  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86521",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0334-2021",
      "product_description": "ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 51 mm    Item Number: 00-8065-548-26",
      "product_quantity": "412 units total",
      "reason_for_recall": "Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant.  Cannot rule out the potential for adverse tissue reactions.",
      "recall_initiation_date": "20200924",
      "center_classification_date": "20201102",
      "termination_date": "20220520",
      "report_date": "20201111",
      "code_info": "All lots"
    }
  ]
}