      "status": "Terminated",
      "city": "Dublin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66635",
      "recalling_firm": "Carl Zeiss Meditec, Inc.",
      "address_1": "5160 Hacienda Dr",
      "address_2": "",
      "postal_code": "94568-7562",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution.",
      "recall_number": "Z-0334-2014",
      "product_description": "Zeiss brand IOLMaster 500, Model 500, Biomicroscope, Slit-Lamp, AC-Powered,     Product is manufactured by Carl Zeiss Meditec AG (Jena),  Carl Zeiss Promenade 10, Jena, Germany 07745; and distributed by Carl Zeiss Meditec, Inc., Dublin, CA    This device is intended for use primarily by physicians and health care workers and may be only used under the supervision of a physician. This device will not be sold to the general public. The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-to white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation",
      "product_quantity": "1,201 units",
      "reason_for_recall": "The user database delivered for download contained an erroneous value. Specifically the \"aO\" constant of the lens AMO Sensar AR40E is listed as \"0.0\" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wrong diopter value.",
      "recall_initiation_date": "20131023",
      "center_classification_date": "20131115",
      "termination_date": "20140417",
      "report_date": "20131127",
      "code_info": "Serial Numbers:  817123  1038929  1038930  1044483  1044485  1044487  1044488  1044489  1044490  1045639  1046055  1046056  1046057  1046343  1046344  1046345  1046353  1046355  1046364  1046373  1046374  1046376  1046377  1046378  1046380  1046381  1046488  1046489  1046500  1046509  1046514  1046516  1046523  1047089  1047090  1047091  1047092  1047093  1047101  1047107  1047111  1047266  1047267  1047268  1047269  1047270  1047271  1047272  1047273  1047275  1047279  1047282  1047283  1047284  1047485  1047486  1047494  1047498  1047499  1047507  1047512  1047513  1047518  1047519  1047520  1047521  1047522  1047523  1047524  1047914  1047921  1047922  1047923  1047924  1047933  1047934  1047935  1047936  1047938  1048579  1048580  1048581  1048582  1048583  1048584  1048585  1048587  1048589  1049601  1049602  1049603  1049604  1049605  1049607  1049608  1049609  1049610  1049611  1049620  1049621  1049622  1050560  1050561  1050562  1050563  1050564  1050566  1050567  1050568  1050569  1050748  1050749  1050751  1050752  1050753  1050754  1050755  1050756  1050757  1050758  1050759  1050760  1050761  1050762  1050763  1051518  1051519  1051520  1051521  1051522  1051523  1051524  1051525  1051526  1051527  1051528  1051529  1051530  1051531  1051533  1051534  1051536  1051538  1051550  1051551  1051552  1051553  1051554  1051860  1051861  1051872  1051873  1051874  1051875  1051876  1051877  1051879  1052512  1052513  1052515  1052516  1052517  1052518  1052519  1052520  1052527  1052528  1052529  1052917  1052918  1052919  1052920  1052944  1052945  1052946  1052947  1052948  1052949  1052951  1052952  1052955  1053181  1053183  1053186  1053187  1053188  1053189  1053190  1053191  1053193  1053194  1053206  1053207  1053208  1053209  1053210  1053211  1053212  1053213  1053554  1053555  1053556  1053558  1053561  1053562  1053563  1053564  1053566  1053567  1053568  1053571  1053573  1053574  1053841  1053847  1054330  1054332  1054338  1054339  1054341  1054342  1054343  1054344  1054347  1054351  1054360  1054362  1054363  1054365  1054366  1054367  1054369  1055236  1055244  1055245  1055246  1055247  1055248  1055249  1055251  1055263  1055266  1055267  1055268  1055949  1055950  1055951  1055952  1055953  1055954  1056018  1056019  1056020  1056021  1056023  1056025  1056026  1056027  1056028  1056451  1056452  1056453  1056454  1056455  1056456  1056457  1057114  1057126  1057127  1057128  1057130  1057131  1057163  1057164  1057165  1057166  1057168  1057177  1057178  1057179  1057181  1057182  1057183  1057188  1057445  1057450  1057451  1057453  1057455  1057460  1058039  1058041  1058042  1058043  1058044  1058045  1058046  1058047  1058050  1058051  1058052  1058585  1058592  1058606  1058607  1058608  1058609  1058610  1058611  1058612  1058613  1058614  1058615  1058618  1058619  1058620  1058621  1058624  1058625  1058627  1058628  1058629  1058649  1058650  1058651  1058652  1058661  1058662  1058663  1059225  1059226  1059227  1059228  1059229  1059230  1059233  1059235  1059236  1059237  1059238  1059869  1059870  1059874  1059875  1059876  1059877  1059878  1059880  1059881  1059885  1059888  1060586  1060591  1060592  1060595  1060596  1060597  1060598  1060599  1060602  1060604  1060605  1060606  1060608  1060609  1060610  1060611  1060612  1060613  1060614  1060615  1060616  1060620  1061181  1061182  1061183  1061184  1061185  1061187  1061188  1061189  1061190  1061191  1061192  1061196  1061202  1061203  1061204  1061205  1061206  1061207  1061208  1061209  1061211  1061212  1061214  1061215  1061217  1061219  1062545  1062546  1062547  1062548  1062549  1062550  1062551  1062553  1062554  1062555  1062556  1062557  1062558  1062559  1062560  1062783  1063470  1063471  1063472  1063473  1063474  1063475  1063476  1063479  1063480  1063481  1063499  1063500  1063501  1063503  1064061  1064070  1064071  1064072  1064073  1064074  1064075  1064076  1064077  1064078  1064086  1064087  1064089  1064090  1064091  1064092  1064093  1064097  1064098  1064329  1064332  1064333  1064334  1064335  1064336  1064337  1064338  1064341  1064945  1064948  1064949  1064953  1064954  1064955  1064956  1064959  1064960  1064961  1064962  1064963  1064964  1064965  1064966  1064967  1064968  1064969  1064970  1064971  1064972  1064973  1065400  1065414  1065415  1065416  1065417  1065418  1065419  1065420  1065421  1065422  1065423  1065424  1065425  1065426  1065427  1065533  1065534  1065535  1065536  1065537  1065538  1065539  1065540  1065541  1065577  1065579  1065580  1065581  1065582  1065583  1065584  1065585  1065594  1065635  1065636  1065637  1065638  1065639  1065640  1065641  1065642  1065643  1065653  1065654  1065655  1065656  1065658  1065659  1065660  1065661  1065665  1065666  1065667  1065668  1065669  1065670  1065671  1065673  1065674  1066443  1066444  1066445  1066446  1066447  1066448  1066450  1066451  1066452  1066675  1067061  1067062  1067064  1067065  1067066  1067067  1067068  1067069  1067636  1067637  1067638  1067639  1067640  1067641  1069993  1069994  1069995  1069996  1069997  1069998  1069999  1070000  1070001  1070002  1070003  1070013  1070154  1070155  1070156  1070169  1070170  1070171  1070172  1070174  1070175  1070176  1070177  1070178  1070204  1070205  1070207  1070208  1070209  1070759  1070762  1070772  1070773  1070774  1071210  1071211  1071212  1071213  1071214  1071590  1071591  1071592  1071594  1071595  1071596  1071597  1071617  1071618  1071619  1071621  1071622  1071624  1071714  1071715  1071716  1071717  1071718  1071719  1071720  1071721  1071726  1071736  1071737  1071738  1071739  1071740  1071741  1071742  1071743  1071745  1071746  1071747  1071748  1071749  1071750  1071952  1071963  1071965  1071966  1071967  1071973  1071974  1071975  1072831  1072832  1072833  1072834  1072836  1072837  1072838  1072839  1072840  1072991  1072992  1072993  1072994  1072995  1072996  1072997  1072998  1072999  1073000  1073005  1073006  1073007  1073008  1073009  1073010  1073011  1073012  1073013  1074561  1074562  1074563  1074564  1074565  1074566  1074568  1074569  1074570  1074571  1074934  1074935  1074938  1074939  1074980  1074981  1074982  1074983  1075269  1075270  1075271  1075272  1075273  1075276  1075277  1075529  1075531  1075532  1075533  1075534  1075536  1075537  1075538  1075561  1075562  1075563  1075564  1075565  1075566  1075568  1075569  1075570  1075576  1075792  1075793  1075794  1075795  1075796  1075797  1075798  1075799  1075800  1075801  1075804  1076541  1076542  1076543  1076544  1076545  1076546  1076547  1076548  1076550  1076553  1077011  1077012  1077014  1077015  1077016  1077017  1077018  1078156  1078158  1078160  1078162  1078163  1078164  1078165  1078173  1078651  1078652  1078655  1078656  1080020  1080021  1080023  1080024  1080026  1080033  1080034  1080165  1080166  1080168  1080169  1080170  1080171  1080172  1080173  1080174  1080182  1080423  1080425  1080443  1080444  1080445  1080447  1080448  1080449  1080450  1080452  1080457  1080459  1080460  1080464  1080655  1080658  1080659  1080660  1080663  1080666  1080668  1080669  1080874  1080877  1080880  1080881  1080887  1080888  1080890  1080892  1080893  1080894  1080895  1080896  1080899  1080901  1080902  1080903  1080904  1081148  1081153  1081154  1081155  1081184  1081185  1081188  1081191  1082370  1082372  1082373  1082376  1082377  1082378  1082379  1082384  1082389  1082395  1082416  1082417  1082418  1082419  1082420  1082421  1082422  1082425  1082429  1082430  1082433  1083469  1083470  1083471  1083472  1083473  1083474  1083475  1083476  1083480  1083481  1083484  1083485  1083487  1083489  1083494  1083495  1083501  1083503  1083505  1083506  1083508  1083511  1083512  1083517  1084044  1086647  1086651  1086653  1086654  1086655  1086656  1086657  1086658  1086659  1086661  1086662  1086663  1086664  1086667  1086672  1086676  1086680  1086684  1086690  1086692  1086806  1086809  1086811  1086813  1086815  1086816  1086817  1086818  1086819  1086820  1086821  1086846  1086848  1086850  1088220  1088221  1088225  1088292  1088296  1088297  1088299  1088321  1088322  1088324  1089272  1089273  1089276  1089277  1089278  1089279  1089280  1089281  1089285  1090583  1090585  1090586  1090589  1090593  1090594  1090595  1090596  1092632  1092633  1092634  1092635  1092636  1092638  1092639  1092640  1092683  1092686  1092688  1092690  1092692  1092702  1094030  1094032  1094033  1094034  1094035  1094039  1094044  1094046  1094048  1094049  1094050  1094051  1094053  1094054  1094056  1094057  1094058  1094060  1094061  1094062  1094063  1094064  1094065  1094066  1094067  1094068  1094069  1094071  1094072  1094074  1094075  1094076  1095354  1095356  1095357  1095358  1095360  1095361  1095362  1095364  1095365  1095366  1095375  1095378  1095389  1095390  1095392  1095822  1095823  1095824  1095826  1095827  1095833  1095834  1095835  1095836  1095837  1095838  1095839  1095841  1097061  1097063  1097066  1097068  1097069  1097074  1097075  1097077  1097078  1097079  1097081  1097084  1097088  1097090  1097091  1097092  1097096  1097660  1097661  1097662  1097663  1097664  1097665  1097666  1097678  1097684  1097685  1097686  1097687  1097689  1097691  1097694  1097699  1098281  1098282  1098285  1098286  1098287  1098294  1098295  1098299  1098303  1098305  1098311  1098795  1098797  1098798  1098801  1098802  1098804  1099560  1099568  1099569  1099571  1099572  1099575  1099588  1099592  1099594  1100687  1100688  1100689  1100690  1100720  1100723  1100724  1100725  1100726  1100728  1100729  1101675  1101704  1101706  1101711  1101713  1101714  1101715  1101716  1101717  1101723  1101724  1102075  1102078  1102080  1102081  1102082  1102083  1102084  1102085  1102086  1102087  1102092  1102093  1102095  1102099  1103003  1103005  1103006  1103007  1103008  1103011  1103019  1103020  1103021  1103022  1103024  1103025  1103035  1103036  1103037  1103038  1103039  1103040  1103041  1104998  1105001  1105002  1105003  1105004  1105005  1105007  1105603  1105604  1105605  1105606  1105607  1105608  1105609  1105610  1105611  1105612  1105626  1105627  1105630  1105634  1105635  1105895  1105896  1105898  1105900  1105901  1105902  1105903  1105904",
      "more_code_info": ""
    }
  ]
}