  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63314",
      "recalling_firm": "Mindray DS USA, Inc. d.b.a. Mindray North America",
      "address_1": "800 MacArthur Blvd.",
      "address_2": "N/A",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Australia, Chile, Colombia, and Venezuela.",
      "recall_number": "Z-0290-2013",
      "product_description": "Mindray A3 Anesthesia Delivery System   Manufactured in China for Mindray DS USA Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430      This is a device used to administer to a patient continuously or intermittently a general inhalation anesthetic and to maintain a patient's ventilation.",
      "product_quantity": "69 units (61 US - 8 Foreign)",
      "reason_for_recall": "Mindray has identified the potential for a system leak resulting an issue related to the carbon dioxide absorber canister gasket on A3/A5 Anesthesia Delivery Systems. A potential system leak resulting from the canister gasket will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case.",
      "recall_initiation_date": "20120813",
      "center_classification_date": "20121116",
      "termination_date": "20130604",
      "report_date": "20121128",
      "code_info": "0633F-01000-00  Various serial numbers    0633F-01000-00 FA-1A000033  0633F-01000-00 FA-1A000023  0633F-01000-00 FA-21000072  0633F-01000-00 FA-1A000029  0633F-01000-00 FA-21000085  0633F-01000-00 FA-19000011  0633F-01000-00 FA-19000008  0633F-01000-00 FA-19000009  0633F-01000-00 FA-1A000027  0633F-01000-00 FA-1A000047  0633F-01000-00 FA-1A000055  0633F-01000-00 FA-1A000041  0633F-01000-00 FA-1A000053  0633F-01000-00 FA-1B000056  0633F-01000-00 FA-21000080  0633F-01000-00 FA-19000007  0633F-01000-00 FA-19000013  0633F-01000-00 FA-21000066  0633F-01000-00 FA-21000078  0633F-01000-00 FA-19000012  0633F-01000-00 FA-19000017  0633F-01000-00 FA-1A000021  0633F-01000-00 FA-1A000028  0633F-01000-00 FA-1A000038  0633F-01000-00 FA-1A000045  0633F-01000-00 FA-1A000035  0633F-01000-00 FA-1A000037  0633F-01000-00 FA-1A000026  0633F-01000-00 FA-1A000030  0633F-01000-00 FA-1A000048  0633F-01000-00 FA-1A000040  0633F-01000-00 FA-1B000059  0633F-01000-00 FA-1A000034  0633F-01000-00 FA-1A000049  0633F-01000-00 FA-21000062  0633F-01000-00 FA-21000064  0633F-01000-00 FA-21000074  0633F-01000-00 FA-21000082  0633F-01000-00 FA-21000090  0633F-01000-00 FA-1A000022  0633F-01000-00 FA-1A000051  0633F-01000-00 FA-1A000052  0633F-01000-00 FA-1A000042  0633F-01000-00 FA-1A000050  0633F-01000-00 FA-1B000057  0633F-01000-00 FA-1B000061  0633F-01000-00 FA-1A000025  0633F-01000-00 FA-1A000031  0633F-01000-00 FA-1A000036  0633F-01000-00 FA-1A000044  0633F-01000-00 FA-21000076  0633F-01000-00 FA-21000077  0633F-01000-00 FA-21000088  0633F-01000-00 FA-1A000043  0633F-01000-00 FA-1A000032  0633F-01000-00 FA-1A000046  0633F-01000-00 FA-1B000058  0633F-01000-00 FA-1B000060  0633F-01000-00 FA-19000010  0633F-01000-00 FA-19000015  0633F-01000-00 FA-21000086"
    }
  ]
}