221060 ,   13433299 ,   13434817 ,   99222089 ,   97970974 ,   99170292 ,   13440406 ,   13434759 ,   12471030 ,   14181047 ,   12471742 ,   13432918 ,   12474110 ,   13433825 ,   13432794 ,   12470871 ,   97970116 ,   13439657 ,   98021726 ,   13440224 ,   12474186 ,   99221786 ,   14180621 ,   13431541 ,   13435455 ,   13430207 ,   14180440 ,   13430792 ,   13431829 ,   14180178 ,   12470619 ,   12474746 ,   13439433 ,   13434210 ,   13436451 ,   99222280 ,   97970389 ,   12470644 ,   13431360 ,   12474613 ,   13430599 ,   12473175 ,   12473377 ,   13434603 ,   13434020 ,   98020101 ,   98020417 ,   98022207 ,   98021803 ,   12473807 ,   14180351 ,   13434095 ,   12473182 ,   12473203 ,   12474695 ,   99222333 ,   13431397 ,   13430298 ,   97970559 ,   13431474 ,   12474341 ,   12470615 ,   13431951 ,   13431755 ,   13430853 ,   12474717 ,   13438602 ,   99220388 ,   14180450 ,   13437416 ,   99401584 ,   99401641 ,   12470520 ,   13430030 ,   12470858 ,   13432939 ,   13435085 ,   12470465 ,   13430945 ,   98120683 ,   98120690 ,   98120692 ,   98120851 ,   98121198 ,   98121177 ,   98120765 ,   98121212 ,   98121228 ,   13432762 ,   12471473 ,   13431281 ,   13432411 ,   13438839 ,   99220844 ,   13438699 ,   13438677 ,   12471452 ,   13439841 ,   99222720 ,   99401657 ,   13440154 ,   99222498 ,   13431663 ,   99222476 ,   13434655 ,   13430325 ,   14180328 ,   13433568 ,   13431193 ,   13433097 ,   12471445 ,   14180997 ,   12473546 ,   13438206 ,   97970487 ,   97970573 ,   97970579 ,   13438631 ,   13439025 ,   99222765 ,   13432838 ,   14180860 ,   13430778 ,   12473176 ,   12470399 ,   13430675 ,   14180159 ,   97972797 ,   12470573 ,   13431775 ,   12472960 ,   14180734 ,   13431744 ,   13430373 ,   13439497 ,   13434520 ,   13432237 ,   14180310 ,   13440659 ,   13430131 ,   14180053 ,   13435327 ,   13436405 ,   13436423 ,   13432332 ,   13437020 ,   13438629 ,   14180977 ,   14180651 ,   13432397 ,   14180131 ,   13433506 ,   13439733 ,   13435275 ,   13436547 ,   12470736 ,   99222578 ,   99220589 ,   97970477 ,   13439822 ,   13433190 ,   13436022 ,   13431029 ,   12471666 ,   99400011 ,   13433257 ,   13430319 ,   13438556 ,   13435539 ,   13432632 ,   13432221 ,   13432481 ,   13433169 ,   14180354 ,   13430994 ,   13433330 ,   13434671 ,   99222622 ,   99222389 ,   99222501 ,   13433194 ,   13430395 ,   13431083 ,   12470257 ,   12473173 ,   13435152 ,   14180094 ,   13431537 ,   13434473 ,   13433438 ,   14180658 ,   13432406 ,   99222730 ,   99222401 ,   97970395 ,   97970690 ,   97970755 ,   97970788 ,   13439062 ,   13431573 ,   13432930 ,   12472023 ,   13435316 ,   12474702 ,   14180269 ,   12473627 ,   13434245 ,   13434301 ,   99221489 ,   99221071 ,   99222568 ,   99222739 ,   99222639 ,   99220646 ,   13433565 ,   13432050 ,   13437722 ,   99220930 ,   13436046 ,   13440126 ,   13440212 ,   13430876 ,   12472500 ,   12471057 ,   14180749 ,   13432776 ,   14180311 ,   13435358 ,   13436443 ,   99222590 ,   99220908 ,   12471146 ,   12472109 ,   13432212 ,   13438476 ,   13431588 ,   97972950 ,   99220584 ,   12473120 ,   12470872 ,   13432662 ,   97970040 ,   97971938 ,   12473967 ,   12471672 ,   13434500 ,   13432552 ,   97970992 ,   12470317 ,   13435270 ,   13435585 ,   13436034 ,   13436181 ,   99220622 ,   13432755 ,   13430654 ,   13439030 ,   12471281 ,   12471354 ,   99220462 ,   13432014 ,   13440040 ,   13437132 ,   13431449 ,   99221429 ,   13435974 ,   13435238 ,   13431552 ,   13431819 ,   99222713 ,   99222439 ,   99221239 ,   13433235 ,   13436226 ,   13438711 ,   99222540 ,   13431822 ,   12470375 ,   13434249 ,   99744476 ,   13433143 ,   13436216 ,   97971597 ,   13430996 ,   13433258 ,   13435109 ,   14181090 ,   13432693 ,   99222738 ,   13430471 ,   13431404 ,   13432842 ,   14180065 ,   13437483 ,   13430399 ,   13433839 ,   13431782 ,   13431517 ,   99222399 ,   13431635 ,   13431639 ,   13436277 ,   12470506 ,   99220439 ,   99222692 ,   13440109 ,   99222406 ,   13434550 ,   99220848 ,   13438812 ,   99222642 ,   13430125 ,   12471135 ,   13435415 ,   13432376 ,   13437425 ,   13433283 ,   13437385 ,   13439089 ,   99222695 ,   13433444 ,   13433451 ,   13433558 ,   13440615 ,   13433410 ,   99220900 ,   13432969 ,   13440202 ,   12470973 ,   13432575 ,   14180410 ,   13430928 ,   13434649 ,   13439075 ,   98020117 ,   13439057 ,   13430497 ,   12471859 ,   13433175 ,   13439900 ,   13439851 ,   97972983 ,   99222046 ,   99220866 ,   12471600 ,   12470351 ,   13431196 ,   14181069 ,   12474734 ,   12474429 ,   12470550 ,   14180220 ,   12474317 ,   13433296 ,   13433289 ,   97970902 ,   97971667 ,   13431194 ,   12472889 ,   13432345 ,   13432654 ,   13431201 ,   14180766 ,   13436096 ,   13431619 ,   99222654 ,   14180258 ,   12474714 ,   98022264 ,   12473896 ,   97970590 ,   97970663 ,   13434686 ,   13430555 ,   13430460 ,   14180887 ,   13432957 ,   13432260 ,   13430831 ,   13431309 ,   12474435 ,   99222365 ,   13433437 ,   12471337 ,   13435946 ,   97970443 ,   12473471 ,   99220412 ,   12473751 ,   99222014 ,   99221257 ,   13432925 ,   13430845 ,   13435079 ,   12473009 ,   99220638 ,   13433287 ,   12474266 ,   13432501 ,   99222395 ,   99220488 ,   12473267 ,   13434524 ,   13432256 ,   13436262 ,   13438909 ,   99744321 ,   99222718 ,   13432028 ,   99220436 ,   12472468 ,   12470350 ,   12470676 ,   12474785 ,   99220928 ,   99221172 ,   99222757 ,   13435998 ,   13435961 ,   99222262 ,   13436223 ,   13435375 ,   13435440 ,   13439698 ,   99222751 ,   13437487 ,   13433123 ,   13438356 ,   13438527 ,   99220923 ,   99220406 ,   12471409 ,   13434656 ,   13433149 ,   12472760 ,   98022207 ,   99222181 ,   99222491 ,   13435130 ,   13435408 ,   13435049 ,   13430041 ,   97970415 ,   12471644 ,   99220602 ,   13434024 ,   99220498 ,   14180998 ,   97970286 ,   98020428 ,   13440139 ,   99222171 ,   13433179 ,   12473483 ,   13439694 ,   13430363 ,   13440601 ,   13437365 ,   99222346 ,   14180214 ,   14180132 ,   13432462 ,   13434530 ,   13436209 ,   13439922 ,   13430772 ,   13439496 ,   13433627 ,   13436025 ,   14181077 ,   14180605 ,   14180423 ,   12471050 ,   12472876 ,   13439896 ,   13436529 ,   13434661 ,   12474025 ,   99221643 ,   13434475 ,   13434240 ,   13434638 ,   97970339 ,   13436365 ,   99220868 ,   14181024 ,   14180907 ,   14181050 ,   99221170 ,   97970125 ,   97970364 ,   13432391 ,   97972791 ,   99222623 ,   12473460 ,   13433816 ,   99220489 ,   99222371 ,   99221840 ,   97970614 ,   13432178 ,   13431099 ,   13431650 ,   13437180 ,   13435997 ,   99222710 ,   98120671 ,   99220621 ,   13433559 ,   13431102 ,   12474000 ,   13434155 ,   12474633 ,   13436030 ,   99401667 ,   13436255 ,   13436306 ,   97970335 ,   99401227 ,   97970623 ,   99222582 ,   13438472 ,   97970526 ,   13430284 ,   13432770 ,   13434041 ,   13434689 ,   13440290 ,   13432409 ,   13438509 ,   12470911 ,   13430498 ,   13436714 ,   13430145 ,   99220444 ,   12473866 ,   12471904 ,   98022262 ,   12473023 ,   97971868 ,   13432890 ,   12472018 ,   99222470 ,   13431368 ,   13433054 ,   13432576 ,   13435197 ,   13431958 ,   99220555 ,   97970003 ,   13433191 ,   12473171 ,   12472888 ,   12473859 ,   13435437 ,   97972952 ,   13437794 ,   13430474 ,   13433091 ,   13431619 ,   13432765 ,   13433093 ,   12471724 ,   13438598 ,   99222749 ,   13438601 ,   13431170 ,   13436161 ,   13434769 ,   98020399 ,   13430236 ,   13433251 ,   12472098 ,   12473115 ,   13433171 ,   12474053 ,   13432296 ,   99222319 ,   99222660 ,   99222275 ,   97970480 ,   97970444 ,   99222307 ,   13436019 ,   99220866 ,   13436535 ,   13436542 ,   14180747 ,   12471437 ,   12472997 ,   12472978 ,   13434696 ,   13433290 ,   13435860 ,   13432751 ,   14180707 ,   13435351 ,   13432563 ,   13435183 ,   13431674 ,   13431431 ,   98121018 ,   98120863 ,   98120830 ,   13437496 ,   12470351 ,   98121125 ,   13439669 ,   99220667 ,   99222711 ,   99222252 ,   13436971 ,   13438783 ,   13438454 ,   99220159 ,   99222341 ,   12473001 ,   13438826 ,   13431904 ,   14180780 ,   13438821 ,   12471756 ,   99221431 ,   13434637 ,   13432871 ,   13437737 ,   13433422 ,   13434571 ,   99220445 ,   97970461 ,   97970358 ,   99222715 ,   99400830 ,   13440208 ,   99220559 ,   13436649 ,   14180129 ,   99220897 ,   13430303 ,   13438619 ,   13431633 ,   13437663 ,   99222416 ,   13436620 ,   99221695 ,   13431993 ,   13439772 ,   13434577 ,   99222165 ,   99222352 ,   99222267 ,   13439863 ,   99222709 ,   14180223 ,   13437685 ,   13431704 ,   13430036 ,   12472998 ,   97971965 ,   99744235 ,   99221938 ,   13433440 ,   99221639 ,   13435893 ,   13436013 ,   99221928 ,   13432814 ,   13432943 ,   13432761 ,   13435337 ,   13435579 ,   99222754 ,   99222328 ,   97971652 ,   98022207 ,   97970987 ,   97971591 ,   97970545 ,   13435319 ,   13435317 ,   13434208 ,   13434241 ,   98022275 ,   99222680 ,   13438441 ,   13437148 ,   13437270 ,   99401593 ,   13433172 ,   98411600 ,   13439470 ,   12472287 ,   13432988 ,   99400030 ,   99222764 ,   13440086 ,   97972646 ,   97972688 ,   97970616 ,   99222116 ,   13432678 ,   13431388 ,   13439887 ,   13440184 ,   13440186 ,   13433100 ,   99220502 ,   14180073 ,   13440007 ,   13440068 ,   13440550 ,   12470578 ,   13432652 ,   14180076 ,   13438645 ,   99221585 ,   99222360 ,   13435359 ,   12474618 ,   99222489 ,   98022164 ,   97971911 ,   13438585 ,   13439727 ,   14180645 ,   14180742 ,   14180884 ,   14180938 ,   14180771 ,   14180833 ,   13431332 ,   13432895 ,   13431988 ,   13430606 ,   13431064 ,   12472993 ,   99222681 ,   99220464 ,   13431262 ,   13431078 ,   13431583 ,   13435485 ,   13434175 ,   13431579 ,   13432934 ,   13432483 ,   97970079 ,   13432559 ,   13432802 ,   13433647 ,   13432670 ,   13434344 ,   12474015 ,   99221291 ,   99222648 ,   99222537 ,   13433870 ,   99220379 ,   12472781 ,   99222562 ,   13432136 ,   99221734 ,   99220795 ,   14180808 ,   13439402 ,   13434118 ,   13430514 ,   13432972 ,   13437263 ,   99221265 ,   99220458 ,   99222598 ,   99222292 ,   13436385 ,   97970454 ,   13439684 ,   13432322 ,   13432615 ,   12473710 ,   97970585 ,   97971109 ,   97972479 ,   99220243 ,   13438475 ,   99222565 ,   13432098 ,   13434368 ,   97971897 ,   13436275 ,   12474321 ,   13433707 ,   13433815 ,   13436588 ,   13433033 ,   12474302 ,   13434615 ,   14180209 ,   13436544 ,   14180232 ,   99222354 ,   97970412 ,   13437320 ,   99222646 ,   99222357 ,   99222627 ,   99221289 ,   14181052 ,   13435311 ,   13439371 ,   13438573 ,   12474163 ,   13433241 ,   99222541 ,   99220595 ,   99220577 ,   99222512 ,   13439379 ,   13433507 ,   14180909 ,   12470576 ,   13431100 ,   13439534 ,   13433834 ,   99220581 ,   12474042 ,   13431129 ,   13439658 ,   12472777 ,   13432029 ,   13431654 ,   13438885 ,   14180794 ,   97971804 ,   97972720 ,   99220929 ,   13431772 ,   99221184 ,   99222657 ,   99222266 ,   97970398 ,   13434059 ,   13432215 ,   13432156 ,   12472802 ,   99221006 ,   12473544 ,   13439724 ,   13438589 ,   13433157 ,   99221801 ,   13433464 ,   97970459 ,   99221824 ,   99221240 ,   99222342 ,   12471546 ,   13436090 ,   13431895 ,   13437847 ,   13436218 ,   13435477 ,   13432614 ,   12472074 ,   12473457 ,   13432517 ,   13436240 ,   13436887 ,   13430022 ,   13430058 ,   14180617 ,   14180554 ,   13435382 ,   13438710 ,   99222301 ,   13438790 ,   13430979 ,   13438478 ,   13438538 ,   97971061 ,   13433964 ,   13437972 ,   13432526 ,   99222298 ,   97970546 ,   13434450 ,   13437599 ,   13431066 ,   99221483 ,   99221201 ,   12472298 ,   13439711 ,   13433375 ,   99400862 ,   13430738 ,   13438604 ,   13431459 ,   99221500 ,   12473226 ,   13438519 ,   13434984 ,   13435122 ,   13434990 ,   99401748 ,   13437366 ,   12473372 ,   13431738 ,   13432758 ,   13430972 ,   99401587 ,   99221123 ,   13439103 ,   13437029 ,   13431212 ,   13438888 ,   13435841 ,   13436295 ,   99221221 ,   13432240 ,   12471354 ,   99221143 ,   99221160 ,   13434044 ,   13437067 ,   13434562 ,   14180300 ,   12474742 ,   14180421 ,   99221411 ,   12474636 ,   97970760 ,   13439676 ,   14180322 ,   13432465 ,   99222428 ,   13438667 ,   13438646 ,   13432285 ,   12471264 ,   13432699 ,   13439672 ,   13431395 ,   13436196 ,   13435949 ,   97970445 ,   97971948 ,   97970871 ,   13433840 ,   13430738 ,   99222278 ,   99222442 ,   13430436 ,   13436106 ,   99220619 ,   13430890 ,   99220624 ,   99222683 ,   12473794 ,   14181051 ,   99401596 ,   99222649 ,   99222499 ,   13434672 ,   99221275 ,   98020128 ,   98021824 ,   99222637 ,   13430976 ,   13437150 ,   13437811 ,   13437882 ,   99221728 ,   99221130 ,   99220424 ,   13433477 ,   12472808 ,   13432226 ,   13435357 ,   99222706 ,   13432914 ,   99221211 ,   13430750 ,   13439117 ,   13437348 ,   13437277 ,   13437903 ,   13436130 ,   99222686 ,   13438501 ,   13433343 ,   13430762 ,   12474499 ,   13430487 ,   13439624 ,   13439202 ,   13439784 ,   13430744 ,   99222574 ,   97971891 ,   99221118 ,   99222150 ,   13431721 ,   97970120 ,   99221379 ,   14180931 ,   13436237 ,   13434402 ,   13434555 ,   13433452 ,   13436063 ,   13435158 ,   13436078 ,   12471569 ,   13433730 ,   13439021 ,   13439135 ,   13438726 ,   98121196 ,   13436684 ,   13432272 ,   13436584 ,   13434687 ,   13434688 ,   13435587 ,   13438304 ,   12471790 ,   13436765 ,   13436353 ,   13435539 ,   13438613 ,   13432488 ,   13432521 ,   13436194 ,   12471539 ,   99220514 ,   97970794 ,   13440178 ,   13440204 ,   13440173 ,   13440125 ,   13440213 ,   13440080 ,   13440196 ,   13440209 ,   99222391 ,   13439889 ,   13436344 ,   13437624 ,   99221196 ,   99221188 ,   12473801 ,   99221259 ,   13437120 ,   13430265 ,   13432821 ,   13432801 ,   12471071 ,   99220453 ,   13437692 ,   13432746 ,   13433210 ,   99221648 ,   13438230 ,   13430790 ,   13435172 ,   99220585 ,   99222182 ,   12470621 ,   12473400 ,   13434306 ,   13434258 ,   13434718 ,   13435263 ,   13434122 ,   13438455 ,   13438412 ,   13438470 ,   13438822 ,   12472013 ,   13432000 ,   99220918 ,   99222679 ,   99222486 ,   99221503 ,   99221524 ,   13436205 ,   12474607 ,   13433976 ,   13434037 ,   12472153 ,   12473899 ,   12473858 ,   12473863 ,   12471618 ,   99220913 ,   13432356 ,   12471179 ,   13431455 ,   12470618 ,   12474658 ,   13439028 ,   13434828 ,   13438857 ,   13439041 ,   13439019 ,   13438856 ,   13439032 ,   13438851 ,   13437501 ,   13431475 ,   12474773 ,   13431481 ,   12470457 ,   13431558 ,   99220433 ,   99744517 ,   13439775 ,   99401691 ,   13437137 ,   13436307 ,   99220526 ,   99222466 ,   99221387 ,   99222702 ,   13438840 ,   14180679 ,   13434608 ,   13434536 ,   99221567 ,   99221786 ,   99222131 ,   99222334 ,   99222454 ,   13439919 ,   99220809 ,   13439717 ,   13439504 ,   13439483 ,   99222653 ,   13440538 ,   13430076 ,   13430107 ,   13439828 ,   13439744 ,   13439686 ,   13439808 ,   13439832 ,   13439825 ,   13439834 ,   13439691 ,   98022314 ,   99221323 ,   13430584 ,   13434035 ,   13434429 ,   13438868 ,   13439043 ,   12472955 ,   99220459 ,   13438784 ,   99220365 ,   13438451 ,   12471924 ,   13430888 ,   13432650 ,   13431762 ,   13440148 ,   13437172 ,   12473635 ,   13433345 ,   14180909 ,   14180051 ,   14181053 ,   13431081 ,   13430206 ,   13437540 ,   98022235 ,   99222469 ,   13440242 ,   12473164 ,   13436386 ,   12470739 ,   13436491 ,   13436387 ,   14180567 ,   13437211 ,   13435323 ,   99744506 ,   13432390 ,   99222520 ,   99221288 ,   13439650 ,   13438354 ,   13437863 ,   13437928 ,   13430285 ,   13440214 ,   12471440 ,   13437291 ,   13439078 ,   13439826 ,   13435245 ,   13434117 ,   13438411 ,   98020970 ,   99222276 ,   13432384 ,   12473080 ,   13439191 ,   13437334 ,   13432398 ,   13437505 ,   13437387 ,   12474619 ,   99220588 ,   99220568 ,   99220570 ,   99401625 ,   99220670 ,   99744581 ,   99222269 ,   99221482 ,   99222388 ,   13438833 ,   12471507 ,   99401670 ,   99221261 ,   14180381 ,   13436118 ,   12470589 ,   13439196 ,   13440203 ,   13434981 ,   99222742 ,   99221612 ,   99401763 ,   99222620 ,   99221208 ,   99222310 ,   13439845 ,   12472521 ,   99220815 ,   99222728 ,   99221189 ,   13430496 ,   13439171 ,   99222715 ,   99220662 ,   13440211 ,   13433505 ,   13436458 ,   13433262 ,   13437169 ,   13431535 ,   13430455 ,   13440119 ,   12473642 ,   99222393 ,   13440221 ,   13431241 ,   99221232 ,   99222513 ,   99220461 ,   13439633 ,   13439715 ,   13439729 ,   99222264 ,   99401673 ,   99401623 ,   99221205 ,   13438517 ,   99222543 ,   99401581 ,   13430737 ,   14180891 ,   14180944 ,   13438877 ,   99222075 ,   13436734 ,   97970044 ,   98022169 ,   13434273 ,   12473216 ,   13435333 ,   13440518 ,   13434292 ,   12474218 ,   99222516 ,   13438779 ,   14180701 ,   13432657 ,   13432304 ,   13439666 ,   13439756 ,   13431629 ,   13432660 ,   13433308 ,   99222410 ,   13437135 ,   13434234 ,   13439789 ,   99221074 ,   13432900 ,   13437208 ,   13431277 ,   14180514 ,   12473595 ,   13436204 ,   99220784 ,   99221117 ,   13435491 ,   13435357 ,   13440265 ,   13431039 ,   98022294 ,   13430433 ,   13431068 ,   13438537 ,   13435119 ,   13435426 ,   14180667 ,   13440379 ,   13434640 ,   13434763 ,   13437423 ,   13430450 ,   13438707 ,   13438794 ,   13432006 ,   14180630 ,   99220565 ,   99220737 ,   13439099 ,   99220914 ,   13437843 ,   97970599 ,   13439061 ,   13438875 ,   12470841 ,   14180325 ,   99220676 ,   99221255 ,   13430006 ,   12470672 ,   13432355 ,   99222763 ,   14180380 ,   13440639 ,   12472259 ,   13439763 ,   13430706 ,   99222311 ,   99222315 ,   13433934 ,   99222125 ,   12471218 ,   13434395 ,   13433472 ,   12471626 ,   99221985 ,   12473898 ,   13438336 ,   99221066 ,   13440146 ,   13431103 ,   99221412 ,   14180829 ,   13438892 ,   99221456 ,   12472049 ,   13436638 ,   13430198 ,   13439723 ,   99222553 ,   13436429 ,   13438546 ,   13434690 ,   12474382 ,   12470653 ,   99222005 ,   13430018 ,   99222602 ,   13437443 ,   12473854 ,   99222556 ,   12473382 ,   12473152 ,   12473155 ,   13433502 ,   99220829 ,   13433480 ,   13433509 ,   99220364 ,   13437313 ,   13438550 ,   12470517 ,   99221062 ,   99220580 ,   13430374 ,   13440270 ,   13433339 ,   13430953 ,   99221083 ,   99222154 ,   99222445 ,   99221003 ,   99220859 ,   99222265 ,   13435543 ,   13435954 ,   14180835 ,   13439259 ,   13437481 ,   14180855 ,   13437378 ,   13436836 ,   13437420 ,   14180712 ,   14180714 ,   99221936 ,   99222012 ,   99221836 ,   99222450 ,   13439778 ,   13434006 ,   13439739 ,   13433434 ,   14180843 ,   14180496 ,   12470354 ,   13437238 ,   13439620 ,   99221043 ,   13438447 ,   99222685 ,   13430071 ,   99221819 ,   99221318 ,   13436732 ,   99222126 ,   99222659 ,   13437413 ,   13440268 ,   13432898 ,   99220837 ,   13440604 ,   13438343 ,   13438355 ,   99221092 ,   13433247 ,   13431114 ,   99222336 ,   99221530 ,   12473066 ,   13438879 ,   13438881 ,   99220416 ,   13438008 ,   13432612 ,   99222180 ,   13438724 ,   13438639 ,   13436690 ,   13439722 ,   99220658 ,   99220954 ,   13432476 ,   99222533 ,   14180752 ,   14181025 ,   13433399 ,   14180957 ,   99221237 ,   99221304 ,   99222650 ,   99401664 ,   13433923 ,   13432196 ,   99220671 ,   13440298 ,   13440458 ,   99221335 ,   99401591 ,   99222285 ,   13439887 ,   99221719 ,   12473109 ,   13432556 ,   13433616 ,   13432457 ,   13433346 ,   99221182 ,   99222415 ,   99222444 ,   13433439 ,   14180910 ,   13437455 ,   99222152 ,   99401671 ,   13438653 ,   14180909 ,   13439811 ,   13439795 ,   99220867 ,   13434049 ,   13434000 ,   99220823 ,   14180753 ,   13436318 ,   99222365 ,   13437268 ,   13432208 ,   13440245 ,   13440371 ,   13437215 ,   13433938 ,   13438514 ,   13434833 ,   13437236 ,   99222506 ,   13434453 ,   99222736 ,   99221535 ,   99222261 ,   13433548 ,   12471077 ,   12473736 ,   13438437 ,   99221198 ,   13432798 ,   13430124 ,   13440014 ,   12470607 ,   13431900 ,   12471138 ,   99221401 ,   99220813 ,   13440323 ,   13438208 ,   13432711 ,   12472008 ,   13430278 ,   12474512 ,   13430407 ,   13430417 ,   12474529 ,   13433112 ,   13438690 ,   99221274 ,   12474220 ,   13439450 ,   13439105 ,   98121329 ,   98121022 ,   98121126 ,   98121300 ,   98121131 ,   98120719 ,   98121346 ,   98121028 ,   98121068 ,   98120863 ,   13439738 ,   13439367 ,   13440307 ,   12470616 ,   99222609 ,   99222563 ,   12472824 ,   12473824 ,   99221227 ,   13434588 ,   13433917 ,   13432179 ,   13432597 ,   98121219 ,   98070965 ,   13438900 ,   13433546 ,   13431153 ,   13435179 ,   99401616 ,   99401717 ,   99222449 ,   99221687 ,   99220856 ,   13438004 ,   13439538 ,   99220788 ,   13432837 ,   99222629 ,   13430593 ,   99221578 ,   99220614 ,   12473950 ,   13430422 ,   14181071 ,   99222564 ,   13436723 ,   13432933 ,   99220691 ,   99222594 ,   99221053 ,   99222529 ,   99221229 ,   13431074 ,   13430269 ,   13439204 ,   99221667 ,   99222586 ,   13433204 ,   13439092 ,   99221099 ,   99221216 ,   99221283 ,   99220576 ,   13431900 ,   99221608 ,   13437578 ,   13440483 ;     *** 2) List Number 13100, Serial Numbers:     13530228 ,   13530338 ,   13530203 ,   13530205 ,   13530328 ,   13530312 ,   13530387 ,   13530209 ,   13530312 ,   12771348 ,   13530386 ,   13530385 ,   13530331 ,   13530373 ,   13530370 ;     *** 3) List Number 13150, Serial Numbers:     98431537 ,   98430945 ,   98431018 ,   98431041 ,   13637227 ,   99271058 ,   98431008 ,   99270417 ,   98930032 ,   98930073 ,   98930074 ,   98930083 ,   13631592 ,   99749369 ,   98430113 ,   98430192 ,   99450473 ,   98410749 ,   13630146 ,   98431142 ,   98431205 ,   13630031 ,   98930058 ,   98431409 ,   98410005 ,   98410015 ,   98411582 ,   98410379 ,   98411404 ,   99450476 ,   98410027 ,   98410353 ,   98410363 ,   98411239 ,   98431167 ,   98410231 ,   98410048 ,   98410580 ,   99270394 ,   13631348 ,   98431086 ,   99749251 ,   98410752 ,   99450050 ,   98410387 ,   99749532 ,   13631527 ,   98410741 ,   98410836 ,   98930402 ,   99749112 ,   98430321 ,   98411440 ,   12720251 ,   99450479 ,   98431440 ,   13636963 ,   98431000 ,   99270409 ,   98431149 ,   98411588 ,   98410556 ,   98410563 ,   98410910 ,   13630816 ,   98431200 ,   98410569 ,   98411468 ,   99450565 ,   98432002 ,   12720318 ,   13630186 ,   13630384 ,   98410078 ,   99450463 ,   98410722 ,   99450478 ,   98410579 ,   98410596 ,   98411312 ,   98411581 ,   98930497 ,   98410972 ,   98411080 ,   98411582 ,   98411622 ,   98410622 ,   99270406 ,   99270571 ,   98430968 ,   99270689 ,   98432021 ,   98430724 ,   99749554 ,   13630112 ,   12721156 ,   99749431 ,   98431174 ,   13631400 ,   13630492 ,   98432187 ,   99749552 ,   98432043 ,   12721163 ,   98431184 ,   13630909 ,   13631823 ,   12721091 ,   13635636 ,   13630480 ,   13631107 ,   13631673 ,   13630317 ,   13630426 ,   99270387 ,   13630880 ,   13631711 ,   13636201 ,   12720577 ,   13630338 ,   13630015 ,   12721082 ,   13633967 ,   12720583 ,",
      "more_code_info": "12720873 ,   12720160 ,   13632682 ,   13633145 ,   12720055 ,   13630583 ,   12721276 ,   13636416 ,   13636362 ,   13631294 ,   13637031 ,   12720614 ,   12720554 ,   13630739 ,   13631251 ,   12721250 ,   99270878 ,   98431435 ,   13631811 ,   13634026 ,   99749538 ,   13630419 ,   13631754 ,   12720275 ,   13631063 ,   12720631 ,   13633172 ,   12720765 ,   13631177 ,   13631192 ,   13631010 ,   12721182 ,   13631863 ,   12720082 ,   12721184 ,   13636378 ,   13631227 ,   99270505 ,   13633174 ,   13633143 ,   13630948 ,   12721167 ,   12720621 ,   13630976 ,   12720410 ,   13631671 ,   13630109 ,   13630489 ,   12721118 ,   13635670 ,   13635603 ,   99749580 ,   13632189 ,   12720382 ,   13637420 ,   13637301 ,   13631136 ,   13630868 ,   13630858 ,   13630640 ,   99270058 ,   13630536 ,   13632824 ,   13631882 ,   99749575 ,   13633173 ,   13630924 ,   13635652 ,   13631744 ,   13630900 ,   13630633 ,   13630740 ,   13631345 ,   98431170 ,   13631809 ,   13631652 ,   98430984 ,   12720822 ,   98431158 ,   13630401 ,   13635834 ,   13635794 ,   13631210 ,   13631243 ,   12720388 ,   98410485 ,   12721108 ,   13630585 ,   12720917 ,   12720850 ,   12720480 ,   13631691 ,   99749571 ,   98430910 ,   13630062 ,   13630823 ,   98432023 ,   13631551 ,   13631311 ,   13631605 ,   13630911 ,   13633941 ,   12720210 ,   12721018 ,   13631850 ,   13631923 ,   99270067 ,   98431140 ,   98431121 ,   13632028 ,   13631916 ,   13631548 ,   12720032 ,   13631451 ,   13631559 ,   13630537 ,   13635336 ,   13630596 ,   13635403 ,   13631515 ,   12720219 ,   13634316 ,   13630159 ,   12720125 ,   13632002 ,   12720791 ,   13630165 ,   13631300 ,   13631069 ,   13636142 ,   13631398 ,   13632160 ,   13630505 ,   12721209 ,   13631993 ,   99270698 ,   99270829 ,   13630726 ,   98432275 ,   99270470 ,   13636580 ,   13631655 ,   13631622 ,   12720720 ,   13632104 ,   99749540 ,   13637859 ,   99270747 ,   99270327 ,   99749514 ,   13630158 ,   98432111 ,   13636539 ,   13630422 ,   13630337 ,   13633129 ,   12720013 ,   13631009 ,   13631918 ,   13631898 ,   13630397 ,   13631763 ,   12721245 ,   12721227 ,   13630695 ,   13631800 ,   13637326 ,   99749506 ,   13632582 ,   13631545 ,   13636822 ,   13631344 ,   12720162 ,   12721014 ,   12720537 ,   13630837 ,   13632818 ,   13631241 ,   13631714 ,   13630567 ,   13630071 ,   13632035 ,   13634335 ,   12721168 ,   12721093 ,   13636349 ,   13636285 ,   13636483 ,   12720373 ,   12721044 ,   98432194 ,   98431010 ,   98431082 ,   98430964 ,   13634368 ,   13630322 ,   13636363 ,   12720228 ,   13631838 ,   13631604 ,   13634238 ,   13631055 ,   13631350 ,   13631258 ,   13630918 ,   13630838 ,   13631821 ,   12720630 ,   12721210 ,   13635545 ,   13637151 ,   13633127 ,   13630334 ,   13636171 ,   13631661 ,   13637779 ,   12720544 ,   13633944 ,   13630939 ,   12720681 ,   12720802 ,   13635936 ,   13631026 ,   13630460 ,   13631862 ,   13635249 ,   13635292 ,   12720650 ,   13634514 ,   12720202 ,   13631193 ,   13634240 ,   13635438 ,   13635655 ,   13635597 ,   13631404 ,   13631143 ,   13635799 ,   12720153 ,   13633170 ,   13630581 ,   12721226 ,   13631128 ,   13632683 ,   13636106 ,   98431518 ,   13630221 ,   13636197 ,   98430970 ,   13636222 ,   13636259 ,   13636272 ,   13630716 ,   12720920 ,   12720494 ,   13635896 ,   13636408 ,   13633607 ,   13631117 ,   13630499 ,   13634126 ,   13631339 ,   13630842 ,   12720137 ,   13630810 ,   13630682 ,   13631734 ,   12720392 ,   13630016 ,   12720668 ,   13635419 ,   13630815 ,   13636220 ,   98430971 ,   12720425 ,   13632203 ,   99270141 ,   13631563 ,   12721160 ,   13630915 ,   12721129 ,   13632186 ,   13630182 ,   13630709 ,   13630701 ,   12721101 ,   13631431 ,   13631500 ,   12720572 ,   12721046 ,   99270125 ,   99749527 ,   13631608 ,   13635993 ,   12720302 ,   13631040 ,   13632250 ,   98432177 ,   98410238 ,   98410671 ,   99749541 ,   12720114 ,   99749509 ,   13631002 ,   13631315 ,   99749523 ,   99749539 ,   13630378 ,   13635607 ,   13630267 ,   12721140 ,   13631245 ,   12720586 ,   13636487 ,   13630790 ,   13630827 ,   13630605 ,   12720448 ,   12720787 ,   13630775 ,   12721157 ,   13635590 ,   13631615 ,   13631226 ,   13631038 ,   12721047 ,   12721217 ,   13630434 ,   12720111 ,   99271144 ,   13634349 ,   98432020 ,   13630932 ,   12720229 ,   13630246 ,   13631237 ,   13636506 ,   98431301 ,   99270338 ,   13631878 ,   13631722 ,   12720664 ,   13633194 ,   13634767 ,   13630981 ,   13632075 ,   13632827 ,   12721251 ,   13631414 ,   12720360 ,   13636204 ,   13631913 ,   13631649 ,   13636117 ,   13630754 ,   13632010 ,   13632082 ,   13630654 ,   13630105 ,   13635797 ,   13633511 ,   13632152 ,   12720258 ,   13635816 ,   13631394 ,   13631768 ,   13632102 ,   12720190 ,   13631223 ,   12720447 ,   13631514 ,   13631264 ,   12720077 ,   98430891 ,   13636532 ,   13632206 ,   13630984 ,   13630836 ,   13630982 ,   13634411 ,   12720184 ,   13631558 ,   13632834 ,   13630149 ,   12721223 ,   13635488 ,   12720722 ,   12720054 ,   12720993 ,   13634208 ,   99749360 ,   99749252 ,   12721054 ,   98432072 ,   98432224 ,   13630211 ,   13631169 ,   13631999 ,   13631927 ,   13631997 ,   13631846 ,   13636102 ,   98432261 ,   13631447 ,   13631990 ,   13632025 ,   13631865 ,   13631879 ,   13630504 ,   13630530 ,   12720603 ,   12720078 ,   13630394 ,   13632708 ,   13630923 ,   12721040 ,   13632034 ,   13634552 ,   13633626 ,   13630336 ,   13633601 ,   13631609 ,   13632051 ,   13631472 ,   13635381 ,   13636809 ,   13633594 ,   13634073 ,   12721141 ,   12720171 ,   13634261 ,   13630069 ,   13636466 ,   13630687 ,   13633552 ,   13635752 ,   13632076 ,   13633677 ,   13634154 ,   13631956 ,   13630256 ,   13631832 ,   12720602 ,   13635544 ,   13631302 ,   13630617 ,   13635445 ,   99270871 ,   13630913 ,   13631682 ,   12720169 ,   13633951 ,   13630784 ,   13635442 ,   98432167 ,   13631481 ,   13630623 ,   13636329 ,   12720512 ,   13630104 ,   13631282 ,   12721197 ,   13631352 ,   13630467 ,   12721084 ,   13631740 ,   13630141 ,   13631015 ,   12720433 ,   13632089 ,   13631820 ,   13632125 ,   13630027 ,   13631713 ,   13631573 ,   13630719 ,   13634370 ,   13632588 ,   13635534 ,   98432097 ,   98431056 ,   13631502 ,   12720322 ,   13631626 ,   13631586 ,   13632346 ,   13633570 ,   13635869 ,   98431297 ,   13631417 ,   13632221 ,   13632191 ,   13631743 ,   99271127 ,   13636802 ,   13632240 ,   99271084 ,   99271080 ,   99271091 ,   13633633 ,   13631519 ,   12720983 ,   13631462 ,   12721222 ,   13637206 ,   13631584 ,   13631599 ,   13637326 ,   13635800 ,   12720737 ,   13632138 ,   12721081 ,   13630248 ,   13634917 ,   98431260 ,   13635706 ,   13635632 ,   13635758 ,   13635625 ,   98431180 ,   13636299 ,   13630372 ,   13631426 ,   13631046 ,   13630525 ,   13632172 ,   12721203 ,   13630352 ,   13630783 ,   98430673 ,   98432184 ,   99270486 ,   13637032 ,   13633958 ,   13631541 ,   13630954 ,   13630861 ,   13630910 ,   13635285 ,   12720140 ,   12720095 ,   13630125 ,   13631824 ,   13635764 ,   13637272 ,   13630710 ,   13631041 ,   13635374 ,   13631232 ,   12720238 ,   13635849 ,   13635497 ,   13631001 ,   13634214 ,   13635298 ,   98430636 ,   98432199 ,   13636202 ,   13630951 ,   13633647 ,   13636059 ,   13631164 ,   13631316 ,   13634405 ,   13635449 ,   13630662 ,   12720282 ,   98432098 ,   12720191 ,   13632081 ,   13632241 ,   13632033 ,   13631697 ,   13630207 ,   13631577 ,   12720729 ,   13630933 ,   13631413 ,   13631695 ,   13633951 ,   13631636 ,   13631639 ,   13631200 ,   99270518 ,   13636502 ,   12721143 ,   13635681 ,   13631668 ,   13633939 ,   13635410 ,   13635310 ,   13631202 ,   13631091 ,   13631664 ,   13634133 ,   13633945 ,   13631209 ,   13631843 ,   13636310 ,   12720684 ,   13633649 ,   13630316 ,   13633554 ,   13631785 ,   13633506 ,   13631939 ,   13632442 ,   13637155 ,   13637071 ,   13630059 ,   12721268 ,   13631837 ,   12720429 ,   12720089 ,   13637161 ,   13631978 ,   12720498 ,   13630442 ,   13630758 ,   13634834 ,   13637159 ,   12720732 ,   12720686 ,   12720361 ,   98432237 ,   13635855 ,   99270662 ,   13631709 ,   99270378 ,   13637804 ,   13630886 ,   12720811 ,   13634573 ,   13631627 ,   13635538 ,   13631342 ,   13631277 ,   12720147 ,   13637920 ,   13630622 ,   12720035 ,   12720792 ,   13632048 ,   13630658 ,   13634350 ,   13634383 ,   13630970 ,   12720957 ,   13633943 ,   13630445 ,   13631566 ,   13630300 ,   13631775 ,   13630568 ,   98432001 ,   13631653 ,   13634517 ,   13631029 ,   13631581 ,   13631547 ,   13631645 ,   98432230 ,   99270457 ,   12720495 ,   13633705 ,   13631925 ,   13632162 ,   98432060 ,   13635437 ,   13631753 ,   13632471 ,   13631930 ,   13630665 ,   13632020 ,   13636016 ,   13632756 ,   13636055 ,   13637283 ,   98432022 ,   13631090 ,   13635714 ,   13634665 ,   13633603 ,   98431225 ,   13631411 ,   12721165 ,   13631147 ,   13635599 ,   13635628 ,   13630862 ,   13631371 ,   13634399 ,   13631561 ,   13630752 ,   98431152 ,   13635244 ,   13630215 ,   13630680 ,   13631826 ,   13634599 ,   13630813 ,   13630660 ,   99270977 ,   99749419 ,   98432181 ,   98430870 ,   98430641 ,   99749061 ,   98432253 ,   12721004 ,   13636526 ,   12720545 ,   98432176 ,   98431800 ,   98431071 ,   13630940 ,   13636291 ,   13636294 ,   13635440 ,   13635392 ,   13635425 ,   13631284 ,   13631522 ,   12720131 ,   13630405 ,   13630140 ,   13630193 ,   13631706 ,   13631098 ,   13630853 ,   13634081 ,   13630400 ,   13635380 ,   12720240 ,   12720004 ,   13633122 ,   13632063 ,   13631944 ,   13635441 ,   13631953 ,   13635930 ,   13635899 ,   12720311 ,   13631252 ,   99270442 ,   12721155 ,   13630991 ,   13631225 ,   13631025 ,   13631336 ,   13635314 ,   13631286 ,   12720756 ,   13635760 ,   13635740 ,   13636449 ,   13630553 ,   12720234 ,   13630285 ,   98432277 ,   13630908 ,   98431602 ,   99749546 ,   12720286 ,   13630028 ,   13630788 ,   98431214 ,   13635317 ,   99270001 ,   98411617 ,   12720214 ,   13630889 ,   12721248 ,   13634166 ,   13630054 ,   13633133 ,   13630878 ,   13630603 ,   13634331 ,   13636986 ,   13632005 ,   13631534 ,   13631555 ,   13636642 ,   13630953 ,   13630635 ,   13635929 ,   98431264 ,   13634506 ,   13630233 ,   13636050 ,   98432188 ,   12721070 ,   12721017 ,   98432209 ,   13631508 ,   13631635 ,   12721009 ,   13631479 ,   12720872 ,   13631528 ,   13631681 ,   13631287 ,   13631488 ,   13631456 ,   12720188 ,   13630638 ,   12721144 ,   13635498 ,   13635645 ,   12721193 ,   13636194 ,   98431290 ,   98432018 ,   13633982 ,   13631248 ,   98431217 ,   98431285 ,   13633496 ,   13636153 ,   13631084 ,   13630111 ,   12720566 ,   13631390 ,   13635766 ,   13637137 ,   12721001 ,   12721126 ,   13637245 ,   98432265 ,   13635687 ,   13635982 ,   13630251 ,   13630406 ,   13630766 ,   13633638 ,   12721264 ,   13631313 ,   13634528 ,   13631194 ,   13630129 ,   12720567 ,   12720525 ,   13631303 ,   13630943 ,   13632047 ,   12720192 ,   13633151 ,   13631629 ,   13637687 ,   12720247 ,   13630561 ,   13633713 ,   13637590 ,   99271055 ,   13636081 ,   13637481 ,   13635840 ,   13631816 ,   12720723 ,   13635145 ,   13635796 ,   13632113 ,   13636750 ,   13636845 ,   13633562 ,   13630203 ,   13631594 ,   13630367 ,   13635399 ,   13631520 ,   13632038 ,   13630877 ,   12720869 ,   13633589 ,   12720301 ,   13630522 ,   13637685 ,   99270325 ,   13633640 ,   13637106 ,   13634010 ,   13633998 ,   13630914 ,   13632060 ,   13636083 ,   13635793 ,   13631780 ,   13636290 ,   99270179 ,   13631557 ,   13637209 ,   13633650 ,   12720018 ,   13633590 ,   12720023 ,   13631752 ,   13633659 ,   13633701 ,   13633673 ,   13633616 ,   13630628 ,   13635829 ,   98431097 ,   98432210 ,   13633543 ,   12721247 ,   13630074 ,   13630044 ,   13630363 ,   13632123 ,   12720404 ,   13630358 ,   13631568 ,   13631517 ,   13634132 ,   13634289 ,   12720705 ,   13634890 ,   13631623 ,   13632214 ,   13637870 ,   12720295 ,   13633711 ,   13631688 ,   13632217 ,   13635359 ,   13635306 ,   13633586 ,   13633669 ,   98432251 ,   13633195 ,   13630254 ,   13637133 ,   13631080 ,   13637010 ,   13633995 ,   13630685 ,   12720164 ,   98432228 ,   99271057 ,   12720591 ,   13630068 ,   13637234 ,   13637099 ,   13631889 ,   12720176 ,   13631694 ,   13636008 ,   13632054 ,   13633718 ,   13636818 ,   13635267 ,   13634343 ,   13634813 ,   13635333 ,   13635332 ,   13635344 ,   13635237 ,   13634024 ,   12720003 ,   12720303 ,   12720408 ,   13630803 ,   12721096 ,   13630619 ,   13635662 ,   13632216 ,   13630435 ,   13631773 ,   13630216 ,   13635698 ,   13636330 ,   13631660 ,   13630184 ,   13631921 ,   13631957 ,   12720547 ,   12720719 ,   13637039 ,   13635405 ,   13631060 ,   13630308 ,   13631942 ,   13631869 ,   13631949 ,   13632170 ,   13637635 ,   13630436 ,   13630725 ,   13635388 ,   13635245 ,   13635266 ,   13635205 ,   13635877 ,   13632198 ,   13631772 ,   13636331 ,   13634365 ,   13630271 ,   12720029 ,   13633628 ,   13633732 ,   13630209 ,   13630494 ,   12720581 ,   13635120 ,   13631081 ,   13636038 ,   13631552 ,   13636017 ,   13635127 ,   13633148 ,   13631905 ,   13633734 ,   98431530 ,   13632019 ,   13634369 ,   12720779 ,   13634315 ,   13631443 ,   13631650 ,   12720771 ,   99270915 ,   13631027 ,   12720138 ,   13630830 ,   13633981 ,   13634221 ,   13636691 ,   13630449 ,   13635253 ,   13632052 ,   13631621 ,   12721033 ,   98431313 ,   13631218 ,   13636275 ,   13637185 ,   13633188 ,   99270646 ,   99749420 ,   98431714 ,   13632006 ,   13636341 ,   13636316 ,   13635416 ,   13637448 ,   13634060 ,   13630713 ,   13634263 ,   13634389 ,   98431512 ,   13630866 ,   12720890 ,   13630967 ,   13634689 ,   13630228 ,   13630850 ,   13630050 ,   13634265 ,   13630883 ,   13636048 ,   12720891 ,   13635667 ,   12720278 ,   13632735 ,   13631337 ,   13634789 ,   13630226 ,   13632594 ,   12721166 ,   99271109 ,   12721006 ,   98431749 ,   12721131 ,   13635148 ,   13632110 ,   13636112 ,   13637732 ,   12720986 ,   13630870 ,   13632757 ,   13636295 ,   13635483 ,   13630032 ,   13635732 ,   13635772 ,   98431050 ,   12720414 ,   13630904 ,   99749534 ,   13634327 ,   13634181 ,   13630696 ,   13630274 ,   13631174 ,   13631618 ,   13631445 ,   13635162 ,   13631322 ,   13633545 ,   13631764 ,   13634376 ,   13630828 ,   13630702 ,   13630440 ,   13637639 ,   13632091 ,   13635436 ,   13635358 ,   13634183 ,   13634210 ,   13634334 ,   13635742 ,   13635761 ,   13633513 ,   13636719 ,   13637129 ,   13637204 ,   13631895 ,   13632184 ,   13635415 ,   13631979 ,   12720345 ,   13636753 ,   13630297 ,   13635478 ,   13636688 ,   13635624 ,   13631859 ,   13635593 ,   12720083 ,   13635407 ,   12721094 ,   99749327 ,   99749558 ,   99749563 ,   13635547 ,   13636189 ,   13636485 ,   13634283 ,   13630358 ,   13635791 ,   13631415 ,   13631485 ,   13631428 ,   13634684 ,   12720587 ,   12721275 ,   12720700 ,   13634381 ,   13630785 ,   13637140 ,   13632136 ,   13631158 ,   99270929 ,   13637036 ,   13635499 ,   13635734 ,   13634021 ,   13637050 ,   13632055 ,   12720077 ,   13630094 ,   13636148 ,   13631389 ,   13634929 ,   13633694 ,   13637152 ,   13636057 ,   13633707 ,   13630941 ,   13635832 ,   12720687 ,   13631593 ,   13631518 ,   13635475 ,   13635458 ,   13635397 ,   13635493 ,   12720254 ,   13630067 ,   12720246 ,   13630832 ,   13630857 ,   12720552 ,   99271105 ,   13636806 ,   13635519 ,   99749234 ,   99749317 ,   13633639 ,   13632200 ,   13630439 ,   13635456 ,   13634169 ,   13633726 ,   13636296 ,   13631585 ,   13631442 ,   13637509 ,   13637491 ,   13632105 ,   13632145 ,   12720682 ,   13632753 ,   13633631 ,   13637527 ,   13631915 ,   99271018 ,   13634558 ,   13634486 ,   13637892 ,   13634330 ,   13637115 ,   13636103 ,   13632249 ,   13632023 ,   13631459 ,   13637917 ,   13632785 ,   13635844 ,   13635831 ,   13634585 ,   13637006 ,   13630113 ,   13632134 ,   13635466 ,   13636833 ,   13636646 ,   13635455 ,   13637795 ,   13635513 ,   13631000 ,   13637158 ,   13636282 ,   12720271 ,   13631733 ,   13631590 ,   13635418 ,   13634733 ,   13637052 ,   13635859 ,   13636786 ,   13633679 ,   13632100 ,   13632150 ,   12720619 ,   13635559 ,   13636167 ,   13637312 ,   13634042 ,   12720306 ,   13630521 ,   12721170 ,   13632159 ,   13636093 ,   13631173 ,   13637128 ,   13634271 ,   13633975 ,   13635972 ,   13637278 ,   13633190 ,   13636582 ,   13630029 ,   13634630 ,   13636255 ,   12720755 ,   13634850 ,   13636080 ,   12720502 ,   13636839 ,   13631146 ,   13631589 ,   13636859 ,   13636205 ,   13634001 ,   12720548 ,   12720775 ,   13635426 ,   13635512 ,   13635506 ,   13632158 ,   13631022 ,   12720593 ,   13631992 ,   12720344 ,   13631467 ,   13631600 ,   13636229 ,   13635421 ,   13635432 ,   13635549 ,   13634641 ,   13632111 ,   12720557 ,   13635412 ,   13636381 ,   13632114 ,   13636663 ,   13636163 ,   13637758 ,   13630779 ,   13635481 ,   13632124 ,   13633671 ,   13634025 ,   13637402 ,   13630952 ,   13637067 ,   13632824 ,   13634520 ,   13631928 ,   13635642 ,   13633985 ,   13636710 ,   13634409 ,   13636619 ,   13636601 ,   13637120 ,   13633964 ,   13637351 ,   13631297 ,   13634148 ,   13630122 ,   13631233 ,   13630930 ,   12720159 ,   13633668 ,   13637840 ,   13636452 ,   13634815 ,   99270960 ,   13631220 ,   13631206 ,   13637648 ,   13637626 ,   13636268 ,   13634360 ,   13630162 ,   13630214 ,   13636828 ,   13636669 ,   13637012 ,   13637005 ,   12721204 ,   13631506 ,   13630812 ,   13631933 ,   13632202 ,   13631405 ,   13635790 ,   12720734 ,   13637486 ,   13637432 ,   13637560 ,   13636467 ,   13636280 ,   13637405 ,   13631103 ,   13632210 ,   13630674 ,   13637910 ,   13637727 ,   13634097 ,   13633128 ,   13635833 ,   13635406 ,   13631872 ,   12720577 ,   13631016 ,   12721130 ,   13630591 ,   12721028 ,   13636643 ,   99749572 ,   13637403 ,   12720487 ,   13631132 ,   13630500 ,   13635339 ,   13634362 ,   12720439 ,   13636593 ,   13635434 ,   12720908 ,   13633629 ,   13631611 ,   13635783 ,   13635716 ,   13632544 ,   13634515 ,   13630048 ,   13631061 ,   13637077 ,   13631802 ,   13633519 ,   13633523 ,   13633662 ,   13631030 ,   13631746 ,   12720966 ,   12720721 ,   13633961 ,   13634091 ,   13633942 ,   13631932 ,   13630741 ,   13637943 ,   13634119 ,   13636793 ,   13636607 ,   13633731 ,   12720136 ,   13632690 ,   13634912 ,   12721146 ,   13633714 ,   13630303 ,   12721138 ,   13635598 ,   13630060 ,   13634581 ,   13631940 ,   13630700 ,   12720346 ,   12720456 ,   13636455 ,   13633598 ,   13631188 ,   13636996 ,   12720428 ,   13637737 ,   13631203 ,   13634361 ,   99749567 ,   13632230 ,   13636286 ,   13636247 ,   13636248 ,   13634300 ,   13634028 ,   12720980 ,   13636795 ,   13637136 ,   13633704 ,   13630399 ,   13630544 ,   13630341 ,   13632795 ,   13636686 ,   13636762 ,   13637589 ,   13637015 ,   13637146 ,   12720767 ,   12721117 ,   12720129 ,   13633182 ,   13635940 ,   12720417 ,   13637708 ,   13637802 ,   12720390 ,   13634363 ,   13631149 ,   13636221 ,   13636645 ,   13633952 ,   13634079 ,   13630295 ,   13635922 ,   13632149 ,   13636907 ,   13631727 ,   13635617 ,   12720531 ,   13637821 ,   13636546 ,   13637702 ,   13632011 ,   13634182 ,   13631065 ,   13636061 ,   13635735 ,   13636567 ,   13630572 ,   13635684 ,   12721111 ,   13634382 ,   13636063 ,   13636342 ,   13636141 ,   13636332 ,   13636121 ,   99270916 ,   13637616 ,   13631935 ,   12720657 ,   13636924 ,   13633547 ,   13637258 ,   13632802 ,   13637273 ,   12720688 ,   13637915 ,   13634820 ,   13633646 ,   13637546 ,   13630487 ,   13636219 ,   13631703 ,   98410659 ,   13631747"
    }
  ]
}