      "country": "United States",
      "city": "Portage",
      "address_1": "3800 E Centre Ave",
      "reason_for_recall": "It was identified by a customer complaint that some I.V. poles were  retracting/lowering from an extended position to the low height unexpectedly. An  investigation was initiated which identified that the I.V. Latch Housing was  nonconforming.  potential hazards:  \" Suspended I.V. pole retracts unexpectedly  \" Suspended I.V. pole retracts unexpectedly with pump/bags attached",
      "address_2": "",
      "product_quantity": "19,876",
      "code_info": "Stryker 2-Stage and 3-Stage I.V. Poles distributed on beds, stretchers, and in service  kits. This product field action is limited to I.V. Poles distributed between  01/07/2014 - 06/24/2014.    The following I.V. Pole option numbers come affixed to beds and stretchers:    Manufacturer Model Number and Product Description  2035-112-010; Head End Left, 2 Stage I.V. Pole  2035-113-011; Head End Right, 2 Stage I.V. Pole  0785-035-101; 26\", 2 Stage IV Pole Assembly  0785-035-200; 26\", 3 Stage IV Pole Assembly  0785-035-300; 30\", 3 Stage IV Pole Assembly  0785-035-401; 30\", 2 Stage IV Pole Assembly      The following I.V. Pole option numbers ship as service kits.    Manufacturer Service Kit Number and Product Description  1105-035-368; 30\" IV POLE 2 STAGE HE PR  1105-035-367; 30\" IV POLE 2 STAGE HE PL  1105-035-343; 30\" IV POLE 2 STAGE FE PR  1105-035-341; 30\" IV POLE, 2 STAGE HE, PR  1105-035-340; 30\" IV POLE, 2 STAGE FE, PL  1105-035-338; 30\" IV POLE, 2 STAGE FE, PL  1040-035-338; 2-STAGE IV POLEHEAD LT30\" ZM  0785-035-340; 2-STAGE IV POLE,HEAD LT, 30\"  0785-035-341; 2-STAGE IV POLEHEAD RT 30\"  0785-035-343; 2-STAGE IV POLE,FOOT RT, 30\"  1105-035-344; 30\" IV POLE, 3 STAGE FE, PL  1105-035-342; 30\" IV POLE, 3 STAGE HE, PL  1105-035-339; 30\" IV POLE, 3 STAGE FE, PR  1105-035-369; 30\" IV POLE, 3 STAGE HE, PL  1105-035-366; 30\" IV POLE, 3 STAGE HE, PR  0785-035-339; 3-STAGE IV POLE,FOOT RT, 30\"  0785-035-342; 3-STAGE IV POLE,HEAD LT, 30\"  1105-035-644; 26\" IV POLE, 3 STAGE FL  1105-035-642; 26\" IV POLE, 3 STAGE HL  1105-035-639; 26\" IV POLE 3 STAGE FR  1105-035-637; 26\" IV POLE, 3 STAGE HR  0785-035-637; 3-STAGE IV POLE,HEAD RT, 26\"  0785-035-639; 3-STAGE IV POLE,FOOT RT, 26\"  0785-035-642; 3-STAGE IV POLE,HEAD LT, 26\"  1105-035-643; 26\" IV POLE, 2 STAGE FR  1105-035-641; 26\" IV POLE, 2 STAGE HR  1105-035-640; 26\" IV POLE, 2 STAGE FL  1105-035-638; 26\" IV POLE, 2 STAGE HL  0785-035-638; 2-STAGE IV POLE,HEAD LT, 26\"  0785-035-640; 2-STAGE IV POLE,FOOT LT, 26\"  0785-035-641; 2-STAGE IV POLE,HEAD RT, 26\"  0785-035-643; 2-STAGE IV POLE,FOOT RT, 26\"  2035-112-000; IV POLE - 2 STAGE H/E  2035-113-000; DUAL H/E 2 STAGE IV POLE OPT",
      "center_classification_date": "20151130",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including DC and Guam, and  the countries of New Zealand Canada Chile Hong Kong Colombia Argentina Brazil France Germany Italy Netherlands Spain Switzerland United Kingdom India Japan South Korea Mexico South Africa Poland Singapore.",
      "state": "MI",
      "product_description": "Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher.  The I.V. poles are stationary, height adjustable stands intended to hold infusion liquids, infusion accessories, and/or other medical devices. They may be used on the following \u001cparent\u001d devices: Model 2030 (EPIC II), model 3005 S3 beds and stretcher.  models 1105, 1105E, 1105X, 1105XE, 1115, 1115E, 1115X, 1115XE, 1125E, 1125X  and 1125XE.",
      "report_date": "20151209",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Medical Division of Stryker Corporation",
      "recall_number": "Z-0344-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72466",
      "termination_date": "20160906",
      "more_code_info": "",
      "recall_initiation_date": "20151013",
      "postal_code": "49002-5826",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}