  "results": [
    {
      "status": "Terminated",
      "city": "Blue Ash",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74701",
      "recalling_firm": "Ethicon Endo-Surgery Inc",
      "address_1": "4545 Creek Rd",
      "address_2": "N/A",
      "postal_code": "45242-2803",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.",
      "recall_number": "Z-0296-2017",
      "product_description": "Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM  The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses",
      "product_quantity": "213",
      "reason_for_recall": "Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.",
      "recall_initiation_date": "20160715",
      "center_classification_date": "20161020",
      "termination_date": "20170228",
      "report_date": "20161026",
      "code_info": "N91L2N, N91L2P, N91N1J, N91N1K, N91L0T, N91L0U, N91N29,   N91N2G, N91R8X, N91R8Y."
    }
  ]
}