  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Shirley",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91178",
      "recalling_firm": "Mobius Imaging, LLC",
      "address_1": "2 Shaker Rd Ste F100",
      "address_2": "",
      "postal_code": "01464-2535",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to consignees in Colorado, Tennessee, and Utah.",
      "recall_number": "Z-0344-2023",
      "product_description": "MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1",
      "product_quantity": "3",
      "reason_for_recall": "Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.",
      "recall_initiation_date": "20221118",
      "center_classification_date": "20221130",
      "termination_date": "20250226",
      "report_date": "20221207",
      "code_info": "UDI-DI 00869346000200  Serial Numbers AIRO-0124, AIRO-0145, AIRO-0288",
      "more_code_info": ""
    }
  ]
}