  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81007",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution",
      "recall_number": "Z-0297-2019",
      "product_description": "Multix Fusion system is a radiographic system:  Models:  Multix Fusion Analog (model number 10746665) with Tube stand 3D V (model number 7042042 or 7042059);  Multix Fusion Digital Portable (model number 10746666) with Tube stand 3D V (model number 7042042);  Multix Fusion Digital Integrated (model number 10746667) with Tube stand 3D V (model number 7042042);  Multix Fusion Digital Wireless (model number 10893300) with Tube stand 3D V (model number 7042042)",
      "product_quantity": "35 units",
      "reason_for_recall": "Steel cables inside the lifting column of the overhead tube,  may rapture without triggering the safety lock, leading  to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning",
      "recall_initiation_date": "20180803",
      "center_classification_date": "20181030",
      "termination_date": "20191126",
      "report_date": "20181107",
      "code_info": "Serial Numbers:   40279  10037  1176  40026  1220  40106  40208  40209  40109  40114  40140  40104  40126  40144  10040  10041  40027  40118  40105  40128  40200  40146  10039  40102  40101  40022  40125  40142  40127  40117  40206  1192  1185  40116  40119"
    }
  ]
}