      "status": "Terminated",
      "city": "Redmond",
      "state": "WA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78574",
      "recalling_firm": "Physio-Control, Inc.",
      "address_1": "11811 Willows Rd NE",
      "address_2": "N/A",
      "postal_code": "98052-2003",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-0294-2018",
      "product_description": "LIFEPAK 20e DEFIBRILLATOR/MONITOR, 3201273    Product Usage:  The LIFEPAK 20e defibrillator/monitor is intended for use in the hospital, physician s office, and clinic setting by personnel who are authorized by a physician/medical director. It is intended for use on a  crash cart  as well as for portable emergency response throughout a hospital.",
      "product_quantity": "3831 units",
      "reason_for_recall": "Physio-Control is aware that some devices have had power-related failures as customers prepared their device for initial deployment or during use within the first year of distribution. The symptoms of these failures may include unexpected power on and power off, device lock-up, or a failure to power on or off, any of which has the potential to result in a failure to deliver therapy to the patient and serious injury or death.",
      "recall_initiation_date": "20171204",
      "center_classification_date": "20180111",
      "termination_date": "20240401",
      "report_date": "20180117",
      "code_info": "45039371  45041785  45044636  45057093  45067882  45073270  45450645  45491315  45491441  45491644  45808476  45813092  45598878  45599806  45073745  45435012  45439562  45278002  45307727  45307732  45307742  45308078  45309308  45309310  45309317  45309923  45310487  45310523  45489720  45321756  45605113  45167375  45160349  45540221  45066559  45551722  45253909  45635361  45313588  45314492  45321757  45173154  45807047  45694119  45041460  45041335  45041572  45486374  45439205  45451242  45468563  45860139  45321722  45399194  45040553  45937962  45860005  45889371  45890053  45890181  45890310  45890600  45168699  45486960  45490591  45802737  45802739  45949393  45874190  45802736  45057137  45173422  45175430  45175496  45257327  44963488  45045796  45045990  45046296  45074231  45247320  45314855  45391419  45445721  45579549  45624977  45625366  45933892  45935769  45935894  45936081  45936175  45936811  45371506  45399258  45584143  45584240  45584531  45591943  45592637  45592826  45604856  45625247  45073272  45111392  45260403  45260814  45274367  45304697  45307492  45308174  45390779  45463390  45464179  45543704  45828889  45166687  45312978  45312991  45313074  45313083  45313100  45313174  45313405  45313592  45313824  45313980  45313993  45314061  45314071  45314134  45314230  45314238  45314356  45314516  45314673  45429741  45430061  45430472  45463404  45464415  45471456  45472539  45476795  45543521  45599919  45604525  45624723  45626387  45635109  45635444  45635762  45636110  45873406  45873427  45873809  45873903  45880717  45880719  45880722  45880775  45880807  45880812  45880817  45880843  45880906  45880909  45880912  45881226  45881832  45881834  45881836  45881837  45882215  45158987  45392115  45392169  45469664  45540364  45540508  45542915  45542972  45543122  45259450  45269320  45430201  45435880  45437503  45437998  45439118  45445967  45446520  45446683  45446933  45451695  45451811  45451974  45457626  45457734  45463253  45463296  45463481  45463510  45516006  45584666  45584880  45625468  45663729  45811955  45866408  45058375  45105056  45166654  45166795  45175743  45175963  45335125  45335281  45335440  45337595  45337680  45337902  45337903  45337991  45338061  45369310  45369913  45369965  45370034  45370053  45370177  45370417  45052251  45052252  45111390  45111391  45111394  45160397  45247610  45845768  45054151  45255396  45256441  45260178  45260276  45260323  45260646  45260834  45260984  45261393  45261500  45261647  45261961  45262052  45262072  45268155  45873875  45882795  45886829  45937965  45938312  45938992  45940505  45940605  45942131  45942688  45044212  45247221  45391952  45744210  45841660  45247129  45255076  45255378  45255467  45274524  45274612  45277208  45277796  45278336  45592897  45593019  45596607  45596789  45664056  45664435  45042043  45044053  45052560  45052565  45438619  45451041  45829209  45485282  45486094  45811493  45309481  45394304  45394358  45394824  45394949  45398402  45415453  45415971  45416971  45431332  45431387  45433134  45433408  45433511  45433535  45433680  45433790  45434241  45434337  45434349  45434440  45434977  45435348  45437441  45437460  45437744  45437871  45438030  45438616  45439282  45439890  45592860  45592942  45596492  45613565  45613786  45629061  45631448  45664250  45664534  45664629  45665102  45666012  45668953  45668955  45668956  45408975  45310262  45557034  45558098  45544927  45742531  45743700  45744346  45744461  45744685  45744776  45744845  45755369  45758380  45776597  45737568  45737984  45145880  45439280  45032751  45033009  45274704  45057134  45074089  45102920  45103076  45103187  45104274  45110839  45110840  45110842  45111167  45314374  45529769  45167124  45168465  45176131  45382990  45036900  45052246  45159353  45558305  45159518  45166719  45277579  45251093  45864380  45864535  45864659  45866542  45866603  45871792  45882955  45949255  45937209  45585358  45585359  45585364  45585367  45585369  45585835  45586080  45586091  45586095  45586097  45586101  45586118  45586313  45586391  45586514  45589736  45589839  45816190  45819381  45819493  45614315  45587016  45260209  45527776  45039695  45371011  45385516  45397567  45401088  45408722  45411464  45415177  45435175  45940723  45940914  45052631  45592703  45813394  45614053  45557407  45465321  45543592  45557702  45845970  45846298  45846392  45846735  45159436  45277549  45314158  45335278  45342093  45391019  45391119  45395525  45395815  45395978  45396084  45400748  45400826  45401520  45401727  45408908  45409013  45410463  45410615  45410746  45411593  45412009  45412450  45417053  45599029  45604378  45604874  45604988  45605000  45617230  45622960  45623129  45623249  45623536  45623629  45623850  45624255  45624404  45624580  45624906  45634634  45634735  45634840  45634996  45938302  45288161  45289061  45289086  45289393  45290238  45292504  45292836  45292861  45293048  45296979  45298178  45298213  45298216  45298220  45298222  45298223  45298253  45298344  45298536  45298625  45298677  45298762  45298783  45298789  45298793  45298797  45298864  45298886  45298929  45301146  45301483  45301524  45301721  45301796  45246538  45246661  45268666  45269502  45073966  45883035  45285033  45285231  45285519  45302300  45663892  45882712  45308979  45313433  45320007  45255581  45757170  45073972  45297463  45483375  45485657  45039309  45407891  45073271  45482419  45490824  45883119  45557086  45942352  45664908  45863293  45268511  45255817  45033066  45384710  45435975  45380991  45864832  45246224  45253953  45599027  45253862  45254882  45255227  45255268  45255401  45255583  45256518  45256520  45256521  45257042  45257178  45257265  45257377  45259222  45259277  45259340  45259386  45828718  45044279  45937031  45886276  45342525  45590142  45255592  45160562  45755969  45828899  45307593  45313208  45315603  45370437  45379232  45379283  45379695  45380541  45037270  45057141  45057679  45057837  45255826  45341311  45278755  45579142  45527892  45465403  45544253  45316235  45943360  45631132  45938281  45073755  45584655  45232520  45256000  45275086  45275176  45813260  45376982  45145494  45435959  45303345  45040967  45278000  45302321  45302832  45303744  45303987  45593539  45616982  45278343  45044602  45045661  45045798  45145663  45430928  45435460  45435824  45438531  45439091  45743883  45302705  45303423  45636057  45579806  45580950  45581120  45581546  45743293  45056561  45439427  45471710  45758018  45603432  45056558  45057136  45057409  45485818  45540687  45540801  45255814  45104517  45485623  45495468  45380024  45380398  45810819  45666081  45882647  45882928  45885006  45308270  45314578  45669275  45257044  45259512  45557746  45559211  45783976  45810901  45485024  45485365  45813150  45412582  45584359  45584605  45483384  45518965  45519159  45519530  45880908  45146133  45037407  45743545  45069955  45052685  45520675  45521404  45757162  45310101  45378896  45391598  45395040  45397819  45614366  45255004  45873874  45860452  45158200  45603341  45783978  45811227  45593688  45593754  45596494  45145230  45041667  45304773  45037341  45600035  45307028  45309288  45309617  45310529  45611263  45612530  45612769  45635865  45663489  45540094  45785666  45438251  45552163  45552584  45561892  45564454  45941607  45941767  45941907  45274837  45277164  45277395  45379899  45390703  45393958  45398044  45485383  45486376  45491723  45829146  45829309  45845544  45845884  45485821  45303404  45254893  45254929  45255228  45259999  45260103  45416685  45416846  45057139  45277880  45543927  45593510  45600127  45630384  45545470  45552104  45552234  45552271  45552893  45557801  45558923  45559610  45569292  45569295  45569376  45569772  45569817  45569818  45569819  45570264  45574066  45574262  45574289  45574537  45574778  45578557  45578560  45578561  45578878  45578882  45578883  45579256  45579258  45579259  45581610  45581712  45581830  45581909  45584133  45584134  45584135  45584684  45585230  45585297  45585472  45585655  45256557  45257043  45259211  45269058  45269458  45269788  45269869  45270072  45272460  45272661  45272729  45278709  45278977  45279339  45279341  45279346  45279503  45279526  45279531  45279691  45279730  45279734  45279801  45279926  45279977  45279984  45280011  45280133  45280140  45280246  45280304  45280322  45282081  45282088  45282106  45282170  45282220  45282225  45282290  45282485  45282577  45282581  45282630  45282738  45282740  45282820  45283218  45283238  45283334  45283728  45283731  45283737  45283811  45283891  45283927  45283932  45283974  45284151  45284187  45284682  45284683  45284689  45285087  45285123  45285125  45285167  45285307  45285316  45285361  45285570  45285584  45285622  45285745  45285749  45285798  45286168  45286196  45286197  45286246  45286313  45286314  45286315  45286329  45286350  45286353  45286405  45288152  45288201  45288210  45288258  45288340  45288481  45288491  45288529  45288532  45288557  45288604  45288659  45288672  45288723  45288743  45288860  45288883  45288991  45288999  45289027  45289068  45289094  45289117  45289119  45289216  45289239  45289362  45289784  45289794  45289884  45289914  45290033  45290108  45290156  45290187  45290191  45290337  45290390  45290573  45290617  45290762  45290766  45290831  45291041  45291095  45291137  45291257  45291268  45291276  45291283  45291293  45291303  45292454  45292487  45292505  45292532  45292551  45292552  45292580  45292614  45292616  45292748  45292755  45292766  45292768  45292778  45292781  45292791  45292799  45292804  45292810  45292816  45292819  45292826  45292838  45292847  45292853  45292859  45292878  45292986  45292988  45292990  45292993  45292996  45292998  45293000  45293005  45293006  45293018  45293020  45293024  45293028  45293031  45293032  45293041  45293054  45293069  45293070  45295552  45295565  45295575  45295577  45295581  45295669  45295730  45295885  45296000  45296045  45296146  45296158  45296253  45296299  45296307  45296310  45296317  45296349  45296504  45296621  45296709  45296751  45296770  45296867  45296896  45297065  45297080  45297448  45297457  45297461  45297471  45297571  45297900  45298050  45031072  45391209  45041153  45145501  45145708  45166646  45166748  45275005  45369803  45371039  45379014  45390664  45401213  45408282  45408804  45411814  45412870  45415324  45416206  45416570  45431049  45431260  45433227  45486386  45584661  45585089  45437705  45933548  45810846  45390558  45399116  45758143  45758145  45303380  45259822  45381141  45158106  45073837  45369987  45370138  45321849  45807826  45316041  45890755  45860642  45810822  45446587  45369115  45407693  45416506  45434030  45290239  45849674  45598940  45041695  45369079  45369853  45446486  45557145  45868846  45869314  45593664  45616863  45552582  45580991  45304610  45304694  45304698  45304699  45304701  45304855  45307176  45307189  45307194  45307351  45307357  45307598  45859458  45737796  45593756  45938754  45382925  45175018  45556707  45260818  45277254  45743071  45743512  45483465  45528408  45529693  45538415  45304605  45304742  45307648  45301930  45180887  45559278  45463433  45543313  45543734  45543880  45543942  45599030  45811329  45812631  45813746  45942119  45942766  45942865  45942962  45943443  45943547  45943645  45944139  45948575  45948604  45948721  45948883  45949016  45949663  45949840  45949946  45950011  45950861  45952318  45952368  45952370  45952371  45952597  45382720  45382763  45383266  45394169  45541445  45603473  45669263  45669277  45756140  45756151  45757143  45757151  45758024  45777231  45778459  45057942  45256008  45277801  45872036  45260030  45259221  45631454  45471495  45552399  45597462  45794152  45808121  45828924  45941406  45942591  45390460  45041042  45041571  45044019  45044123  45044588  45044600  45045559  45068220  45953044  45953128  45953198  45309136  45310175  45310178  45376546  45378865  45379055  45383054  45383585  45383685  45383795  45384615  45811059  45828696  45407747  45873973  45874304  45874315  45663358  45145943  45596499  45519440  45882555  45889251  45807259  45145822  45886431  45412677  45415033  45871645  45938750  45938758  45941171  45941173  45471854  45757159  45384818  45408333  45408568  45408664  45052248  45052441  45068218  45068223  45158775  45269624  45272549  45272601  45063297  45066561  45457806  45472449  45472637  45476382  45476970  45482729  45482822  45489434  45464356  45597868  45600646  45030658  45110841  45146038  45147871  45147962  45153397  45153490  45153586  45153726  45153816  45154274  45154282  45156355  45158192  45543239  45545019  45552488  45556804  45557283  45557391  45557871  45558643  45558755  45569831  45569904  45570130  45572265  45572331  45572404  45572517  45572648  45572708  45572844  45572889  45572947  45573094  45573203  45573263  45573363  45573491  45573589  45573653  45573839  45573900  45573978  45574028  45574236  45578498  45578582  45578821  45578925  45579267  45579691  45580114  45953426  45953998  45955706  45955745  45955836  45955947  45956162  45956410  45956619  45957138  45599345  45073473  45604320  45623947  45412794  45383153  45431457  45256519  45259218  45259443  45259531  45259741  45485161  45046372  45309388  45320101  45320107  45940979  45941505  45626102  45632311  45045806  45612388  45613638  45277258  45278004  45581695  45581886  45166942  45777814  45245606  45246748  45246972  45247040  45665463  45068214  45270019  45046189  45070011  45073957  45073961  45259668  45544347  45158240  45412224  45073967  45041441  45044606  45044773  45052333  45584673  45882699  45324663  45888842  45167299  45408201  45521321  45882791  45519274  45936915  45622924  45303933  45888442  45777606  45223952  45315964  45846766  45846811  45846876  45848401  45544037  45544908  45545115  45568865  45568951  45569023  45859768  45464142  45881362  45604532  45777417  45777493  45307555  45758023  45070589  45274469  45873947  45874261  45874277  45269247  45485974  45545559  45797822  45438489  45439902  45623456  45662703  45662816  45662954  45663159  45625621  45845336  45845436  45438475  45073753  45391526  45394728  45398502  45398635  45398905  45398960  45399011  45407996  45408114  45599052  45599171  45600306  45665536  45880907  45881700  45881893  45882127  45882299  45883066  45883191  45884622  45045093  45664714  45158378  45937941  45255763  45259855  45268244  45482383  45485168  45490232  45490441  45883093  45883152  45888410  45935577  45483605  45613637  45277794  45485366  45032948  45037565  45039386  45039546  45039665  45039728  45040051  45040421  45040579  45040639  45040672  45073277  45103284  45103499  45103802  45103959  45104052  45104589  45104688  45104986  45260695  45260812  45260830  45260839  45277250  45301001  45301113  45301195  45301350  45307766  45307993  45309765  45310087  45312841  45313253  45313803  45314254  45314438  45314555  45315733  45315897  45319608  45321276  45321565  45321584  45321712  45321744  45321816  45321899  45321986  45322030  45322097  45324335  45324429  45329939  45367949  45368097  45368130  45368180  45368545  45368830  45368839  45368867  45370948  45370960  45371496  45374093  45374368  45374449  45374732  45375661  45376138  45376446  45377198  45578401  45587025  45593348  45593364  45593590  45597054  45597176  45597441  45597448  45597453  45580228  45580399  45301317  45301333  45301343  45301359  45302298  45302308  45304069  45307048  45307369  45308861  45552237  45597708  45600015  45600030  45603349  45603391  45603434  45603465  45603585  45603640  45603717  45603808  45603870  45603876  45603900  45603918  45604017  45604075  45605198  45605231  45611072  45611079  45611080  45611155  45611261  45611264  45611419  45611611  45611612  45612385  45612387  45612622  45612628  45612630  45612978  45613636  45614049  45614052  45288399  45292579  45297467  45297640  45297733  45246428  45468357  45485260  45485602  45557163  45557605  45559720  45559801  45561893  45561894  45564452  45564455  45569166  45037113  45039805  45636244  45564453  45541092  45542905  45543234  45543466  45872996  45873286  45301445  45302065  45302186  45302301  45302575  45304142  45304341  45307191  45307288  45308732  45439935  45465723  45469783  45482805  45227223  45262133  45262221  45268115  45268290  45268566  45603578  45269953  45392054  45485888  45540970  45541470  45542966  45543104  45543296  45543802  45544000  45544241  45545240  45545375  45545631  45551129  45551130  45551172  45551489  45551520  45551522  45551569  45551625  45551753  45551754  45551757  45552040  45552102  45552106  45552115  45587030  45591261  45593402  45596486  45600021  45600652  45600655  45600814  45611613  45612079  45612652  45612980  45613464  45613676  45613896  45614271  45619763  45619771  45619772  45619776  45619965  45620448  45625818  45625905  45625937  45625996  45626039  45626042  45626100  45629192  45629413  45629777  45629795  45629808  45630385  45630387  45630390  45630393  45599463  45665400  45863721  45873019  45873857  45619756  45813043  45612381  45846172  45859603  45873053  45260730  45303396  45433679  45489277  45489879  45495469  45785349  45949059  45949527  45949592  45058046  45068219  45880811  45553050  45559504  45037321  45039620  45040814  45066555  45068127  45068558  45074206  45301322  45302291  45302421  45303389  45303399  45303405  45303699  45304053  45304062  45304073  45304079  45304211  45307316  45308824  45308841  45310543  45312825  45385412  45439469  45446055  45446781  45465603  45467139  45467294  45482866  45483047  45483111  45485025  45485079  45485219  45485555  45485822  45873358  45611259  45811462  45464090  45465667  45472812  45473204  45520989  45872914  45873335  45859708  45864185  45867610  45869288  45873285  45848168  45860754  45885621  45888601  45888676  45614756  45813423  45933463  45936585  45936708  45860935  45809207  45829012  45485252  45551131  45551234  45551890  45552219  45490010  45068336  45828765  45453829  45453654  45341139  45307718  45308866  45310530  45316540  45465782  45475802  45476223  45544809  45552583  45557663  45322083  45309254  45315852  45341261  45037517  45039305  45039795  45040280  45040453  45806738  45806855  45806968  45807113  45807174  45434666  45251523  45252266  45253812  45262286  45631442  45631836  45324822  45324824  45324825  45324826  45156530  45158105  45158390  45158791  45159123  45166825  45167188  45167244  45179914  45809225  45813169  45044598  45052566  45158202  45158395  45160563  45166683  45166726  45166757  45166986  45166987  45168514  45168727  45247554  45251377  45321624  45321845  45322092  45325378  45330266  45335670  45612256  45383517  45552236  45883270  45884837  45885223  45885819  45445699  45037459  45037481  45039747  45040202  45041763  45041883  45042179  45057554  45066501  45066557  45066558  45066589  45066712  45073274  45889767  43384379  45041775  45147804  45275227  45070029  45145732  45145780  45030742  45052564  45177117  45430843  45481970  45482088  45482135  45486378  45528206  45528304  45709951  45710105  45737566  45737567  45744217  45785364  45795479  45797193  45435376  45385090  45054727  45670604  45670605  45219278  45219455  45251235  45252028  45252354  45252464  45252568  45252633  45252850  45253690  45331243  45346409  45346752  45346794  45354974  45355760  45355951  45355958  45362852  45567252  45578385  45483571  45486448  45315167  45315340  45316618  45322230  45329304  45329362  45147442  45069123  45316006  45486881  45493567  45493625  45494153  45494543  45495768  45539541  45540340  45545370  45554085  45554147  45556850  45562822  45567891  45568231  45568343  45568607  45568955  45568958  45568975  45315352  45868219  45364409  45823903  45592622  45591738  45591999  45603121  45604160  45604245  45447205  45604255  45370495  45412956  45360758  45361814  45849725  45849761  45053709  45362844  45155382  45455109  45827558  45871900  45864556  45355704  45228780  45278260  45030796  45031559  45031644  45041797  45046456  45063163  45063688  45063695  45272850  45315168  45315209  45315533  45315568  45315697  45315881  45316866  45348005  45348077  45348134  45348157  45168995  45187725  39281983  45603136  45603244  45628912  45674854  45316606  45169095  45591596  45816083  45842309  45845214  45218581  45737171  45154734  45827623  45425994  45453628  45356145  45356205  45356310  45357022  45357102  45357170  45357180  45357334  45357476  45533953  45454798  45590361  45591187  45849248  45883047  45886321  45868123  45495526  45495783  45508694  45508752  45508948  45509168  45509311  45509541  45509751  45509897  45509997  45510187  45510418  45510510  45510662  45510769  45510911  45511101  45511298  45513318  45358291  45358296  45358469  45358472  45358644  39245365  45030995  45319980  45320151  45325159  45325843  45325844  45935771  45936302  45936564  45936754  45936833  45936960  45937224  45937951  45938104  45938321  45938484  45938684  45938737  45938842  45940574  45942571  45942607  45942728  45942814  45942920  45943199  45943447  45348396  45349920  45360643  45155556  45159657  45346297  45346621  45568121  45356888  45363863  45364327  45364353  45366031  45366123  45819202  45331837  45810957  45737512  45737570  45737629  45319428  45346067  45346439  45346490  45346669  45346745  45346793  45346830  45347126  45355718  45355721  45355953  45356276  45363536  45363844  45364855  45366033  45366263  45848778  45252788  45252929  45346832  45354985  45567616  45567633  45567650  45567655  45568117  45569244  45581280  45665646  45665847  45665908  45665971  45818076  45407639  45058698  45842040  45842830  45863612  45252851  45618309  45172728  45885183  45902475  45218222  45316728  45320283  45325152  45325160  45329621  45329859  45330013  45330089  45330258  45330736  45330746  45331205  45331311  45331500  45331526  45331586  45331737  45331958  45335148  45346059  45346195  45532339  45486531  45357947  45358029  45358230  45358239  45358301  45358379  45358384  45320446  45758063  45737853  45791301  45813660  45814880  45816196  45818073  45818478  45819925  45540749  45366032  45160585  45316575  45316740  45319543  45319595  45320904  45329138  45330581  45330582  45330639  45330967  45331996  45335524  45335527  45335914  45335928  45335929  45335933  45336778  45336788  45337436  45337515  45475688  45477329  45481845  45482093  45482181  45482270  45483510  45495368  45803518  45346367  45346534  45346789  45346790  45355655  45356109  45483097  45540979  45578382  45591337  45314857  45587376  45429650  45430587  45431042  45172884  45062307  45485989  45742179  45058694  45842841  45357977  45701571  40415427  45069136  45347164  45736929  45477409  45481844  45316603  45316846  45319558  45319605  45320419  45320433  45325904  45936632  45361923  45358572  45358575  45049271  45356320  45358177  45358573  45360785  45360832  45360892  45360899  45361025  45361110  45361366  45361413  45361507  45362045  45362692  45362702  45363081  45363241  45363298  45363438  45366107  45366283  45366356  45366358  45366404  45366470  45366631  45366636  45367741  45367753  45367760  45367762  45367763  45367846  45367890  45418671  45418911  45418975  45419151  45419308  45419379  45419442  45419488  45419736  45420102  45420259  45420338  45420687  45420733  45828867  45532430  45349923  45348142  45346721  45354951  45355717  45355739  45355935  45348887  45349919  45170971  45335750  45743616  45743979  45744862  45842722  45845953  45846227  45883196  45047766  45736986  45737446  45058365  45670792  45670793  45670795  45670804  45670981  45671180  45671182  45671198  45671215  45671287  45671325  45671326  45671328  45671330  45671401  45671439  45671441  45671465  45671572  45671681  45674851  45674955  45675149  45675429  45675645  45675673  45675791  45675964  45676156  45676187  45676402  45676454  45676529  45676593  45676666  45676956  45677012  45677098  45677161  45677184  45677213  45677266  45677286  45677341  45677474  45677484  45677550  45677575  45677631  45677641  45677681  45677685  45677753  45677755  45677819  45677827  45678092  45678139  45678208  45678222  45678250  45678260  45678268  45678282  45678288  45315720  45616870  45617765  45617790  45618739  45624458  45624745  45674685  45362610  45430912  45430954  45431330  45431383  45431440  45431622  45435463  45446773  45447113  45447747  45449452  45450214  45450403  45454485  45475533  45532827  45534353  45534438  45535567  45535638  45370631  42918026  45455390  45483270  45220042  45447474  45751377  45943331  45943374  45361267  45361627  45168930  45355952  45356305  45356919  45357048  45758284  45146607  45357952  45361914  45362556  45362616  45362781  45362907  45362934  45363184  45363262  45363363  45363411  45366058  45366131  45366160  45366162  45366165  45366198  45325902  45329161  45329199  45329477  45330300  45330399  45330400  45330401  45330674  45330957  45331035  45335923  45187343  45146881  45529069  45356110  45058357  45062506  45154569  45155780  45603320  45319797  45320583  45322114  45813792  45160391  45171263  45049228  45056912  45059425  45061128  45811481  45815833  45331515  45253234  45054142  45356143  45936290  45515873  45355886  45844132  45846390  45049553  45049751  45049790  45054825  45056679  45056803  45056869  45218413  45218909  45218975  45219339  45818695  45355592  45358178  45358295  45358707  45358737  45358773  45358841  45457192  45867639  45888712  45618748  38376072  45324771  45170880  45792010  45792409  45848954  45849192  45482484  45789318  45791298  45743051  45529635  45529830  45530104  45532714  45533381  45534247  45062091  45319743  45355510  45355928  45483015  45155393  45155443  45155511  45155530  45155640  45155735  45777831  45738002  45827440  45354829  45567562  45567630  45483248  45803516  45815604  45816208  45820564  45822835  45590577  45590722  45603067  45603155  45603274  45349921  45058794  45054706  45530451  45530606  45532247  45533236  45159604  45167585  45168420  45169036  45169152  45170881  45356849  45863681  45881898  45842732  45358898  45473375  45590346  45618192  45864052  45869727  45882714  45486770  45316125  45251378  45454673  45538685  44753155  45744524  45748696  45054194  45054140  45272913  45278377  45842172  45844572  45844912  45845167  45845406  45846738  45849375  45849434  45860708  45863168  45864370  45864425  45868574  45868649  45872052  45882853  45883114  45885620  45885726  45886427  45886559  45902018  45902237  45935591  45940913  45941118  45941273  45155680  45866446  45866851  45809295  45156049  45320793  45776427  45776431  45812325  45812444  45812960  45814921  45814923  45814938  45815225  45815252  45815875  45054288  45337042  45579810  45580164  45580982  45949621  45514305  45758278  45901684  45853159  45853725  45737075  45539317  45153390  45048833  45354984  45355723  45355915  45355998  45356307  45356313  45356349  45449377  45058362  45031431  45041792  45042276  45047910  45049436  45171016  45445810  45446045  45446147  45446583  45804175  45278991  45349828  45519958  45520145  45520447  45530258  45530336  45742734  39849895  45046378  45052987  45053440  45061132  45063164  45066256  45069120  45532555  45167278  45331390  45331553  45331931  45336997  45337420  45341420  45341550  45341740  45356171  45337417  45030806  45863534  45569379  45674704  45062318  45222405  45252816  45252947  45253406  45316822  45325161  45617747  45617762  45617772  45252679  45346792  45867107  45346668  45346860  45347988  45355008  45356108  45842773  45053539  45054445  45059620  45069128  45159570  45159740  45160143  45160870  45170969  45171004  45171115  45171274  45171335  45172739  45174929  45175355  45175521  45187515  45187634  45189595  45429109  45319450  43742518  45363672  45366067  45370510  45370515  45370598  45370806  45370850  45370910  45371017  45938246  45319383  45329119  45330418  45342828  45315210  45315871  45316689  45316690  45316818  45316854  45319391  45320357  45320637  45320790  45320836  45322178  45325969  45329153  45329170  45330408  45330484  45330485  45336611  45343164  45882931  45848736  45859148  45864745  45890459  45890550  45890614  45891684  45738068  45273382  45545249  45553428  45553435  45554073  45568130  45671183  45671386  45671438  45671655  45675293  45047745  45047786  45047822  45047895  45048689  45048735  45048790  45049180  45058228  45058235  45061832  45331085  45356208  45832094  45846790  45848609  45848679  45848988  45849127  45849221  45849414  45849580  45347991  45348263  45348385  45348389  45348496  45348507  45348807  45348899  45349329  45349973  45350049  45350050  45350053  45350058  45350077  45350125  45350158  45350165  45350182  45350189  45354814  45354816  45354830  45354908  45354983  45811181  45814933  45816515  45366055  45355771  45356194  45356300  45356304  45356315  45356324  45356850  45356858  45357037  45357040  45357046  45357074  45357151  45357250  45357279  45357281  45357282  45357316  45357506  45357618  45357639  45357642  45357643  45357974  45867200  45364827  45783369  45253759  45782864  45363748  45175067  45592708  45678275  45678286  45776210  45776215  45802363  45251458  45360462  45445976  45737750  45792006  45370951  45355867  45616854  45777490  45325841  45159566  45160135  45160232  45356127  45370696  45513763  45356848  45363849  45370472  45355028  45530730  45483866  45491828  45252524  45361815  45587413  45589740  45590074  45590331  45811952  45709069  45709517  45734951  45735317  45320003  45322218  45738282  45545106  45146601  45146712  45146832  45147198  45147201  45147433  45147439  45147447  45147450  45147587  45154551  45154667  45156290  45154458  45154766  45154859  45155808  45156476  45159313  45159465  45159746  45160765  45828897  45540569  45819430  45346596  45346689  45777204  45066253  45859462  45859630  45866762  45938932  45811597  45813089  45567836  45792005  45222388  45433141  45491155  45539623  45540136  45545477  45556686  45579807  45579808  45580162  45580163  45580989  45530030  45533845  45535070  45535330  45587200  45587277  45587416  45590767  45591302  45538072  45346587  45356064  41557239  45349933  45058680  45059696  45063191  45063842  45431128  45533078  45533650  45534635  45535482  45539343  45540484  45545571  45783035  45853038  45853591  45859364  45901579  45315929  45445709  45777494  45777827  45155669  45156038  39115037  45316017  45358643  45360613  45361495  45888993  45493346  45493398  45495255  45520049  45520330  45273596  45346310  45347100  45331528  45331615  45331788  45331822  45331830  45331857  45331872  45331997  45335976  45336074  45336400  45336503  45341866  45341920  45884632  45885045  45889111  45360463  45361090  45362742  45518282  45849789  45890302  45486344  45587410  45618509  45623981  45625248  45625586  45802568  45515187  45530530  45538078  45538321  45538822  45538895  45540821  45367343  45418741  45419544  45420198  45420455  45425802  45425927  45426097  45426194  45426305  45426363  45426423  45426572  45426708  45426749  45427063  45427171  45427242  45427299  45427447  45427505  45427554  45427597  45429064  45429232  45429456  45429753  45429920  45430202  45430411  45430853  45518504  45816146  45819876  45052809  45053490  45053495  45053644  45054037  45054113  45054534  45054722  45054728  45056547  45058688  45059431  45059614  45061122  45061125  45061131  45168914  45528631  45319940  45367471  45329165  45347169  45861093  45871663  45883305  45031642  45346711  45412877  45701449  45701753  45701822  45701904  45701941  45702014  45702210  45707281  45707405  45707717  45708524  45736591  45736737  45514834  45515352  45515695  45515784  45775151  45777378  45777838  45778664  45782660  45783209  45802570  45803517  45804166  45804171  45804172  45807291  45807446  45807658  45807771  45807894  45807976  45808347  45808411  45808585  45808706  45809372  45810815  45220118  45251254  45251463  45347968  45347971  45348001  45348003  45348004  45348046  45348170  45348259  45678248  45322177  45348237  45348371  45348373  45348386  45348402  45348548  45348771  45350013  45350177  45354815  45354835  45355012  45355027",
      "more_code_info": "45355040  45355069  45355070  45355116  45355158  45355225  45355237  45355243  45355287  45355291  45355295  45355324  45355327  45355328  45355472  45355475  45355479  45355657  45356142  45356144  45356343  45356859  45356918  45357609  45357673  45357681  45357760  45357813  45357848  45862785  45863355  45863832  45425678  45845073  45322122  38814700  45456946  45456991  45457124  45849496  45737601  45813737  45818079  45819257  45819974  45820028  45820060  45820510  45832164  45832352  45832453  45832538  45848151  45848179  45848213  45848372  45848484  45848501  45848503  45848642  45848739  45848807  45848895  45849073  45849292  45849444  45849522  45617519  45513413  45513642  45513897  45514099  45539023  45539239  45758280  45774916  45775158  45775163  45775676  45775945  45776086  45776420  45777835  45046373  45325157  45319540  45362552  45363510  45518321  45844177  45834154  45840816  45860090  45860988  45849677  45860903  45861117  45863196  45863460  45864260  45816461  45825961  45828692  45168799  45578384  45842655  45355720  45363618  45363737  45058697  45457384  45482390  45349136  45455210  45062641  45063692  45885428  45869042  45844127  39635956  45192066  45355724  45355744  45061966  45153392  45153467  45153476  45153603  45153610  45153809  45153853  45154025  45154298  45154427  45154549  45154634  45155341  45167231  45167336  45167407  45168133  45168371  45172805  45172839  45172910  45348550  45349899  45349971  45074489  45155935  45160488  45171075  45171413  45171980  45189748  45810840  45541186  45058374  45147759  45514405  45624262  45707942  45710206  45710261  45710356  45710408  45736449  45933881  45941205  45941423  45941516  45941637  45941763  45941923  45942176  45943628  45948576  45948787  45949015  45949211  45949538  45949772  45949976  45950164  45950297  45950603  45952317  45952417  45952562  45054152  45347996  45349840  45049347  45361350  45049501  45346748  45347972  45591607  45363160  45173107  45853226  45866379  45866638  45869249  45869548  45872271  45872380  45880716  45884681  45041804  45048117  45054638  45058926  45059606  45061119  45061369  45062399  45062831  45063165  45063397  45063470  45063646  45063740  45063929  45066211  45066250  45066272  45066356  45074485  45074490  45074492  45074494  45075286  45145974  45146146  45154875  45532913  45355533  45360939  45361631  45361854  45361951  45362550  45366156  45885914  45940510  45314785  45350075  45515501  45533677  45743882  45750404  45743492  45057934  45581357  45046377  45053365  45054146  45056551  45059116  45069125  45320190  45173141  45357978  45360955  45337426  45743372  45744318  45744436  45749006  45473839  45475375  45802377  45802569  45803515  45803519  45808271  45809161  45811319  45811816  45812197  45813256  45813369  45842820  45842879  45844200  45844571  45844764  45844858  45845014  45845032  45845305  45845754  45845853  45846653  45849724  45849742  45849787  45849830  45851407  45852277  45852425  45852512  45852698  45859995  45860461  45860601  45860766  45861027  45849027  45849099  45859176  45859209  45859566  45859732  45859807  45864638  45884936  45901472  45056543  45848605  45708698  45735553  45735692  45736138  45743254  45743738  45744741  45744815  45744935  45750536  45750907  45751020  45751101  45751194  45751441  45751444  45751446  45755386  45755519  45755547  45755552  45756123  45756806  45757839  45758065  45758067  45253605  45319877  45325838  45329574  45329694  45329706  45329886  45330238  45330718  45330729  45330731  45331140  45331376  45849577  45358297  45253807  45866552  45252447  45253155  45062233  45063154  45063162  45528881  45748503  45790031  45790835  45791213  45791300  45792008  45792411  45370745  45218627  45219144  45844427  45056624  45062082  45362061  45360971  45823037  45823235  45823495  45823516  45823540  45823566  45823669  45823746  45823858  45825120  45825138  45825210  45825216  45825248  45825363  45825587  45825712  45825762  45355005  45061500  45061589  45061743  45062319  45062322  45062347  45066214  45155404  45940658  45943542  45355950  40704880  45533529  45538582  45545043  45553148  45553304  45553432  45553549  45553923  45554007  45554074  45557922  45558244  45558625  45558714  45558861  45559107  45559424  45559560  45559705  45559811  45559872  45561906  45561959  45562048  45562065  45562066  45562144  45562279  45562356  45562358  45562359  45562373  45562452  45562620  45562621  45562622  45562651  45562996  45562997  45562998  45563091  45563162  45563169  45563171  45563172  45563336  45563351  45563353  45563355  45564152  45564153  45564154  45564155  45564373  45564461  45564754  45564840  45564925  45565131  45565199  45566750  45566754  45566820  45566835  45566899  45566913  45567064  45567172  45567246  45567710  45591029  45591720  45816541  45819622  45822664  45823374  45823432  45823663  45823708  45825338  45825451  45825541  45825600  45825646  45825789  45826197  45826316  45826812  45827071  45827120  45827201  45827287  45827434  45827496  45827544  45827577  45827653  45827668  45828987  45829114  45829184  45829395  45829473  45829655  45830005  45830139  45830326  45830490  45830864  45833245  45833356  45833544  45833613  45833696  45833753  45833930  45834005  45834118  45840818  45840899  45840905  45841016  45841222  45841243  45841289  45841311  45841408  45841425  45841455  45841565  45841572  45841611  45841627  45841700  45841762  45841781  45841815  45841820  45841897  45841918  45841949  45842026  45842118  45842125  45842215  45842220  45842293  45842654  45842775  45053484  45047999  45048043  45048363  45048529  45048604  46718289"
    }
  ]
}