  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Draper",
      "state": "UT",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78738",
      "recalling_firm": "Edwards Lifesciences, LLC",
      "address_1": "12050 S Lone Peak Pkwy",
      "address_2": "N/A",
      "postal_code": "84020-9414",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "United States: All states with the exception of: Alaska, Hawaii, Idaho, New Mexico, Rhode Island, South Dakota and Wyoming",
      "recall_number": "Z-0293-2018",
      "product_description": "Certitude Delivery System and any associated Multipacks Models - 9600CT20A, 9600CT23A, 9600CT26A, 9600CT29A    Product Usage:  The Edwards Certitude Delivery System is used for delivery of the Edwards SAPIEN 3 transcatheter heart valve (THV), typically during a transapical (TA) or transaortic (TAo) approach to transcatheter aortic valve replacement.",
      "product_quantity": "1730 units",
      "reason_for_recall": "A component in the Certitude Loader may have a  molding overflow defect that could detach during advancement of the delivery system and  potentially embolize from the loader into the patient.",
      "recall_initiation_date": "20170721",
      "center_classification_date": "20180110",
      "termination_date": "20190913",
      "report_date": "20180117",
      "code_info": "Lot Numbers:   60677653  60703356  60703356  60733369  60898859  60708907  60708907  60693311  60677343  60678033  60708907  60699067  60685057  60712982  60760615  60792834  60776987  60813952  60782083  60796436  60840392  60842807  60854965  60885861  60930443  60677270  60678035  60678040  60703355  60704423  60703355  60704423  60679106  60689105  60718006  60718007  60792838  60813951  60782088  60816613  60816614  60842808  60842810  60873492  60842809  60885862  60885863  60930444  60697333  60703357  60703357  60678007  60739663  60733367  60733366  60760822  60801453  60808313  60776988  60816615  60883652  60842811  60739661  60718977  60718978  60718979  60724851  60724852  60724853  60724854  60724855  60724856  60724857  60724858  60724859  60724860  60724861  60724862  60724863  60724864  60724865  60724866  60724867  60724868  60724869  60724870  60730280  60730282  60730283  60730284  60730285  60730286  60730287  60730288  60730318  60730289  60730319  60730320  60730321  60735835  60735836  60735789  60735790  60735791  60735792  60735793  60735794  60737235  60737236  60739823  60739824  60744633  60744634  60744636  60744637  60744638  60744635  60744823  60744824  60744825  60745017  60745016  60746726  60746727  60746728  60746729  60746730  60746731  60748826  60748827  60748828  60748829  60748831  60748832  60749200  60749201  60749214  60749213  60750611  60750612  60751302  60751303  60753189  60753192  60753191  60753190  60754451  60754452  60756176  60756177  60756487  60756488  60757572  60758167  60758168  60758169  60758170  60758183  60758184  60757932  60757931  60757766  60757765  60757573  60757976  60757975  60759299  60759300  60759485  60759486  60759588  60759589  60759704  60759705  60760609  60760610  60760901  60760902  60760903  60760904  60760905  60760906  60761419  60761420  60761421  60761422  60763201  60763206  60763207  60763202  60764509  60764510  60764511  60764512  60764532  60764533  60764640  60764641  60764647  60764648  60764725  60764726  60764615  60764616  60764604  60764605  60764733  60764734  60764785  60764786  60764787  60764788  60764812  60764813  60764814  60764815  60764818  60764819  60766845  60766846  60766561  60766562  60767048  60767243  60767242  60767293  60767292  60767408  60767409  60767410  60767411  60767412  60767413  60767008  60767009  60767010  60767011  60767012  60767013  60767014  60767015  60767016  60767017  60767018  60767019  60767020  60767021  60767022  60767023  60767819  60768519  60768520  60768521  60768524  60768522  60768523  60768653  60768654  60770516  60770517  60770736  60770735  60770734  60770660  60770661  60770863  60770864  60770865  60770866  60770867  60770868  60771056  60771185  60771186  60771193  60771194  60772777  60772778  60772779  60772780  60772781  60772782  60772787  60772788  60773097  60773207  60773208  60773209  60773246  60773247  60774347  60774609  60774610  60774794  60774795  60774793  60774796  60774797  60775232  60775231  60775236  60775233  60775234  60775235  60776722  60776723  60776814  60776815  60776853  60776854  60776855  60776856  60776857  60776858  60776868  60776869  60776901  60776902  60776981  60777071  60777206  60777205  60777222  60777223  60778473  60778474  60778492  60778493  60778494  60778495  60778727  60778755  60778756  60778760  60778761  60778762  60778763  60778949  60778948  60778947  60778946  60778945  60778944  60780651  60780672  60780704  60780705  60781015  60781062  60781063  60781064  60781065  60781066  60781067  60781068  60781069  60781076  60781077  60782212  60782213  60782214  60782215  60782216  60782217  60782225  60782224  60782860  60782861  60782916  60783197  60783198  60783040  60783050  60783051  60783169  60783650  60784014  60784015  60784340  60784341  60784345  60784346  60784495  60784496  60784546  60784547  60784548  60784549  60784550  60784551  60784757  60784758  60784759  60784760  60784761  60784762  60784770  60784771  60784772  60784773  60784774  60784775  60784993  60784994  60784995  60784996  60784997  60784998  60785010  60785011  60786506  60786771  60786772  60786773  60786774  60786775  60786776  60786777  60786778  60786779  60786846  60786942  60786943  60787113  60787114  60787115  60787116  60787117  60787118  60787119  60787120  60787121  60787122  60787123  60788038  60788265  60788266  60788352  60788353  60788546  60788547  60788618  60788679  60788680  60788681  60788682  60790035  60790036  60790046  60790145  60790246  60790247  60790248  60790249  60792227  60792228  60792229  60792328  60792329  60792330  60792331  60792332  60792333  60792735  60794244  60794245  60794437  60794438  60794486  60794487  60794496  60794497  60794536  60794537  60796470  60796471  60796487  60796488  60796489  60796490  60796493  60796494  60796495  60796496  60796509  60796510  60796511  60796512  60796527  60796528  60796975  60796976  60796977  60796978  60796979  60796980  60796981  60796982  60796983  60796988  60797057  60797058  60798459  60798460  60798640  60798641  60798801  60798802  60798817  60798816  60798852  60798855  60798854  60798853  60800892  60800893  60801260  60801354  60801361  60801362  60801363  60801364  60801454  60801455  60801800  60801801  60803395  60803396  60804028  60804029  60804030  60804031  60804032  60804033  60805349  60805350  60805351  60805916  60805915  60805917  60805918  60806045  60806044  60806121  60806122  60806120  60806119  60808099  60808381  60808382  60808383  60808384  60809372  60809373  60809374  60809375  60810893  60810894  60811412  60811411  60811409  60811410  60811415  60811416  60811417  60811418  60811503  60811504  60811531  60811648  60811825  60811826  60813146  60813147  60813148  60813149  60813281  60813282  60813389  60813390  60813761  60813762  60813763  60813764  60813765  60813766  60815980  60816245  60816246  60816247  60816248  60816035  60816252  60816251  60816396  60816397  60816594  60816595  60816596  60816597  60816598  60816685  60816686  60818470  60818536  60818535  60818588  60818589  60818698  60818699  60818700  60818701  60818702  60818703  60820467  60820712  60820713  60820714  60820715  60820743  60820744  60820745  60820746  60820747  60820748  60822570  60822571  60822732  60822812  60822813  60822884  60823108  60823107  60824307  60824308  60824449  60825105  60825106  60825131  60825132  60825173  60825174  60825175  60825176  60825177  60825178  60825188  60826903  60826904  60826905  60826906  60826960  60826961  60827233  60827232  60827281  60827282  60827283  60827284  60827285  60827286  60828135  60828409  60828410  60828635  60828634  60828721  60828720  60828725  60828726  60828728  60828729  60830157  60830158  60830311  60830331  60830899  60830900  60832668  60832669  60833056  60833057  60834361  60834362  60834449  60834450  60835007  60836226  60836495  60836227  60836487  60836492  60836488  60836489  60836493  60836494  60836616  60836617  60836547  60836624  60836625  60836626  60836627  60836628  60836629  60836765  60836766  60837414  60838283  60838284  60838316  60838327  60838328  60838364  60838365  60838639  60838640  60838731  60838732  60840708  60840709  60840873  60841290  60841291  60841292  60841293  60841294  60841295  60841296  60841297  60841339  60841340  60841341  60841342  60841473  60841474  60841475  60841476  60842630  60842631  60842785  60842786  60842787  60842788  60842789  60842790  60842805  60842806  60842947  60842948  60843030  60844710  60844711  60844715  60844713  60844714  60844712  60844756  60844757  60845110  60845111  60846114  60846482  60846502  60847067  60847316  60847345  60847346  60847566  60847567  60847568  60847569  60848354  60848535  60848606  60848607  60848608  60848609  60848818  60848819  60848820  60848821  60848822  60848823  60848848  60848849  60848850  60848851  60848863  60849066  60849067  60851787  60851788  60851789  60851790  60851791  60851792  60852681  60852682  60852683  60852684  60852685  60852686  60852687  60852688  60852689  60852690  60852691  60853304  60853303  60853350  60854572  60854573  60854852  60854765  60854766  60854851  60854923  60855009  60855010  60855155  60855208  60855395  60855396  60855543  60855544  60855545  60855546  60857002  60857003  60857005  60857006  60857208  60857220  60857207  60857282  60857281  60857329  60857330  60857331  60857332  60857340  60857341  60858333  60858334  60858513  60858514  60858555  60858556  60858624  60858625  60858654  60858668  60858669  60858670  60858671  60858942  60858943  60858944  60858945  60858946  60858947  60858950  60858951  60858956  60858957  60859436  60859435  60860974  60861106  60861107  60861132  60861133  60861157  60861332  60861333  60861334  60861335  60861336  60861337  60861374  60861375  60861381  60861382  60862813  60862962  60862963  60862964  60862965  60862966  60862967  60863179  60863180  60863229  60863532  60863533  60863537  60863570  60863571  60863572  60863573  60863574  60863575  60863702  60863703  60864757  60864880  60865170  60865171  60865221  60865231  60865232  60865250  60865249  60866874  60867064  60867063  60867072  60867074  60867075  60868112  60868402  60868403  60868404  60868405  60868406  60868378  60868379  60868380  60868381  60868517  60868521  60868522  60868475  60868474  60868562  60868561  60868560  60868559  60868534  60868533  60868532  60868531  60868606  60868607  60868608  60868609  60868610  60868611  60868612  60868626  60868627  60870098  60870099  60870101  60870126  60870127  60870128  60870129  60870256  60870257  60870263  60870335  60870336  60870376  60870377  60871514  60871515  60871516  60871517  60871518  60871519  60871543  60871613  60871614  60871862  60871863  60874100  60874196  60874197  60874198  60874199  60874200  60874201  60875861  60876698  60876699  60877630  60877631  60877770  60877840  60877841  60877847  60877848  60878001  60878099  60878100  60878101  60878102  60878103  60878104  60878170  60878171  60878174  60878175  60878218  60878219  60878214  60878215  60878216  60878217  60879444  60879971  60880060  60880061  60880257  60880258  60880259  60880260  60880261  60880262  60880741  60880742  60881797  60881798  60881842  60881846  60881992  60882106  60882187  60882188  60882189  60882190  60883502  60883503  60883504  60883639  60883665  60883792  60883793  60883794  60883795  60883796  60883797  60883844  60883870  60884135  60884136  60884137  60884138  60884143  60884144  60885735  60885736  60885899  60885901  60885931  60885932  60885956  60885957  60885958  60885959  60886040  60886041  60886042  60886043  60886044  60886045  60886073  60886158  60886258  60886259  60886390  60886389  60886499  60886500  60887282  60887283  60887455  60887456  60887457  60887458  60887617  60887616  60887615  60887614  60887613  60887612  60887535  60887536  60887626  60887636  60887637  60887638  60887639  60887640  60887641  60887793  60888775  60888776  60888777  60888778  60888836  60888835  60889090  60889089  60889188  60889246  60889247  60889332  60889354  60889380  60891086  60891087  60891111  60891112  60891113  60891114  60891115  60891116  60891181  60891182  60891238  60891256  60892131  60892885  60892886  60893025  60893026  60893027  60893028  60893469  60893470  60894629  60894751  60894752  60894843  60894844  60894845  60894846  60894862  60894863  60896384  60896383  60896400  60896464  60896569  60896589  60896644  60896663  60896689  60896688  60896682  60896681  60898257  60898292  60898293  60898294  60898295  60898467  60898468  60898469  60898470  60898680  60898681  60898766  60898767  60898768  60898769  60898770  60898771  60898928  60898929  60900585  60900646  60900647  60902043  60902044  60902045  60902004  60902018  60902497  60902498  60902507  60903744  60903745  60904517  60904518  60904575  60904576  60904584  60904587  60904597  60904730  60904731  60905748  60905747  60905746  60905745  60905744  60905892  60905919  60905918  60906094  60906095  60906096  60906097  60906098  60906099  60906180  60906181  60905569  60905568  60907704  60907813  60907814  60907815  60907816  60907817  60907818  60907842  60907843  60907846  60907847  60907848  60907849  60907850  60907851  60907934  60907935  60907947  60907948  60909638  60909664  60909656  60909657  60909692  60909693  60909746  60909747  60909812  60909813  60911332  60911530  60911531  60911643  60911644  60911662  60911687  60912100  60912101  60913407  60913408  60913564  60913565  60913642  60913643  60913644  60913680  60913738  60913991  60913992  60915192  60915193  60915200  60915201  60915202  60915203  60915334  60915343  60915344  60915345  60915424  60915425  60915528  60915529  60915434  60915435  60915437  60915438  60915471  60915648  60915649  60915661  60915679  60915660  60917690  60917693  60917701  60917712  60917713  60917851  60917891  60917892  60917894  60918023  60918231  60918439  60918440  60919933  60919934  60919935  60919936  60919937  60919938  60920111  60920112  60920113  60920114  60920115  60920285  60920353  60920266  60920267  60921807  60921808  60922009  60922010  60922078  60922258  60922259  60922279  60922715  60922716  60922717  60922788  60922789  60922797  60922798  60922799  60922800  60922801  60922802  60923644  60923732  60923776  60923846  60923954  60923955  60923972  60924007  60924008  60924009  60924010  60924084  60924085  60924086  60924087  60924088  60924089  60924112  60924224  60924225  60924259  60924260  60924261  60924262  60924263  60924264  60924848  60925002  60925003  60925165  60925166  60925200  60925363  60925364  60925374  60925440  60925439  60926775  60927482  60928099  60928100  60928187  60928188  60928189  60928206  60928265  60928424  60928425  60928426  60928427  60928434  60928435  60928513  60928685  60928686  60929688  60929689  60929994  60929995  60930069  60930057  60930058  60930059  60930060  60930061  60930062  60930101  60930117  60930118  60930119  60930120  60930121  60930122  60930294  60930295  60930495  60930496  60930497  60930498  60932132  60932133  60932463  60932464  60932695  60932696  60932790  60932791  60932792  60932793  60934073  60934502  60934503  60934590  60934591  60934619  60934699  60934698  60934734  60934735  60936270  60936271  60936586  60937837  60937702  60937703  60937878  60937879  60938058  60938242  60938241  60938352  60938353  60938354  60938355  60939055  60939663  60939664  60939665  60939666  60939667  60939668  60939669  60939670  60939743  60939744  60939757  60939758  60940079  60940223  60940222  60940201  60940202  60940476  60940477  60940478  60940479  60942368  60942771  60942772  60942959  60943130  60943131  60943051  60943052  60943151  60944170  60944171  60944172  60944173  60944338  60944339  60944352  60944354  60944365  60946314  60946521  60946611  60946612  60946475  60946778  60946779  60947142  60947365  60947366  60947367  60947368  60947318  60947143  60948431  60948430  60948627  60948628  60948629  60948630  60948631  60948632  60948633  60948634  60948635  60948636  60948810  60948811  60948922  60948923  60948924  60948925  60948927  60948928  60949059  60949058  60949114  60949115  60950268  60950814  60950815  60950860  60950913  60951027  60951026  60951055  60951056  60951070  60951196  60951197  60951266  60951267  60952376  60952488  60952602  60952601  60952709  60952710  60952756  60952757  60952758  60952759  60953834  60953835  60953927  60954145  60954146  60954148  60954166  60954297  60954670  60954671  60954672  60954673  60954674  60954675  60954676  60954677  60954678  60954679  60954680  60954681  60954682  60954683  60954684  60954685  60954475  60954476  60954477  60954478  60954479  60954480  60954481  60954482  60954483  60954484  60956102  60956103  60956146  60956147  60956208  60956218  60956217  60956216  60956215  60956229  60956230  60956236  60956334  60956358  60956359  60956393  60956394  60956445  60957500  60957501  60957888  60957889  60957890  60957937  60957938  60958143  60958164  60958172  60959531  60959553  60959844  60959905  60959965  60959966  60959967  60959968  60959969  60959970  60959918  60959946  60960032  60960033  60960035  60960036  60960167  60960289  60960417  60961606  60961607  60962154  60962287  60962355  60962471  60962472  60963363  60963364  60963365  60963366  60963444  60963436  60963437  60963473  60963474  60963897  60963898  60963899  60963900  60963901  60963902  60963903  60963904  60963905  60963906  60963907  60963908  60963909  60963910  60963911  60963912  60965227  60965228  60965501  60965502  60965779  60965780  60965992  60965993  60965816  60965775  60965776  60965994  60965998  60966089  60966090  60966964  60966947  60966946  60966994  60967346  60967731  60967730  60968964  60969160  60969161  60969164  60969215  60969216  60969592  60969752  60969751  60969750  60969749  60969693  60969694  60969818  60969819  60969820  60969821  60969822  60969823  60971031  60971032  60971033  60971034  60971035  60971055  60971056  60971057  60971058  60971048  60971210  60971384  60972563  60972687  60972675  60972839  60972840  60972795  60972956  60972955  60976486  60976713  60976824  60976825  60976877  60976878  60976879  60976880  60976951  60976952  60976978  60977045  60977226  60977227  60977195  60977206  60977208  60977209  60978961  60979086  60979087  60979139  60979460  60981116  60981117  60981282  60981283  60981284  60981285  60981456  60981457  60981458  60981459  60981504  60981564  60981587  60981588  60981664  60981661  60983220  60983221  60983240  60983241  60983266  60983268  60983269  60983270  60983267  60983281  60983282  60983283  60983314  60983315  60983456  60983457  60983525  60983536  60983537  60983863  60983864  60983867  60983868  60983869  60983870  60983880  60983881  60985043  60985058  60985325  60986652  60986653  60986673  60986754  60987043  60987044  60988866  60988867  60988876  60988877  60988924  60989326  60989378  60989435  60989436  60990824",
      "more_code_info": ""
    }
  ]
}