      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78833",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "412 units affected.",
      "recall_number": "Z-0311-2018",
      "product_description": "Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.",
      "product_quantity": "412",
      "reason_for_recall": "If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.",
      "recall_initiation_date": "20170714",
      "center_classification_date": "20180103",
      "termination_date": "20200605",
      "report_date": "20180110",
      "code_info": "9890-010-88152 - Codes included in attached list. Devices were manufactured and distributed from March 2015 - March 2017."
    }
  ]
}