{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Montvale",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86440",
      "recalling_firm": "Pentax of America Inc",
      "address_1": "3 Paragon Dr",
      "address_2": "N/A",
      "postal_code": "07645-1782",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide and Canada",
      "recall_number": "Z-0292-2021",
      "product_description": "9372HD Digital Capture (ver. 1.1.0 + w/ 9263 endoPortal),",
      "product_quantity": "39 total",
      "reason_for_recall": "There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.",
      "recall_initiation_date": "20200922",
      "center_classification_date": "20201023",
      "termination_date": "20220516",
      "report_date": "20201104",
      "code_info": "All systems with version 1.1.0 or higher;    Computer Serial Numbers: 84086-01  84533-01  84090-16  84090-15  84090-14  84090-11  84090-09  84090-10  84090-08  84088-13  84088-11  84528-08  84087-06  84087-03  84538-06  84088-14  84089-02  84089-06  84089-03  84529-04  84528-05  81536-06  84091-01  84089-16  81536-12  84532-11  81538-05  84090-02  84089-13  84089-12  84090-06  84087-04  84531-12  84531-11  84531-10  84531-13  84091-16  84533-11  84533-12"
    }
  ]
}