  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78636",
      "recalling_firm": "Vyaire Medical",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0346-2018",
      "product_description": "Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290.    The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.",
      "product_quantity": "16,670 units in total",
      "reason_for_recall": "Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber.  The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.",
      "recall_initiation_date": "20141129",
      "center_classification_date": "20180124",
      "termination_date": "20210205",
      "report_date": "20180131",
      "code_info": "0001100127  0001116871  0001116872  0001124936  0001129065  0001129066  0001129067  0001129069  0001129070  0001131426  0001131428  0001134350  0001135400  0001136027  0001136028  0001136029  0001145151  0001145152  0001145153  0001157125"
    }
  ]
}