{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80960",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, and WY.",
      "recall_number": "Z-0299-2019",
      "product_description": "Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.",
      "product_quantity": "142",
      "reason_for_recall": "In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.",
      "recall_initiation_date": "20180806",
      "center_classification_date": "20181030",
      "termination_date": "20200410",
      "report_date": "20181107",
      "code_info": "Serial Number 23071  22776  23623  23678  22787  23021  22848  22288  22960  23082  22501  23029  23163  23164  23518  23519  22517  23538  23097  22781  23147  22506  22634  22844  23427  23169  23529  22880  23699  22429  23640  22650  22737  22542  22176  22500  22350  22247  22347  22724  22725  22895  22833  22955  22058  23400  22847  22852  22367  22600  23650  23651  22705  22928  22609  23136  23133  23137  23556  22543  23474  22821  23517  23118  23410  22789  22208  22632  23022  23024  23132  23663  23149  22743  22851  22175  23602  23018  23019  22166  22170  22945  23476  22869  23726  23035  22431  22777  22859  22861  22824  22693  22757  22651  22171  22630  23619  22941  22957  22366  23098  22998  22427  23168  22729  23459  23625  23627  22748  23054  23063  23578  23177  22352  22353  22354  23597  22736  22618  22828  22458  22695  22644  23141  22211  23059  23079  23078  23167  23464  22184  22804  22392  22234  22235  22289  22795  23172  23173  22592  22271  22914"
    }
  ]
}