  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "If the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with the user due to the length of the accessory.",
      "address_2": "",
      "product_quantity": "9 units",
      "code_info": "Model #728306, Serial #'s: 100016, 100018, 100019, 100020, 100021, 100022, 100023, 100024 and 100025.",
      "center_classification_date": "20131115",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) Distribution in the states of DC, GA, IL, IN, NY and OH and the countries: Germany, Israel and Netherlands.",
      "state": "OH",
      "product_description": "Brilliance iCT Computed Tomography X-Ray System,       Product Usage:  The Brilliance iCT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.",
      "report_date": "20131127",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-0335-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66601",
      "termination_date": "20150804",
      "more_code_info": "",
      "recall_initiation_date": "20131011",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}