      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78636",
      "recalling_firm": "Vyaire Medical",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0345-2018",
      "product_description": "Adult Air-Life single-limb circuit (Model AH202) with AirLife Humidification Chamber, Model Number AH290.    The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.",
      "product_quantity": "16,670 units in total",
      "reason_for_recall": "Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber.  The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.",
      "recall_initiation_date": "20141129",
      "center_classification_date": "20180124",
      "termination_date": "20210205",
      "report_date": "20180131",
      "code_info": "0001115270  0001121993  0001124929  0001 124930  0001 124932  0001129063  0001 129064  0001131424  0001131425  0001 134347  0001134348  0001135401  0001135402  0001137252  0001137253  0001139266  0001139267  0001145147  0001145148  0001145149  0001156843"
    }
  ]
}