{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88953",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave Ne",
      "address_2": "",
      "postal_code": "55432-3568",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide, including PR.  There was government distribution but no military distribution was identified.",
      "recall_number": "Z-0293-2022",
      "product_description": "Medtronic Wireless Recharger Model WR9220 included in the RS5200 Recharger Kit used by the InterStim Micro (Model 97810) implantable neurostimulator.",
      "product_quantity": "10,081 devices",
      "reason_for_recall": "The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.",
      "recall_initiation_date": "20211104",
      "center_classification_date": "20211124",
      "report_date": "20211201",
      "code_info": "UDI 00763000186265",
      "more_code_info": ""
    }
  ]
}