{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78879",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "107 domestic accounts  15 US Govt accounts  682 Foreign Accounts  3 Mexico accounts  12 Canada accounts",
      "recall_number": "Z-0333-2018",
      "product_description": "Philips Healthcare IQon Spectral CT computed tomography x-ray system",
      "product_quantity": "50",
      "reason_for_recall": "There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs,  the part(s) will not be expelled, however, a noise may be heard coming from the gantry.",
      "recall_initiation_date": "20171108",
      "center_classification_date": "20180111",
      "termination_date": "20200417",
      "report_date": "20180117",
      "code_info": "IQon Spectral CT 728332    Devices manufactured 03/05/2009 to present.",
      "more_code_info": ""
    }
  ]
}