      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63314",
      "recalling_firm": "Mindray DS USA, Inc. d.b.a. Mindray North America",
      "address_1": "800 MacArthur Blvd.",
      "address_2": "N/A",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Australia, Chile, Colombia, and Venezuela.",
      "recall_number": "Z-0291-2013",
      "product_description": "Mindray A5 Anesthesia Delivery System   Manufactured in China for Mindray DS USA Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430    This is a device used to administer to a patient continuously or intermittently a general inhalation anesthetic and to maintain a patient's ventilation.",
      "product_quantity": "346 units (311 US - 35 Foreign)",
      "reason_for_recall": "Mindray has identified the potential for a system leak resulting an issue related to the carbon dioxide absorber canister gasket on A3/A5 Anesthesia Delivery Systems.  A potential system leak resulting from the canister gasket will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case.",
      "recall_initiation_date": "20120813",
      "center_classification_date": "20121116",
      "termination_date": "20130604",
      "report_date": "20121128",
      "code_info": "0631F-01000-00  Various serial numbers      0631F-01000-00 EJ-1A001411  0631F-01000-00 EJ-1A001426  0631F-01000-00 EJ-1A001431  0631F-01000-00 EJ-1A001439  0631F-01000-00 EJ-1A001443  0631F-01000-00 EJ-1C001577  0631F-01000-00 EJ-1C001592  0631F-01000-00 EJ-1C001593  0631F-01000-00 EJ-1C001622  0631F-01000-00 EJ-1A001469  0631F-01000-00 EJ-1A001416  0631F-01000-00 EJ-1A001423  0631F-01000-00 EJ-1B001497  0631F-01000-00 EJ-1B001519  0631F-01000-00 EJ-1B001524  0631F-01000-00 EJ-1B001538  0631F-01000-00 EJ-1B001544  0631F-01000-00 EJ-1B001546  0631F-01000-00 EJ-1B001549  0631F-01000-00 EJ-1B001550  0631F-01000-00 EJ-1C001572  0631F-01000-00 EJ-1C001617  0631F-01000-00 EJ-22001684  0631F-01000-00 EJ-1B001540  0631F-01000-00 EJ-1B001548  0631F-01000-00 EJ-21001634  0631F-01000-00 EJ-21001641  0631F-01000-00 EJ-21001642  0631F-01000-00 EJ-21001644  0631F-01000-00 EJ-21001650  0631F-01000-00 EJ-21001643  0631F-01000-00 EJ-21001654  0631F-01000-00 EJ-22001704  0631F-01000-00 EJ-23001760  0631F-01000-00 EJ-23001765  0631F-01000-00 EJ-23001775  0631F-01000-00 EJ-1A001437  0631F-01000-00 EJ-1A001445  0631F-01000-00 EJ-22001710  0631F-01000-00 EJ-22001724  0631F-01000-00 EJ-22001743  0631F-01000-00 EJ-1A001418  0631F-01000-00 EJ-1B001511  0631F-01000-00 EJ-23001764  0631F-01000-00 EJ-23001768  0631F-01000-00 EJ-23001769  0631F-01000-00 EJ-23001779  0631F-01000-00 EJ-23001792  0631F-01000-00 EJ-22001696  0631F-01000-00 EJ-22001701  0631F-01000-00 EJ-22001705  0631F-01000-00 EJ-22001707  0631F-01000-00 EJ-22001723  0631F-01000-00 EJ-1C001573  0631F-01000-00 EJ-1C001589  0631F-01000-00 EJ-1C001590  0631F-01000-00 EJ-1C001599  0631F-01000-00 EJ-1C001610  0631F-01000-00 EJ-1C001619  0631F-01000-00 EJ-21001667  0631F-01000-00 EJ-22001679  0631F-01000-00 EJ-22001686  0631F-01000-00 EJ-22001688  0631F-01000-00 EJ-22001721  0631F-01000-00 EJ-15001187  0631F-01000-00 EJ-1A001420  0631F-01000-00 EJ-22001699  0631F-01000-00 EJ-1B001512  0631F-01000-00 EJ-1A001459  0631F-01000-00 EJ-21001635  0631F-01000-00 EJ-21001636  0631F-01000-00 EJ-21001637  0631F-01000-00 EJ-21001658  0631F-01000-00 EJ-22001681  0631F-01000-00 EJ-22001692  0631F-01000-00 EJ-22001693  0631F-01000-00 EJ-22001698  0631F-01000-00 EJ-22001700  0631F-01000-00 EJ-22001703  0631F-01000-00 EJ-22001712  0631F-01000-00 EJ-22001734  0631F-01000-00 EJ-22001741  0631F-01000-00 EJ-1A001449  0631F-01000-00 EJ-1A001471  0631F-01000-00 EJ-1A001478  0631F-01000-00 EJ-1A001479  0631F-01000-00 EJ-1A001440  0631F-01000-00 EJ-1C001585  0631F-01000-00 EJ-1A001473  0631F-01000-00 EJ-1A001438  0631F-01000-00 EJ-1A001442  0631F-01000-00 EJ-1A001456  0631F-01000-00 EJ-1A001474  0631F-01000-00 EJ-1A001463  0631F-01000-00 EJ-1B001559  0631F-01000-00 EJ-1B001564  0631F-01000-00 EJ-1B001496  0631F-01000-00 EJ-1A001446  0631F-01000-00 EJ-23001757  0631F-01000-00 EJ-1A001421  0631F-01000-00 EJ-1A001424  0631F-01000-00 EJ-23001753  0631F-01000-00 EJ-1A001462  0631F-01000-00 EJ-1A001480  0631F-01000-00 EJ-1A001484  0631F-01000-00 EJ-21001657  0631F-01000-00 EJ-21001662  0631F-01000-00 EJ-21001670  0631F-01000-00 EJ-22001675  0631F-01000-00 EJ-22001676  0631F-01000-00 EJ-22001689  0631F-01000-00 EJ-22001694  0631F-01000-00 EJ-22001697  0631F-01000-00 EJ-22001702  0631F-01000-00 EJ-22001708  0631F-01000-00 EJ-22001713  0631F-01000-00 EJ-22001716  0631F-01000-00 EJ-22001717  0631F-01000-00 EJ-22001718  0631F-01000-00 EJ-22001720  0631F-01000-00 EJ-22001722  0631F-01000-00 EJ-22001727  0631F-01000-00 EJ-22001736  0631F-01000-00 EJ-22001740  0631F-01000-00 EJ-23001762  0631F-01000-00 EJ-23001778  0631F-01000-00 EJ-23001783  0631F-01000-00 EJ-1B001514  0631F-01000-00 EJ-1B001518  0631F-01000-00 EJ-1B001523  0631F-01000-00 EJ-1B001526  0631F-01000-00 EJ-1C001597  0631F-01000-00 EJ-1C001606  0631F-01000-00 EJ-23001750  0631F-01000-00 EJ-23001767  0631F-01000-00 EJ-23001772  0631F-01000-00 EJ-23001777  0631F-01000-00 EJ-22001732  0631F-01000-00 EJ-22001745  0631F-01000-00 EJ-23001766  0631F-01000-00 EJ-23001773  0631F-01000-00 EJ-23001780  0631F-01000-00 EJ-23001782  0631F-01000-00 EJ-23001786  0631F-01000-00 EJ-21001656  0631F-01000-00 EJ-21001660  0631F-01000-00 EJ-23001755  0631F-01000-00 EJ-23001761  0631F-01000-00 EJ-1A001435  0631F-01000-00 EJ-22001682  0631F-01000-00 EJ-22001730  0631F-01000-00 EJ-22001737  0631F-01000-00 EJ-23001747  0631F-01000-00 EJ-23001751  0631F-01000-00 EJ-23001756  0631F-01000-00 EJ-1A001447  0631F-01000-00 EJ-1A001450  0631F-01000-00 EJ-1A001455  0631F-01000-00 EJ-22001711  0631F-01000-00 EJ-1A001434  0631F-01000-00 EJ-1B001532  0631F-01000-00 EJ-1B001533  0631F-01000-00 EJ-1A001419  0631F-01000-00 EJ-1B001493  0631F-01000-00 EJ-1B001507  0631F-01000-00 EJ-1B001528  0631F-01000-00 EJ-1A001444  0631F-01000-00 EJ-1B001502  0631F-01000-00 EJ-1A001453  0631F-01000-00 EJ-1A001467  0631F-01000-00 EJ-1B001498  0631F-01000-00 EJ-1B001506  0631F-01000-00 EJ-23001749  0631F-01000-00 EJ-23001759  0631F-01000-00 EJ-23001763  0631F-01000-00 EJ-23001771  0631F-01000-00 EJ-23001784  0631F-01000-00 EJ-1C001574  0631F-01000-00 EJ-1C001580  0631F-01000-00 EJ-1C001588  0631F-01000-00 EJ-1C001591  0631F-01000-00 EJ-1C001594  0631F-01000-00 EJ-1C001598  0631F-01000-00 EJ-22001714  0631F-01000-00 EJ-23001774  0631F-01000-00 EJ-1A001412  0631F-01000-00 EJ-1A001422  0631F-01000-00 EJ-1A001425  0631F-01000-00 EJ-1C001570  0631F-01000-00 EJ-1C001611  0631F-01000-00 EJ-1C001612  0631F-01000-00 EJ-1B001555  0631F-01000-00 EJ-1B001556  0631F-01000-00 EJ-1C001566  0631F-01000-00 EJ-1C001567  0631F-01000-00 EJ-1C001613  0631F-01000-00 EJ-1B001541  0631F-01000-00 EJ-1C001571  0631F-01000-00 EJ-1C001578  0631F-01000-00 EJ-1C001586  0631F-01000-00 EJ-1C001596  0631F-01000-00 EJ-1C001607  0631F-01000-00 EJ-1C001609  0631F-01000-00 EJ-1C001614  0631F-01000-00 EJ-1C001615  0631F-01000-00 EJ-1C001616  0631F-01000-00 EJ-1C001618  0631F-01000-00 EJ-1C001620  0631F-01000-00 EJ-1C001621  0631F-01000-00 EJ-1C001623  0631F-01000-00 EJ-1C001626  0631F-01000-00 EJ-1C001628  0631F-01000-00 EJ-1C001629  0631F-01000-00 EJ-1C001630  0631F-01000-00 EJ-21001669  0631F-01000-00 EJ-21001671  0631F-01000-00 EJ-21001672  0631F-01000-00 EJ-1A001430  0631F-01000-00 EJ-1A001441  0631F-01000-00 EJ-23001770  0631F-01000-00 EJ-23001785  0631F-01000-00 EJ-22001729  0631F-01000-00 EJ-22001739  0631F-01000-00 EJ-23001746  0631F-01000-00 EJ-23001748  0631F-01000-00 EJ-23001754  0631F-01000-00 EJ-23001758  0631F-01000-00 EJ-1A001413  0631F-01000-00 EJ-1A001417  0631F-01000-00 EJ-1A001458  0631F-01000-00 EJ-1A001485  0631F-01000-00 EJ-1B001520  0631F-01000-00 EJ-21001645  0631F-01000-00 EJ-21001663  0631F-01000-00 EJ-21001665  0631F-01000-00 EJ-21001666  0631F-01000-00 EJ-1B001491  0631F-01000-00 EJ-1B001494  0631F-01000-00 EJ-1B001503  0631F-01000-00 EJ-1B001508  0631F-01000-00 EJ-1B001513  0631F-01000-00 EJ-1B001522  0631F-01000-00 EJ-1B001525  0631F-01000-00 EJ-21001649  0631F-01000-00 EJ-22001680  0631F-01000-00 EJ-22001715  0631F-01000-00 EJ-1A001464  0631F-01000-00 EJ-21001673  0631F-01000-00 EJ-21001674  0631F-01000-00 EJ-1B001529  0631F-01000-00 EJ-1C001627  0631F-01000-00 EJ-21001638  0631F-01000-00 EJ-21001661  0631F-01000-00 EJ-21001664  0631F-01000-00 EJ-21001668  0631F-01000-00 EJ-1A001472  0631F-01000-00 EJ-1A001486  0631F-01000-00 EJ-1B001490  0631F-01000-00 EJ-1B001495  0631F-01000-00 EJ-1A001448  0631F-01000-00 EJ-1A001452  0631F-01000-00 EJ-1A001454  0631F-01000-00 EJ-1A001466  0631F-01000-00 EJ-1A001477  0631F-01000-00 EJ-1B001534  0631F-01000-00 EJ-1B001547  0631F-01000-00 EJ-1C001575  0631F-01000-00 EJ-1A001414  0631F-01000-00 EJ-1A001429  0631F-01000-00 EJ-1B001515  0631F-01000-00 EJ-1B001517  0631F-01000-00 EJ-1B001545  0631F-01000-00 EJ-1B001489  0631F-01000-00 EJ-1B001492  0631F-01000-00 EJ-1B001504  0631F-01000-00 EJ-1B001531  0631F-01000-00 EJ-1B001521  0631F-01000-00 EJ-1A001415  0631F-01000-00 EJ-1B001510  0631F-01000-00 EJ-1B001536  0631F-01000-00 EJ-1A001457  0631F-01000-00 EJ-1A001468  0631F-01000-00 EJ-1B001557  0631F-01000-00 EJ-1B001558  0631F-01000-00 EJ-1B001560  0631F-01000-00 EJ-1C001587  0631F-01000-00 EJ-1C001595  0631F-01000-00 EJ-1B001563  0631F-01000-00 EJ-1C001568  0631F-01000-00 EJ-1C001579  0631F-01000-00 EJ-22001687  0631F-01000-00 EJ-1A001481  0631F-01000-00 EJ-1A001451  0631F-01000-00 EJ-1A001460  0631F-01000-00 EJ-1A001461  0631F-01000-00 EJ-1A001470  0631F-01000-00 EJ-1A001428  0631F-01000-00 EJ-1A001433  0631F-01000-00 EJ-1B001501  0631F-01000-00 EJ-1B001505  0631F-01000-00 EJ-1B001509  0631F-01000-00 EJ-22001691  0631F-01000-00 EJ-22001695  0631F-01000-00 EJ-22001706  0631F-01000-00 EJ-22001719  0631F-01000-00 EJ-22001726  0631F-01000-00 EJ-22001728  0631F-01000-00 EJ-22001731  0631F-01000-00 EJ-22001733  0631F-01000-00 EJ-22001735  0631F-01000-00 EJ-22001738"
    }
  ]
}