{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sandy",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88762",
      "recalling_firm": "Becton Dickinson Infusion Therapy Systems Inc.",
      "address_1": "9450 S State St",
      "address_2": "",
      "postal_code": "84070-3213",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "U.S. Nationwide Distribution: AR, AZ, MA, GA, NC, OH, WA, NY, NJ, HI, CA, MD, RI, MT, IN, PA, IL OK, OR, AK, CT, TX, ME, MO, FL, VA, LA, CO, UT    O.U.S.: None",
      "recall_number": "Z-0306-2022",
      "product_description": "BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO.      BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon\" catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, dual port adapter, vent plug and pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. A BD Q-Syte device or an end cap with protective cover is provided in the unit package.    The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury.    These devices have BD Instaflash\" Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing  during insertion. Both the stabilization platform and dual port adapter are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green).    BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with  consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.",
      "product_quantity": "155,840 devices, 80 units in each box",
      "reason_for_recall": "There is a breach in the product packaging that renders the product non-sterile.",
      "recall_initiation_date": "20210908",
      "center_classification_date": "20211124",
      "report_date": "20211201",
      "code_info": "Catalog number/Model number: 383536; Batch numbers/UDI numbers: Lot: 1193055/(17)240630(10)1193055(30)80/(01)50382903835367, 2024-06-30;   Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30  Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30  Lot: 1188953: (17)240630(10)1188953(30)80/(01)50382903835367, 2024-06-30",
      "more_code_info": ""
    }
  ]
}