  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88987",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide.",
      "recall_number": "Z-0340-2022",
      "product_description": "Puritan Bennett 980 Series Ventilator, 980S1ENDICU",
      "product_quantity": "278 total",
      "reason_for_recall": "Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.",
      "recall_initiation_date": "20211104",
      "center_classification_date": "20211209",
      "termination_date": "20240709",
      "report_date": "20211215",
      "code_info": "GTIN/UDI - 10884521171114  Serial Numbers: 35B1700409 35B1700443 35B1700457 35B1700462 35B1700473 35B1700476 35B1700483 35B1700487 35B1700490 35B1700511 35B1700552 35B1700567 35B1700569 35B1700570",
      "more_code_info": ""
    }
  ]
}