{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88983",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide.",
      "recall_number": "Z-0336-2022",
      "product_description": "Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.",
      "product_quantity": "5,566 (US); 6,354 (OUS)",
      "reason_for_recall": "Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.",
      "recall_initiation_date": "20211115",
      "center_classification_date": "20211215",
      "report_date": "20211222",
      "code_info": "All Serial Numbers.   May also be designated as part number D998 etc.  Model Number, UDI-DI: 0998-00-0800-31, 10607567109053; 0998-UC-0800-31, 10607567109053; 0998-00-0800-32, 10607567111117; 0998-UC-0800-32, 10607567111117; 0998-00-0800-33, 10607567109008; 0998-UC-0800-33, 10607567109008; 0998-00-0800-34, 10607567111940; 0998-UC-0800-34, 10607567111940; 0998-00-0800-35, 10607567109107; 0998-UC-0800-35, 10607567109107; 0998-00-0800-36, 10607567114187; 0998-UC-0800-36, 10607567114187; 0998-00-0800-45, 10607567108421; 0998-UC-0800-45, 10607567108421; 0998-00-0800-52, 10607567108438; 0998-UC-0800-52, 10607567108438; 0998-00-0800-53, 10607567108391; 0998-UC-0800-53, 10607567108391; 0998-00-0800-55, 10607567108414; 0998-UC-0800-55, 10607567108414; 0998-00-0800-65, 10607567113432; 0998-UC-0800-65, 10607567113432",
      "more_code_info": ""
    }
  ]
}