      "status": "Terminated",
      "city": "Athens",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83633",
      "recalling_firm": "Argon Medical Devices, Inc",
      "address_1": "1445 Flat Creek Rd",
      "address_2": "N/A",
      "postal_code": "75751-5002",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was nationwide.  There was military/government distribution.      Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.",
      "recall_number": "Z-0290-2020",
      "product_description": "(1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A, 25/box, Sterile, Rx;    (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A, 25/box, Sterile, Rx;    (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A, 25/box, Sterile, Rx;     (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A, Sterile, Rx;     (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A, 25/box, Sterile, Rx; and    (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590, 25/box, Sterile, Rx.    The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.",
      "product_quantity": "Total number in recall for all products in dist: 49,130 units",
      "reason_for_recall": "The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.",
      "recall_initiation_date": "20181016",
      "center_classification_date": "20191105",
      "termination_date": "20210615",
      "report_date": "20191113",
      "code_info": "(1) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Male Luer Lock, REF 041210002A-Lot numbers 11226656 and 11228887;    (2) Argon Medical Devices Stopcocks, 1-Way w/Male Luer Lock, REF 041216000A - Lot numbers 11226187, 11229619, and 11227967;    (3) Argon Medical Devices Stopcocks, Double 4-Way Stopcock w/Male Luer Lock, REF 041220001A - Lot numbers 11228888, 11230598;    (4) Argon Medical Devices Stopcocks, Clear 4-Way Stopcock w/Rotating Male Luer Lock, REF 041240002A - Lot number 11228245;    (5) Argon Medical Devices Stopcocks, North Carolina Baptist Hospital, Winston-Salem, NC, REF 042597300A - Lot number 11232206; and    (6) Argon Medical Devices 3 Way Stopcock w/Rotating Adapter (1050psi), REF 196590 - Lot numbers 11229800 and 11232485."
    }
  ]
}