  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Rd",
      "reason_for_recall": "Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a batch of product code 631508 was incorrectly labeled.",
      "address_2": "",
      "product_quantity": "9",
      "code_info": "Catalog Number 631508  Serial Numbers: 0001765391, 0001765392, 0001765393, 0001765394, 0001765395, 0001765396, 0001765397, 0001765398, 0001765399",
      "center_classification_date": "20131118",
      "distribution_pattern": "US Distribution in the states of: NC and CA.",
      "state": "MI",
      "product_description": "Vascutek Gelsoft Vascular Graft (peripheral)  Labeling reads in part, \"***Gelsoft Gelatin Impregnated Knitted Vascular Prosthesis***STRAIGHT***STERILE***EO***Item No: 631508***Diameter 8mm***Usable Length 15cm***Manufactured By***VASCUTEK Ltd. a TERUMO Company Newmains Avenue Inchinnan, Renfrewshire PA4 9RR Scotland***\"    The Vascutek Gelsoft Vascular Prosthesis is a gelatin-sealed, woven polyester graft indicated for abdominal and peripheral vascular repair, i.e. replacement or bypass in aneurysmal and occlusive disease of arteries",
      "report_date": "20131127",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-0336-2014",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "66670",
      "termination_date": "20140227",
      "more_code_info": "",
      "recall_initiation_date": "20131023",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}