      "country": "United States",
      "city": "Batesville",
      "address_1": "1069 State Road 46 E",
      "reason_for_recall": "Ceiling Bracket 71 was  shipped with a larger center hole. A potentially  hazardous situation exists that if the Ceiling Bracket 71  with a larger center hole is used for a concrete ceiling  Ultra System installation with the M8 screw and the safe  working load is exceeded, the system could fall.",
      "address_2": "",
      "product_quantity": "188",
      "code_info": "manufacturing date between Dec 01, 2015- Apr 14, 2016",
      "center_classification_date": "20161031",
      "distribution_pattern": "WA  NC  OH  PA  MI  Australia  Sweden",
      "state": "IN",
      "product_description": "Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling  installed rail systems: Multirall 200, Likorall 200, 242, 243, 250,  and LikoGuard L and XL.",
      "report_date": "20161109",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hill-Rom, Inc.",
      "recall_number": "Z-0326-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75197",
      "termination_date": "20170324",
      "more_code_info": "",
      "recall_initiation_date": "20160809",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}