      "country": "United States",
      "city": "Irvine",
      "address_1": "9775 Toledo Way",
      "reason_for_recall": "Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.",
      "address_2": "",
      "product_quantity": "66,429 units",
      "code_info": "9946730  9960999  9965968  9974493  A004327  A011297  A016502  A024591  9952156  9961016  9965970  9975117  A004329  A011515  A016504  A024594  9952519  9961297  9965980  9975118  A004435  A011522  A016507  A024893  9952892  9961308  9965983  9976018  A004436  A011731  A016510  A024894  9953171  9961313  9965984  9976024  A004437  A011735  A016632  A024896  9953177  9963245  9966022  9976049  A004535  A011818  A017149  A025130  9953699  9963247  9966200  9976052  A004537  A011845  A017150  A025857  9953892  9963249  9966203  9976055  A004542  A011972  A017154  A025862  9955091  9963378  9966205  9978598  A004543  A012088  A018239  A026331  9955406  9963380  9966550  A000215  A004544  A012341  A018301  A026336  9955408  9963382  9966562  A000219  A004614  A012591  A018307  A026337  9955530  9963386  9966563  A000220  A004616  A012691  A018310  A026375  9955561  9963622  9966674  A000369  A005866  A012738  A018721  A027449  9955726  9963627  9966674  A000453  A005870  A012743  A018722  A027805  9955727  9963633  9966676  A002304  A005980  A012755  A018741  A027862  9955845  9963851  9966934  A002305  A006045  A013058  A018885  A027922  9956028  9964145  9966945  A002307  A006222  A013211  A019293  A028081  9956396  9964148  9966946  A002404  A006225  A013685  A019401  A028468  9956460  9964157  9971999  A002406  A006229  A013686  A019539  A028566  9957242  9964161  9972000  A002408  A006237  A013688  A019550  A028760  9957512  9964259  9972002  A002595  A006829  A013689  A019567  A028777  9957684  9964494  9972610  A002606  A006831  A013796  A019573  A028780  9957685  9964496  9972613  A002616  A006834  A013969  A019834  A028787  9957927  9964498  9972719  A002898  A007148  A014018  A019912  A029143  9957947  9964756  9972720  A003029  A007150  A014061  A020026  A029145  9957949  9964810  9973291  A003038  A007304  A014063  A020056  A029146  9958187  9964826  9973322  A003039  A007344  A014351  A020244  A029318  9958345  9965055  9973329  A003044  A007345  A014352  A020313  A029460  9958346  9965062  9973567  A003164  A007349  A014359  A020315  A029508  9958393  9965067  9973570  A003291  A007354  A014450  A020457  A029610  9958740  9965070  9973597  A003296  A007462  A014753  A020464  A029613  9958741  9965483  9973779  A003311  A007463  A014768  A020564  A029617  9958750  9965488  9973880  A003316  A008617  A014795  A020569  A029794  9959038  9965489  9974008  A003603  A008767  A014840  A020699  A029862  9959490  9965768  9974013  A003692  A008886  A015227  A021233  A029863  9959502  9965771  9974068  A003703  A009203  A015228  A021384  A029865  9959520  9965805  9974070  A003706  A009221  A015332  A021386  A029866  9959884  9965946  9974071  A003710  A010851  A015362  A021416  A031039  9959885  9965947  9974165  A003805  A010947  A015541  A021885  A031102  9960141  9965950  9974226  A004047  A010948  A015559  A021887  A031106  9960147  9965956  9974230  A004085  A010956  A015563  A021913  A031385  9960150  9965958  9974247  A004116  A010990  A015570  A022134  A031387  9960720  9965962  9974250  A004259  A011059  A015957  A022404  A031391  9960761  9965963  9974426  A004318  A011267  A016217  A022406  A031649  9960776  9965965  9974433  A004321  A011295  A016477  A024199  A031663  A031670  A040153  A049233  A058253  A067173  A075715  A081223  A091333  A031677  A040155  A049236  A059305  A067407  A075721  A081718  A091341  A031963  A040163  A049237  A059307  A067409  A076320  A081725  A091589  A032065  A040168  A050083  A059311  A067412  A076402  A081859  A091708  A032068  A040705  A050084  A059577  A068091  A076406  A081893  A091710  A032070  A040718  A050119  A059691  A068094  A076410  A081895  A091713  A032072  A040725  A050263  A059727  A068098  A076412  A082393  A093603  A032398  A041039  A050356  A059731  A068222  A077046  A083144  A093607  A032835  A041157  A050361  A059733  A068382  A077048  A083166  A093787  A032837  A041166  A050399  A061328  A068419  A077049  A083203  A093789  A032840  A041283  A050404  A061329  A068436  A077557  A083206  A093795  A033206  A041359  A050744  A062218  A068563  A077565  A083545  A093797  A033229  A041404  A050745  A062266  A068565  A077607  A083546  A094057  A033230  A041456  A050756  A062267  A068773  A077650  A085047  A094083  A033233  A041488  A050855  A062268  A068835  A077664  A085059  A094085  A034330  A041495  A050863  A062269  A068839  A078031  A085063  A094100  A034351  A041754  A051742  A062987  A068842  A078033  A085066  A094794  A034352  A041861  A051875  A062991  A068850  A078870  A085070  A094796  A034353  A041866  A051939  A063168  A068966  A078879  A085073  A094801  A034354  A041867  A051940  A063235  A068981  A078950  A085329  A095119  A034509  A042017  A051941  A063275  A072687  A078962  A086510  A095305  A034511  A043949  A051943  A063399  A072735  A078963  A086658  A095308  A034512  A043960  A051958  A063400  A072738  A079311  A087223  A095397  A034517  A043964  A052051  A063960  A072740  A079320  A087228  A096868  A034568  A046465  A052195  A063968  A072810  A079321  A087231  A096899  A034585  A046470  A052217  A063976  A073361  A079324  A087234  A097061  A034586  A046485  A052227  A064156  A073364  A079334  A088753  A097090  A034587  A046487  A052237  A064238  A073371  A079456  A088756  A097095  A037007  A046488  A052246  A064250  A073382  A079606  A089280  A097098  A037008  A046610  A052542  A064251  A073384  A080017  A089296  A097100  A037009  A046611  A052828  A064252  A074651  A080018  A089334  A097427  A037010  A046735  A052873  A064254  A075001  A080019  A089673  A097429  A037014  A046739  A052874  A064373  A075030  A080021  A089675  A097430  A037410  A046868  A052876  A064908  A075034  A080022  A089837  A097433  A037415  A046873  A053105  A064910  A075040  A080241  A090423  A097435  A037416  A047032  A053180  A064912  A075046  A080269  A090577  A097436  A037513  A048479  A053183  A064999  A075177  A080272  A090598  A097437  A038097  A048480  A053452  A065056  A075183  A080276  A090712  A098264  A038103  A048487  A053459  A065597  A075298  A080387  A091142  A100171  A038785  A048645  A058121  A065598  A075361  A080592  A091143  A100172  A038794  A048646  A058149  A065602  A075362  A080600  A091145  A100240  A038802  A048808  A058152  A066531  A075366  A080613  A091147  A100243  A039450  A048815  A058154  A066948  A075619  A080712  A091224  A100251  A039461  A048821  A058157  A066965  A075697  A081143  A091312  A100584  A040138  A048825  A058160  A066966  A075709  A081213  A091331  A100588  A100593  A117561  A123429  A135805  A142934  A156901  A163572  A179306  A100895  A117826  A123434  A136017  A142937  A156904  A164739  A181904  A101575  A117907  A123805  A136086  A143067  A156909  A164745  A181919  A101628  A117969  A123807  A136448  A143074  A157053  A164825  A181921  A101634  A117971  A124180  A136451  A144556  A157135  A164903  A181952  A101638  A118192  A124276  A136456  A144585  A157322  A164981  A182269  A101641  A118425  A124277  A136580  A144591  A157325  A164989  A182274  A101941  A118687  A124464  A136583  A144597  A157367  A165291  A182284  A101944  A118689  A124476  A137622  A144736  A157895  A165294  A182289  A101950  A118708  A125090  A137641  A144873  A157896  A165297  A183791  A102234  A118842  A126669  A137643  A145017  A157898  A165563  A183801  A102251  A118850  A126670  A137645  A145023  A158504  A165564  A183802  A102287  A118854  A127819  A137757  A145031  A158512  A165565  A183928  A102289  A119044  A127937  A137803  A145039  A158557  A165932  A184118  A102293  A119046  A127949  A137873  A145165  A159131  A165935  A184221  A102646  A119144  A128031  A137879  A153798  A159132  A165946  A184222  A102685  A119346  A128212  A137885  A153890  A159133  A165985  A184357  A102686  A119879  A128333  A137886  A153892  A159134  A165986  A185869  A103326  A119891  A128942  A137888  A153894  A159337  A172223  A185871  A103395  A120131  A128953  A138208  A153896  A159338  A172229  A185872  A103398  A120204  A129090  A138219  A154486  A159813  A172232  A185875  A103401  A120240  A129091  A138220  A154498  A159830  A172236  A186101  A103406  A120283  A129327  A138225  A154505  A159844  A174414  A186135  A103410  A120400  A129335  A138231  A154514  A159848  A174416  A186140  A103412  A120413  A129898  A138490  A154522  A159851  A175520  A186333  A104092  A121145  A130244  A138906  A154633  A160096  A175556  A186481  A104114  A121151  A130945  A138908  A154634  A161031  A175695  A189181  A104120  A121155  A130985  A138921  A154813  A161039  A175710  A189190  A105499  A121156  A131044  A138923  A154816  A161042  A175715  A189192  A105549  A121772  A131047  A138926  A155028  A161051  A175935  A189194  A105553  A121779  A131939  A139599  A155131  A161142  A176084  A189815  A105757  A121785  A131943  A139607  A155133  A161279  A176157  A189816  A105880  A122401  A131966  A139610  A155140  A161288  A176158  A189818  A105883  A122468  A132855  A139615  A155147  A161289  A176216  A189825  A105887  A122621  A133983  A139617  A155157  A161292  A178717  A190048  A105895  A122651  A134515  A139993  A155434  A161463  A178718  A190117  A105913  A122655  A134519  A140160  A155668  A162951  A178726  A190118  A106016  A122656  A134630  A140235  A156090  A162954  A178883  A190283  A106026  A123032  A135227  A140561  A156286  A163149  A178912  A190446  A116756  A123037  A135229  A140671  A156380  A163157  A178918  A190453  A116759  A123039  A135232  A142410  A156402  A163168  A178921  A190455  A117473  A123202  A135380  A142419  A156409  A163353  A179058  A190456  A117485  A123205  A135786  A142422  A156415  A163499  A179077  A191757  A117494  A123211  A135789  A142552  A156419  A163513  A179079  A191759  A117516  A123427  A135800  A142556  A156421  A163526  A179304  A191767  A191864  A213290  A228170  A235536  A246111  A253655  A264501  A275676  A200836  A214210  A228172  A235538  A246120  A253663  A264962  A275677  A200884  A214214  A228208  A236147  A246128  A253664  A264965  A275678  A201108  A214215  A228603  A236466  A246131  A255799  A264967  A275679  A201216  A214558  A228604  A237340  A246255  A255814  A264971  A276176  A201220  A214561  A228859  A237344  A246256  A255821  A265128  A276177  A201224  A214564  A228862  A237349  A246570  A255823  A265129  A276284  A201331  A214898  A230195  A237353  A246571  A255827  A265691  A276285  A201490  A214905  A230269  A237466  A246633  A255985  A265699  A277023  A201510  A214908  A230847  A237467  A246634  A255986  A265706  A277024  A201532  A214909  A230930  A237561  A247155  A256542  A265773  A277025  A201542  A214965  A230931  A237564  A247267  A256548  A265775  A277172  A201907  A215131  A230984  A237642  A247287  A256550  A265873  A278439  A203970  A215149  A230985  A237645  A248003  A256555  A267430  A278441  A203975  A215156  A230987  A237647  A248009  A256742  A267432  A278443  A203985  A215164  A230991  A238041  A248020  A256819  A267525  A278627  A203990  A215167  A231005  A238047  A248023  A256822  A267534  A279421  A204483  A215538  A231643  A239537  A248028  A256823  A267535  A279423  A204486  A215602  A231645  A239539  A248215  A256826  A267536  A279439  A204493  A215651  A231649  A239540  A248743  A256827  A267537  A279729  A204871  A215652  A231652  A239550  A248746  A257352  A268139  A279730  A204875  A222349  A231656  A239551  A248753  A257358  A268176  A280700  A204876  A222570  A232346  A240033  A248759  A257360  A268243  A280706  A205105  A222647  A232352  A240038  A248894  A257363  A268246  A280709  A209299  A222648  A232359  A240043  A248896  A257368  A268247  A280855  A209303  A223013  A232363  A240047  A249006  A258088  A268253  A280856  A209304  A223015  A232365  A240159  A249008  A258096  A268254  A280931  A209330  A223124  A232587  A240161  A249010  A258107  A268742  A281003  A209430  A223125  A232588  A243270  A249406  A258255  A272372  A281667  A209707  A223150  A232598  A243277  A249648  A260041  A272384  A282592  A209710  A224545  A233765  A243282  A249650  A260047  A272390  A282622  A209714  A224546  A233766  A243284  A249714  A260055  A272537  A282625  A209717  A224547  A233767  A243415  A249715  A260216  A272612  A282647  A210619  A225374  A233768  A243417  A249953  A260768  A272614  A282650  A210621  A225445  A234758  A243529  A250226  A260776  A272624  A282821  A210623  A225446  A234759  A243532  A250259  A260782  A274619  A282834  A210624  A225449  A234760  A243533  A250825  A260786  A274627  A282840  A210897  A225654  A234761  A243535  A250851  A260946  A274630  A282841  A210906  A227264  A235062  A243994  A250852  A262262  A274641  A282843  A210918  A227265  A235063  A243997  A250856  A262287  A274813  A283310  A212857  A227684  A235210  A244001  A251155  A262288  A275504  A283606  A212865  A227685  A235215  A244012  A253403  A262289  A275505  A283641  A213284  A227714  A235220  A244099  A253411  A262382  A275506  A283642  A213286  A227717  A235221  A244101  A253414  A262419  A275507  A283643  A213288  A227794  A235532  A244139  A253427  A264500  A275666  A283644  A284103  A290773  A297537  A303119  A309234  A316750  A323580  A329235  A284108  A290791  A297612  A303136  A309394  A316751  A323582  A329356  A284109  A291365  A297616  A303137  A309522  A316752  A324273  A329358  A284113  A291405  A297618  A303138  A310667  A317330  A324274  A329359  A284119  A291406  A299056  A303140  A310668  A317394  A325147  A329761  A284907  A291407  A299057  A303142  A310803  A317412  A325150  A329765  A284920  A291567  A299059  A303388  A310809  A317437  A325360  A329891  A284927  A292912  A299329  A303393  A312420  A318015  A325361  A329892  A284934  A292918  A299330  A303411  A312421  A318019  A325772  A329935  A284939  A292921  A299331  A303412  A312461  A318187  A325773  A330004  A284944  A293023  A299593  A303413  A314037  A320445  A325949  A330008  A284948  A293329  A300051  A303688  A314041  A320448  A326557  A330010  A286589  A293332  A300060  A303840  A314042  A321077  A326564  A330011  A286591  A293333  A300237  A303853  A314043  A321079  A326565  A335114  A286649  A293540  A300319  A303859  A314044  A321202  A326723  A335138  A288843  A293541  A300569  A305247  A314477  A321291  A326787  A335146  A288845  A293856  A300570  A305283  A314479  A321294  A326805  A335151  A288854  A293858  A300595  A305285  A314492  A321295  A326806  A335171  A288946  A293861  A300982  A305287  A315198  A321297  A326808  A335537  A288960  A295763  A300994  A305291  A315204  A322497  A326809  A335543  A288961  A296040  A300995  A305950  A315207  A322498  A327070  A336022  A288965  A296042  A300996  A305960  A315210  A322504  A328301  A336359  A288966  A296043  A301001  A305962  A315214  A322669  A328310  A336361  A290567  A296134  A301556  A305966  A316607  A323235  A328319  A336692  A290575  A296135  A301558  A305969  A316610  A323237  A328442  A336700  A290582  A296701  A301560  A309144  A316612  A323241  A328474  A336702  A290599  A297361  A301563  A309230  A316746  A323579  A329194  A336704  9945407  9945399  9946910  9946737  9948666  9949668  9949448  9948915  9945725  9946271  9945991  9948660  9963219  A006885  A028706  A046108  A092226  A125138  A197490  A217465  9972750  A006886  A032436  A046114  A092708  A125142  A197771  A218093  9972981  A017413  A032439  A049550  A092709  A125735  A198001  A218094  9972987  A017429  A033725  A049557  A092735  A125819  A198024  A218095  9973009  A017430  A033841  A060565  A092741  A125820  A198044  A221276  9976672  A017432  A033856  A060578  A092742  A125943  A198049  A221279  9976687  A020842  A033858  A072386  A119501  A126381  A205358  A221280  9976691  A020853  A033861  A072396  A119534  A126394  A205362  A221281  9977163  A020921  A034550  A081121  A119537  A126396  A211498  A229624  9977489  A021063  A036785  A081129  A119547  A126424  A211509  A229671  A001785  A021473  A036788  A081140  A119552  A127015  A211511  A265947  A001914  A025088  A036790  A091993  A119802  A127019  A217418  A265948  A001919  A025090  A044657  A091996  A120128  A127211  A217448  A266807  A001959  A025123  A044663  A091999  A120130  A153128  A217449  A266811  A002231  A025129  A044668  A092004  A124985  A197485  A217460  A280246  A003190  A028569    Additional Lots  A335134  A335148  A335154  9952403  9952144  9950754  9950731  9950392  9951451  9950007  9951453  A076257  A336031  A336033  A335161  A335378  A335379  A335404  A335541  A335542  A336029  A336036  A336354  A336688  A336706  A337379  A338056  A338065  A338068  A086973",
      "center_classification_date": "20161109",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Korea, Latvia, Lithuania, Malaysia, Mexico, Nepal, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Russian Federation, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.",
      "state": "CA",
      "product_description": "Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter  Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled selective infusion of physician-specified therapeutics agents such as embolization materials and of diagnostic materials such as contrast media. UltraFlow HPC Flow: The device is intended for the controlled selective infusion of physician-specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuroanatomy. It is not intended for use in the coronary vasculature.    Catalog No.  105-5055V02  105-5055V02  105-5055V05  105-5065V02  105-5065V05  105-5066V02  105-5066V05",
      "report_date": "20161116",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Micro Therapeutics Inc, Dba Ev3 Neurovascular",
      "recall_number": "Z-0309-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75358",
      "termination_date": "20171011",
      "more_code_info": "",
      "recall_initiation_date": "20161005",
      "postal_code": "92618-1811",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}