      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78636",
      "recalling_firm": "Vyaire Medical",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0344-2018",
      "product_description": "Infant AirLife Dual-,limb, dual-heat, high-flow circuit (Model AH265) with AirLife Humidification Chamber, Model Number AH290.    The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.",
      "product_quantity": "16,670 units in total",
      "reason_for_recall": "Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber.  The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.",
      "recall_initiation_date": "20141129",
      "center_classification_date": "20180124",
      "termination_date": "20210205",
      "report_date": "20180131",
      "code_info": "0001115635  0001115640  0001124941  0001124942  0001124943  0001124944  0001129628  0001134758  0001134759"
    }
  ]
}