  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88987",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide.",
      "recall_number": "Z-0342-2022",
      "product_description": "Puritan Bennett 980 Series Ventilator, 980U1ENASAA",
      "product_quantity": "278 total",
      "reason_for_recall": "Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.",
      "recall_initiation_date": "20211104",
      "center_classification_date": "20211209",
      "termination_date": "20240709",
      "report_date": "20211215",
      "code_info": "GTIN/UDI - 10884521201385  Serial Numbers: 35B1700252 35B1700365 35B1700399 35B1700407 35B1700489 35B1700505 35B1700513 35B1700555 35B1700557 35B1700562 35B1700563 35B1700565 35B1700566 35B1700605 35B1700612 35B1700634 35B1700651 35B1700662 35B1700663",
      "more_code_info": ""
    }
  ]
}