  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69127",
      "recalling_firm": "Philips Medical Systems, Inc.",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and to the countries of:  Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Denmark Estonia, Ethiopia. France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Italy,  Japan, Jordan, Korea, Republic of, Libya, Malaysia, Mexico, Nepal, Netherlands,  New Zealand, Norway, Oman, Palestinian territory, Peru, Poland, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates and United Kingdom.",
      "recall_number": "Z-0325-2016",
      "product_description": "Ingenia, Intera, Achieva and Achieve dStream MR systems on Software  versions R5.1.7 and R5.1.8 with the Mobi View software option.  Diagnostic imaging system.",
      "product_quantity": "307 units",
      "reason_for_recall": "When a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed.  This occurs with software release R5.1.7/R5.1.8.",
      "recall_initiation_date": "20140821",
      "center_classification_date": "20151123",
      "termination_date": "20210512",
      "report_date": "20151202",
      "code_info": "Software  versions R5.1.7 and R5.1.8"
    }
  ]
}