  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88999",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.",
      "recall_number": "Z-0328-2022",
      "product_description": "RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum  REF: CH200",
      "product_quantity": "15 kits",
      "reason_for_recall": "When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range",
      "recall_initiation_date": "20211101",
      "center_classification_date": "20211201",
      "termination_date": "20230309",
      "report_date": "20211208",
      "code_info": "Lot Numbers/Exp Date: 559344         28th June 2022; 572730         28th July 2023. GTIN: 05055273201130",
      "more_code_info": ""
    }
  ]
}