  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78727",
      "recalling_firm": "Encore Medical, Lp",
      "address_1": "9800 Metric Blvd",
      "address_2": "N/A",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "One medical device distributor in California.",
      "recall_number": "Z-0326-2018",
      "product_description": "DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE,   REF 801-05-042, NON STERILE, Qty 1  For use in orthopedic surgery.",
      "product_quantity": "2 units",
      "reason_for_recall": "The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.",
      "recall_initiation_date": "20171212",
      "center_classification_date": "20180110",
      "termination_date": "20220613",
      "report_date": "20180117",
      "code_info": "Lot 252567",
      "more_code_info": ""
    }
  ]
}