      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78879",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "107 domestic accounts  15 US Govt accounts  682 Foreign Accounts  3 Mexico accounts  12 Canada accounts",
      "recall_number": "Z-0332-2018",
      "product_description": "Philips Healthcare Brilliance iCT SP computed tomography x-ray system",
      "product_quantity": "134",
      "reason_for_recall": "There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs,  the part(s) will not be expelled, however, a noise may be heard coming from the gantry.",
      "recall_initiation_date": "20171108",
      "center_classification_date": "20180111",
      "termination_date": "20200417",
      "report_date": "20180117",
      "code_info": "Brilliance iCT SP - 728311    Devices manufactured 03/05/2009 to present.",
      "more_code_info": ""
    }
  ]
}