{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ball Ground",
      "state": "GA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90999",
      "recalling_firm": "Caire, Inc.",
      "address_1": "2200 Airport Industrial Dr Ste 500",
      "address_2": "N/A",
      "postal_code": "30107-4686",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide - Worldwide Distribution:  CA, NC, OH, AZ, and Canada, Chile, Colombia, and Germany",
      "recall_number": "Z-0338-2023",
      "product_description": "CAIRE Liberator 30, MODEL 13337403, Liquid Oxygen System Unit",
      "product_quantity": "5 devices",
      "reason_for_recall": "An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.",
      "recall_initiation_date": "20221010",
      "center_classification_date": "20221202",
      "termination_date": "20240819",
      "report_date": "20221214",
      "code_info": "UDI/DI M766133374030, Serial Numbers:  CBB3022300368, CBB3022300390, CBB3022300470, CBB3022300472, CBB3022300479"
    }
  ]
}