  "results": [
    {
      "status": "Terminated",
      "city": "Laval",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72529",
      "recalling_firm": "Pega Medical Inc.",
      "address_1": "1111 Autoroute Chomedy",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in the states of FL and TN.",
      "recall_number": "Z-0324-2016",
      "product_description": "SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System.     This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number.      The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients.   Orthopedic.",
      "product_quantity": "4 units distributed in the US",
      "reason_for_recall": "SCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.",
      "recall_initiation_date": "20151020",
      "center_classification_date": "20151120",
      "termination_date": "20160816",
      "report_date": "20151202",
      "code_info": "Part Number: SCF-MLD265; LOT#: 130703-01  Part Number: SCF-MLD273; LOT#: 130703-02"
    }
  ]
}