  "results": [
    {
      "status": "Ongoing",
      "city": "Flagstaff",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86421",
      "recalling_firm": "W L Gore & Associates, Inc.",
      "address_1": "3750 W Kiltie Ln",
      "address_2": "N/A",
      "postal_code": "86005-8712",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. - Nationwide    O.U.S. - Australia, Austria, Belgium, Croatia, Cypress, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Monaco, Netherlands, Norway, New Zealand, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden,  Switzerland, United Arab Emirates, and United Kingdom",
      "recall_number": "Z-0291-2021",
      "product_description": "GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System",
      "product_quantity": "19013 units",
      "reason_for_recall": "When the delivery catheter was attempted to be removed, there was an observed connection between the stent graft and the delivery catheter that prevented the delivery catheter from being withdrawn.",
      "recall_initiation_date": "20200909",
      "center_classification_date": "20201023",
      "report_date": "20201104",
      "code_info": "CATALOG NUMBERS  TGM212110X*  TGM212115X*  TGM212120X*  TGM262610X*  TGM262615X*  TGM262620X*  TGM282810X*  TGM282815X*  TGM282820X*  TGM313110X*  TGM313115X*  TGM313120X*  TGM343410X*  TGM343415X*  TGM343420X*  TGM373710X*  TGM373715X*  TGM373720X*  TGM404010X*  TGM404015X*  TGM404020X*  TGM454510X*  TGM454515X*  TGM454520X*  TGM262110X*  TGM312610X*  TGMR313110X*  TGMR313115X*  TGMR313120X*  TGMR312610X*  TGMR373710X*  TGMR373715X*  TGMR373720X*  TGMR404010X*  TGMR404015X*  TGMR404020X*  *X is a placeholder for enumeration code: E=EMEA, Australia, New Zealand; J=Japan    All Lot Numbers"
    }
  ]
}