  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irvine",
      "address_1": "9775 Toledo Way",
      "reason_for_recall": "Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.",
      "address_2": "",
      "product_quantity": "11,860 units",
      "code_info": "A010308  A020911  A026317  A032375  A036898  A043417  A046146  A053154  A010309  A020912  A026319  A032378  A036899  A043663  A047160  A053878  A010310  A020913  A026835  A032381  A036900  A043667  A047161  A053881  A010312  A020914  A026836  A032383  A036901  A043668  A047162  A053882  A010314  A020916  A026842  A033000  A036904  A043911  A047808  A053993  A010316  A021047  A026843  A033001  A036905  A043912  A047809  A053994  A013030  A021410  A026939  A033237  A036906  A043913  A047810  A054421  A013031  A021411  A026940  A033238  A037896  A043914  A047811  A054422  A013591  A021412  A026941  A033240  A037897  A043915  A048395  A054423  A013592  A021479  A026942  A033242  A038406  A044125  A048396  A054424  A013594  A021483  A027316  A033243  A038407  A044126  A048397  A054425  A014538  A021484  A027317  A033246  A038410  A044127  A048398  A054426  A014542  A021677  A027319  A033738  A038411  A044128  A049717  A054427  A014543  A021678  A028325  A033740  A039064  A044129  A049718  A054714  A015450  A021679  A028326  A033831  A039066  A044130  A049721  A054777  A015451  A021680  A028327  A033832  A039067  A044131  A049722  A054778  A015557  A021681  A028328  A033833  A039068  A044278  A049724  A054779  A016139  A021688  A028588  A033879  A039727  A044279  A049726  A054780  A016140  A021691  A028591  A033880  A040185  A044284  A049727  A055473  A016152  A022538  A028609  A033881  A040188  A044285  A049933  A055475  A016155  A022707  A029206  A033882  A040330  A044287  A049938  A055476  A016797  A022710  A029208  A034154  A040331  A044378  A049939  A055477  A016798  A023091  A029223  A034156  A040480  A044379  A049942  A056174  A016801  A023092  A029315  A034201  A041006  A044380  A049945  A056178  A016802  A023094  A029316  A034203  A041009  A044381  A049947  A056180  A017024  A023255  A029965  A034334  A041510  A044382  A049949  A056181  A017485  A023256  A029968  A034335  A041511  A044826  A050932  A056182  A017486  A023261  A030489  A034336  A041513  A044827  A050933  A056768  A017487  A023269  A030490  A034337  A041514  A044829  A050935  A056771  A017489  A024218  A030542  A036036  A041870  A044831  A050936  A056775  A017861  A024228  A030543  A036189  A041871  A044834  A050973  A056778  A017862  A024268  A030544  A036338  A042005  A044873  A050974  A056794  A018467  A024577  A030545  A036339  A042392  A044883  A050976  A057026  A018469  A024908  A031358  A036340  A042396  A045690  A051066  A057027  A018470  A024917  A031360  A036391  A042399  A045691  A051067  A058140  A019062  A024920  A031361  A036404  A042400  A045692  A051069  A058142  A019063  A024923  A031364  A036514  A042658  A045693  A052643  A058143  A019065  A025001  A031794  A036606  A043061  A045695  A052644  A058145  A019499  A025002  A031795  A036607  A043063  A045697  A052645  A058146  A019500  A025634  A031948  A036608  A043064  A045698  A052646  A058147  A019616  A025635  A031949  A036749  A043065  A046133  A052648  A058148  A020032  A025654  A031951  A036750  A043272  A046134  A052649  A058505  A020186  A025655  A031953  A036777  A043273  A046139  A053151  A058506  A020488  A026312  A032012  A036778  A043285  A046141  A053152  A058507  A020909  A026314  A032372  A036780  A043416  A046142  A053153  A058508  A058509  A069632  A081553  A098121  A113419  A125915  A134228  A139869  A058510  A069633  A081557  A098122  A113420  A125916  A134229  A139870  A059244  A070370  A081558  A098123  A113421  A125919  A134230  A139871  A059245  A070380  A086827  A098124  A113791  A125921  A134886  A139872  A059246  A070552  A086831  A100283  A113792  A125922  A134887  A140870  A059247  A070553  A086837  A100284  A113793  A125924  A134888  A140872  A061332  A070554  A086840  A100295  A113794  A125925  A134889  A140873  A061934  A070555  A087759  A100297  A113795  A125926  A134890  A140875  A061935  A073701  A087760  A100298  A113796  A125927  A134891  A140876  A062283  A073702  A087761  A100299  A115088  A127285  A135409  A140877  A062284  A073704  A088653  A100323  A115090  A127287  A135410  A140878  A062752  A073705  A088654  A100324  A115091  A127291  A135411  A142033  A062988  A073707  A088655  A100327  A115605  A127292  A135412  A142036  A062989  A073709  A088658  A100328  A115606  A128359  A135413  A142037  A063057  A073711  A088659  A106144  A115607  A128360  A135414  A142038  A063745  A074018  A088660  A106155  A117448  A128361  A135415  A142039  A063747  A074019  A090010  A106160  A117449  A128363  A135417  A142040  A064308  A074021  A090012  A106223  A117665  A128364  A135418  A142041  A064309  A074022  A090013  A106880  A117667  A129828  A135419  A142042  A064310  A074023  A090016  A107034  A117696  A129830  A137811  A142043  A064313  A074026  A090018  A107035  A117697  A129831  A137812  A143592  A064973  A074892  A090019  A107036  A117698  A131570  A137813  A143593  A064975  A074893  A091998  A107039  A117700  A131573  A137814  A143594  A065074  A074895  A092000  A107673  A117701  A131575  A137815  A143595  A065075  A074896  A092002  A107674  A117702  A131576  A137819  A148018  A065522  A076813  A092003  A107675  A119160  A131577  A137820  A148019  A065523  A076815  A092005  A107679  A119161  A131579  A137821  A148022  A065524  A076817  A092006  A107680  A119162  A131581  A137823  A148024  A065526  A076818  A092007  A107681  A119164  A131582  A138559  A148025  A065639  A076820  A093615  A107682  A120209  A132269  A138560  A148026  A065642  A076821  A093616  A108850  A120210  A132270  A138561  A148028  A065649  A076822  A093617  A108856  A120212  A132271  A138562  A148035  A065651  A076823  A093618  A108857  A120213  A132272  A138563  A148042  A065653  A078964  A093619  A108858  A120215  A132276  A139221  A148043  A066073  A078965  A094468  A108859  A122026  A133598  A139222  A148044  A066074  A080848  A094469  A108860  A122027  A133599  A139223  A148045  A066078  A080849  A094470  A110200  A122028  A133600  A139224  A150199  A066079  A080850  A094473  A110201  A122029  A133602  A139225  A150200  A066692  A080851  A096665  A110203  A122033  A133605  A139226  A150201  A066694  A080852  A096669  A110206  A122035  A133606  A139227  A150202  A066695  A080853  A096670  A110207  A123930  A133607  A139229  A150203  A066696  A080854  A096671  A110208  A123931  A134224  A139853  A150204  A069628  A080855  A096673  A110209  A123932  A134225  A139854  A150205  A069630  A081550  A096675  A110210  A123933  A134226  A139864  A150894  A069631  A081551  A098120  A113417  A123935  A134227  A139867  A150898  A150899  A162848  A185769  A225554  A261101  A297968  A317168  A015452  A150900  A162850  A185771  A225555  A261102  A297970  A317171  A017863  A150901  A162851  A185772  A226043  A261106  A297971  A317937  A019067  A150902  A162853  A185773  A226051  A261109  A297973  A317940  A019687  A151239  A162854  A188190  A227129  A263738  A300193  A318732  A019498  A151240  A162855  A188195  A229179  A265492  A300194  A319040  A019689  A151241  A164383  A190077  A232637  A266994  A300991  A319043  A019686  A151242  A164384  A190079  A232638  A266997  A300992  A319130  A019688  A151243  A164385  A190080  A232640  A266999  A300993  A320823  A020051  A151244  A165179  A190081  A234756  A267000  A302644  A320826  A020185  A151245  A165180  A190082  A238664  A267001  A302645  A320827  A020487  A151246  A166121  A193900  A238666  A269767  A302646  A322378  A020486  A151248  A166122  A193902  A238667  A269771  A304830  A322381  A020485  A152139  A166123  A193904  A239875  A269773  A304842  A322384  A020484  A152140  A167455  A193905  A239877  A269774  A304844  A327069  A020184  A152141  A167457  A193906  A241057  A269775  A304846  9961450  A020035  A152142  A169229  A197312  A242547  A271509  A304847  9961902  A020047  A152143  A169230  A197314  A242548  A272748  A306243  9961900  A020910  A153769  A169231  A199979  A242550  A272750  A306245  9961899  A021481  A153770  A171151  A199983  A242559  A272752  A306251  9962409  A021690  A153771  A171156  A204514  A242561  A274216  A306253  9962411  A021689  A153772  A171163  A204515  A245762  A274223  A306254  9962666  A022709  A153773  A171165  A204516  A245765  A276038  A306256  9962420  A022520  A154448  A171168  A211767  A245767  A276041  A307932  9962984  A022522  A154450  A171248  A211769  A245768  A278715  A307935  9963258  A022519  A154452  A174423  A214097  A245770  A278718  A307940  9963003  A024283  A154453  A174424  A214779  A245771  A281388  A307941  9963797  A024276  A154454  A174426  A214781  A249221  A281389  A309598  9963255  A023270  A156152  A174427  A214782  A249225  A281390  A309599  9963792  A024213  A156157  A174428  A214783  A250673  A283037  A309601  9962996  A021046  A157201  A174430  A214784  A250675  A283040  A309603  9964272  A029219  A157202  A174432  A215491  A252980  A283041  A311498  9963787  A029216  A157203  A177772  A215494  A252981  A283042  A311499  9963786  A033834  A157204  A177775  A216381  A252983  A287710  A311500  9964271  A044281  A157205  A180608  A217701  A253701  A287712  A311502  9964269  A044288  A157206  A180610  A219813  A253703  A289358  A311503  9965334  A044282  A157207  A180654  A219814  A256125  A289361  A313697  A005065  A044280  A157210  A180656  A220212  A256127  A291847  A313698  A005067  A043664  A157211  A183853  A220502  A256129  A294769  A313699  A008831  A043301  A160075  A183854  A220506  A256855  A294772  A313700  A008829  A044283  A160076  A183875  A221351  A258774  A294776  A313701  A008832  A043661  A160077  A183876  A221352  A258777  A294786  A316285  A010289  A047036  A160919  A183877  A221354  A258784  A296212  A316286  A010917  A047751  A160920  A183879  A223464  A258846  A296214  A316436  A013596  A049094  A160921  A185767  A223465  A258848  A296216  A316437  A014546  A047788  A047786  A157208  A230275  A239181  A033899  A036714  A043575  A059875  A049098  A162846  A230621  A250676  A033921  A036761  A043607  A059883  A047787  A167458  A300192  A263745  A034028  A036762  A043608  A059889  A047790  A169227  A300191  A294784  A034047  A037727  A043620  A059898  A049087  A167454  A256868  A307933  A034056  A037783  A043622  A059901  A047752  A167459  A296219  A024002  A034069  A037792  A043671  A059907  A047789  A167456  A252984  A025079  A034331  A037890  A044712  A059909  A049089  A177776  A253705  A025131  A034348  A037944  A044750  A061683  A049106  A177773  A256867  A025453  A034348  A037949  A047035  A064065  A049780  A177774  A249223  A025469  A034378  A037957  A047070  A064101  A049080  A177777  A268268  A025490  A034387  A037968  A047418  A064122  A043418  A180655  A287709  A025816  A034405  A037975  A047452  A064134  A049104  A185764  A241048  A027429  A034430  A037994  A047473  A064147  A049102  A185766  A249224  A027949  A034763  A038333  A047492  A064154  A049107  A185765  A261105  A027958  A034790  A038529  A047517  A065277  A049096  A185774  A271508  A027960  A034805  A038550  A048296  A065305  A049100  A198686  A287711  A027966  A034818  A034405  A048336  A065317  A050118  A188192  A230623  A028002  A034887  A039292  A048366  A065330  A049781  A188191  A263747  A028019  A035376  A039800  A050293  A065342  A049779  A188194  A249226  A028025  A035407  A039819  A050314  A065354  A049782  A193910  A278717  A028748  A035474  A039819  A050341  A071060  A049105  A199982  A289359  A028816  A035474  A039877  A050375  A071079  A050117  A193907  A289360  A028845  A035484  A039916  A050416  A071095  A086855  A215497  A297969  A029783  A035526  A040676  A050461  A071130  A096668  A204517  A263739  A029776  A035547  A040683  A051858  A071141  A131037  A214100  A263748  A030150  A035554  A040714  A051869  A071167  A131039  A215506  A234754  A030271  A035731  A040803  A051869  A071786  A131040  A215504  A256856  A030287  A035774  A040804  A052182  A071791  A132274  A216380  A271507  A031103  A036013  A040852  A052211  A072457  A132273  A218414  A239184  A031136  A036044  A040853  A052245  A074811  A133596  A216377  A241045  A031151  A036084  A040880  A052282  A074829  A133597  A216378  A250671  A031182  A036127  A040884  A052296  A074839  A137824  A218224  A263742  A031224  A036136  A040948  A052305  A074877  A148031  A217703  A238668  A031240  A036137  A040955  A053710  A074897  A151238  A218226  A239185  A031274  A036176  A040967  A053760  A067913  A152133  A216379  A250672  A032693  A036188  A040968  A056706  A076175  A152136  A215496  A263744  A033113  A036177  A041392  A056721  A076233  A152135  A218413  A230622  A033124  A036219  A041713  A056730  A076207  A152134  A220211  A250674  A033445  A036228  A041734  A056737  A076238  A152137  A227130  A261103  A033490  A036245  A041746  A057442  A076248  A156153  A226053  A271511  A033809  A036603  A041756  A057449  A076258  A156154  A226046  A230625  A033819  A036613  A041760  A057454  A078408  A156158  A229181  A241049  A033850  A036615  A043549  A057461  A078418  A156155  A228013  A253706  A033857  A036696  A043548  A057470  A078451  A157209  A228014  A265490  A033887  A036697  A043567  A057481  A079821  A080450  A089979  A090139  A095566  A097029  A099768  A104413  A110371  A083417  A089991  A090167  A095618  A097729  A100859  A105772  A110376  A084728  A090011  A090181  A095637  A099267  A103256  A112687  A110383  A084761  A090047  A090209  A095694  A099348  A103270  A112690  A110389  A084820  A090070  A090241  A095716  A099367  A104383  A112693  A110396  A084834  A090076  A092811  A095724  A099374  A104400  A105810  A112046  A087689  A090107  A095509  A097001  A099734  A104409  A110354  A112674  A089934  9947374  9956715  9961032  9969715  9972030  9974427  9977717  9936174  9947375  9956717  9965803  9969716  9972031  9974429  9977718  9938400  9947376  9956719  9966844  9969717  9972037  9974430  9979347  9939295  9947429  9957296  9966845  9969762  9972507  9974926  9979595  9939297  9948948  9958024  9966850  9971070  9973294  9974927  9979597  9940045  9948952  9958034  9966852  9971071  9973296  9976336  A000429  9940046  9952154  9958040  9966853  9971072  9973298  9976338  A003369  9941533  9952183  9958091  9967670  9971303  9973299  9976339  A003372  9941535  9953585  9958112  9967674  9971309  9974011  9976342  A010293  9941539  9954488  9958114  9967677  9971310  9974012  9977016  A010297  9941540  9954489  9959779  9969487  9971644  9974026  9977711  A010297  9944947  9955853  9960360  9969489  9971646  9974028  9977712  A010914  9944948  9956647  9960363  9969507  9971648  9974419  9977715  A011604  A011605  A011611  A012315    Additional Lots  A000433  A012314  A017636  A026690  A031793  A033182  9938393  9937699  9937696  9944949  9942856  9936172  9936899  9942855  9944946  9936171  9938396  9953241  9952967  9941531  9971651  9972525  9972509  9978562  9974928  9974929  9979349  9969504  9979593  9978561  9977716  A001480  A001481  A001482  A001483  A001484  A003371  A003456  A003458  A004018  A004019  A004021  A004572  A004574  A005069  A005576  A005578  A005579  A006535  A006536  A006537  A007195  A007197  A007785  A007786  A007788  A007789  A010924  A011567  A011607  A011613  A011615  A011629  A011662  A012039  A012066  A012190  A012232  A012282  A012318  A012320  A012322  A012675  A012707  A012767  A012885  A012954  A012989  A013050  A014279  9968894  9964705  9969438  9967981  9964672  9968318  9967991  9968317  9964671  9968320  9969440  9968895  9980770  9980772  9980771  A004578  A005066  A012292  A012383  A000132  A003455  A004022  A004573  A007198  A008186  9964670  9964704  9968891  9968841  9968844  9969444  9980775  9954287  9954294  9954267  9954308  9954266  9954320  9954288  9954306  9954255  9954256  9959777  9959780  9960366  9959806  9960820  9960821  9960824  9961030  9961033  9959809  9966139  A003373  A007199  A007200  A004577  A015144  A020247  A008184  A008830  A012064  A043162  A069444  A033379  A014270  A026679  9937700  9971073  9971641  9934763  9957293  9957298  9979351  A014261  9934764  9938394  A000434  9965804  A025799  A026695  9939294  A024009  9972033  A012002  A017643  A034305  A059913  A033211  A068601  A068619  A071153  9967675  9958037  A034933  9957297  9978563  9969761  9938395  A023983  A034997  A057476  A057480  A029761  9942854  A035425  A057283  A010926  A032608  9952964  9953580  9969491  9979343  A012106  A025250  A016069  A040916  A040923  A017043  A014289  A029760  9974027  A014271  A016123  A024023  A025484  9955854  9977719  A015155  A018446  9979592  A036782  A016412  A036740  A023987  A033495  A039951  A033505  A033502  A036112  9971653  9934765  A031974  A031624  A016091  A069419  A035720  A014284  A031282  A029786  A025166  A016210  A034342  A014247  9979598  A081437  A012865  A013028  A013029  9977714  A035769  A016168  A069193  A034903  A022010  A023998  A025779  A016137  A034019  A024008  A033202  A034423  A031818  A032503  A033982  A016428  A025487  A035649  A029732  A033362  A035787  A041395  A046089  A051861  A065238  A075536  A076263  A036743  A034773  A035560  A031623  A015159  A035449  A069253",
      "center_classification_date": "20161109",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Korea, Latvia, Lithuania, Malaysia, Mexico, Nepal, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Russian Federation, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.",
      "state": "CA",
      "product_description": "Pipeline Embolization Device (Pipeline Classic)  The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracranial aneurysms (lAs) in the internal carotid artery from the petrous to the superior hypophyseal segements.    Catalog No.     FA-71300-25V09\t  FA-71350-25V16\t  FA-71400-25V10\t  FA-71425-35V13\t  FA-71475-30V16  FA-71300-25V12\t  FA-71350-25V17\t  FA-71400-25V11\t  FA-71425-35V15\t  FA-71475-30V18  FA-71300-25V13\t  FA-71350-25V18\t  FA-71400-25V12\t  FA-71425-35V18\t  FA-71475-35V09  FA-71300-25V15\t  FA-71350-25V19\t  FA-71400-25V13\t  FA-71450-25V09\t  FA-71475-35V10  FA-71300-25V16\t  FA-71350-30V09\t  FA-71400-25V15\t  FA-71450-25V10\t  FA-71475-35V11  FA-71300-25V17\t  FA-71350-30V10\t  FA-71400-25V16\t  FA-71450-25V11\t  FA-71475-35V12  FA-71300-25V18\t  FA-71350-30V12\t  FA-71400-25V18\t  FA-71450-25V12\t  FA-71475-35V13  FA-71300-25V19\t  FA-71350-30V13\t  FA-71400-25V19\t  FA-71450-25V13\t  FA-71475-35V15  FA-71300-30V09\t  FA-71350-30V15\t  FA-71400-30V09\t  FA-71450-25V15\t  FA-71475-35V18  FA-71300-30V12\t  FA-71350-30V16\t  FA-71400-30V10\t  FA-71450-25V16\t  FA-71475-35V19  FA-71300-30V13\t  FA-71350-30V18\t  FA-71400-30V11\t  FA-71450-25V18\t  FA-71500-25V09  FA-71300-30V15\t  FA-71350-35V09\t  FA-71400-30V12\t  FA-71450-25V19\t  FA-71500-25V10  FA-71300-30V16\t  FA-71350-35V10\t  FA-71400-30V13\t  FA-71450-30V09\t  FA-71500-25V11  FA-71300-30V19\t  FA-71350-35V12\t  FA-71400-30V15\t  FA-71450-30V10\t  FA-71500-25V12  FA-71300-35V09\t  FA-71350-35V13\t  FA-71400-30V16\t  FA-71450-30V11\t  FA-71500-25V13  FA-71300-35V12\t  FA-71350-35V15\t  FA-71400-30V18\t  FA-71450-30V12\t  FA-71500-25V15  FA-71300-35V13\t  FA-71350-35V18\t  FA-71400-35V09\t  FA-71450-30V13\t  FA-71500-25V16  FA-71300-35V15\t  FA-71375-25V09\t  FA-71400-35V13\t  FA-71450-30V15\t  FA-71500-25V18  FA-71325-25V09\t  FA-71375-25V10\t  FA-71400-35V15\t  FA-71450-30V16\t  FA-71500-25V19  FA-71325-25V10\t  FA-71375-25V11\t  FA-71400-35V18\t  FA-71450-30V18\t  FA-71500-30V09  FA-71325-25V12\t  FA-71375-25V12\t  FA-71425-25V09\t  FA-71450-30V19\t  FA-71500-30V10  FA-71325-25V13\t  FA-71375-25V13\t  FA-71425-25V10\t  FA-71450-35V09\t  FA-71500-30V11  FA-71325-25V15\t  FA-71375-25V15\t  FA-71425-25V11\t  FA-71450-35V10\t  FA-71500-30V12  FA-71325-25V18\t  FA-71375-25V16\t  FA-71425-25V12\t  FA-71450-35V12\t  FA-71500-30V13  FA-71325-25V19\t  FA-71375-25V17\t  FA-71425-25V13\t  FA-71450-35V13\t  FA-71500-30V15  FA-71325-30V09\t  FA-71375-25V18\t  FA-71425-25V15\t  FA-71450-35V15\t  FA-71500-30V16  FA-71325-30V10\t  FA-71375-25V19\t  FA-71425-25V16\t  FA-71450-35V16\t  FA-71500-30V17  FA-71325-30V12\t  FA-71375-30V09\t  FA-71425-25V18\t  FA-71475-25V09\t  FA-71500-30V18  FA-71325-30V13\t  FA-71375-30V10\t  FA-71425-25V19\t  FA-71475-25V10\t  FA-71500-35V09  FA-71325-30V15\t  FA-71375-30V12\t  FA-71425-30V09\t  FA-71475-25V11\t  FA-71500-35V10  FA-71325-30V16\t  FA-71375-30V13\t  FA-71425-30V10\t  FA-71475-25V12\t  FA-71500-35V11  FA-71325-30V18\t  FA-71375-30V15\t  FA-71425-30V11\t  FA-71475-25V13\t  FA-71500-35V12  FA-71325-30V19\t  FA-71375-30V16\t  FA-71425-30V12\t  FA-71475-25V15\t  FA-71500-35V13  FA-71325-35V09\t  FA-71375-30V18\t  FA-71425-30V13\t  FA-71475-25V16\t  FA-71500-35V15  FA-71325-35V12  FA-71325-35V13  FA-71325-35V15  FA-71350-25V09  FA-71350-25V10  FA-71350-25V13  FA-71350-25V15  FA-77250-10V13  FA-77250-10V19  FA-77250-12V11  FA-77250-12V12  FA-77250-12V13  FA-77250-12V18  FA-77250-14V12  FA-77250-14V13  FA-77250-14V15  FA-77250-14V18  FA-77250-14V19  FA-77250-16V11  FA-77250-16V12  FA-77250-16V13  FA-77250-16V15  FA-77250-16V16  FA-77250-16V19  FA-77250-18V09  FA-77250-18V10  FA-77250-18V11  FA-77250-18V12  FA-77250-18V13  FA-77250-18V16  FA-77250-18V18  FA-77250-18V19  FA-77250-20V10  FA-77250-20V11  FA-71375-35V09  FA-71375-35V12  FA-71375-35V13  FA-71375-35V15  FA-71375-35V16  FA-71375-35V18  FA-71400-25V09  FA-77275-14V11  FA-77275-14V12  FA-77275-14V13  FA-77275-14V17  FA-77275-14V18  FA-77275-14V19  FA-77275-16V12  FA-77275-16V13  FA-77275-16V16  FA-77275-16V18  FA-77275-16V19  FA-772",
      "report_date": "20161116",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Micro Therapeutics Inc, Dba Ev3 Neurovascular",
      "recall_number": "Z-0307-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75358",
      "termination_date": "20171011",
      "more_code_info": "",
      "recall_initiation_date": "20161005",
      "postal_code": "92618-1811",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}