  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89024",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Brazil",
      "recall_number": "Z-0332-2022",
      "product_description": "Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA",
      "product_quantity": "2 units OUS",
      "reason_for_recall": "Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit",
      "recall_initiation_date": "20211112",
      "center_classification_date": "20211203",
      "report_date": "20211215",
      "code_info": "Serial Numbers: MCSB006939, MCSB006947 GTIN: 10884521592896",
      "more_code_info": ""
    }
  ]
}