  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78702",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "1067 US customers  163 Canada, 917 International",
      "recall_number": "Z-0348-2018",
      "product_description": "ECG Out Cable  -  Heartstart MRx  M1783A/M5526A 12-pin Sync cables  All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641.  Manufactured - September 4, 2002   October 8, 2014",
      "product_quantity": "15811",
      "reason_for_recall": "Issue with ECG out cables.  When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs):    \"\tOn the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave.  \"\tOn the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*.     *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning:  When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+.  If the user follows this warning, this problem cannot occur on the XL+.",
      "recall_initiation_date": "20170717",
      "center_classification_date": "20180114",
      "termination_date": "20180712",
      "report_date": "20180124",
      "code_info": "M1783A/M5526A 12-pin Sync cables    All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641"
    }
  ]
}