{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken",
      "address_2": "",
      "product_quantity": "7,692 units",
      "code_info": "1) List  Number 12618; Serial Numbers: 14040001 ,  14040002 ,  14040003 ,  14040004 ,  14040005 ,  14040006 ,  14040008 ,  14040009 ,  14040010 ,  14040011 ,  14040012 ,  14040014 ,  14040015 ,  14040016 ,  14040017 ,  14040018 ,  14040019 ,  14040020 ,  14040021 ,  14040022 ,  14040023 ,  14040024 ,  14040025 ,  14040026 ,  14040027 ,  14040027 ,  14040027 ,  14040028 ,  14040028 ,  14040028 ,  14040029 ,  14040030 ,  14040031 ,  14040033 ,  14040034 ,  14040034 ,  14040034 ,  14040035 ,  14040036 ,  14040037 ,  14040038 ,  14040039 ,  14040040 ,  14040041 ,  14040042 ,  14040043 ,  14040044 ,  14040045 ,  14040047 ,  14040048 ,  14040049 ,  14040050 ,  14040051 ,  14040052 ,  14040053 ,  14040054 ,  14040056 ,  14040059 ,  14040061 ,  14040062 ,  14040063 ,  14040064 ,  14040065 ,  14040066 ,  14040067 ,  14040068 ,  14040069 ,  14040070 ,  14040071 ,  14040072 ,  14040073 ,  14040074 ,  14040075 ,  14040076 ,  14040077 ,  14040078 ,  14040079 ,  14040080 ,  14040081 ,  14040082 ,  14040083 ,  14040084 ,  14040085 ,  14040086 ,  14040087 ,  14040088 ,  14040089 ,  14040090 ,  14040091 ,  14040092 ,  14040093 ,  14040094 ,  14040095 ,  14040096 ,  14040097 ,  14040098 ,  14040099 ,  14040100 ,  14040101 ,  14040103 ,  14040104 ,  14040105 ,  14040106 ,  14040107 ,  14040108 ,  14040109 ,  14040110 ,  14040111 ,  14040112 ,  14040113 ,  14040115 ,  14040116 ,  14040117 ,  14040118 ,  14040119 ,  14040120 ,  14040121 ,  14040122 ,  14040123 ,  14040124 ,  14040125 ,  14040128 ,  14040129 ,  14040130 ,  14040131 ,  14040132 ,  14040133 ,  14040134 ,  14040135 ,  14040136 ,  14040138 ,  14040139 ,  14040140 ,  14040141 ,  14040142 ,  14040143 ,  14040144 ,  14040145 ,  14040146 ,  14040148 ,  14040149 ,  14040150 ,  14040152 ,  14040153 ,  14040154 ,  14040155 ,  14040156 ,  14040156 ,  14040156 ,  14040157 ,  14040158 ,  14040159 ,  14040160 ,  14040161 ,  14040162 ,  14040163 ,  14040164 ,  14040165 ,  14040166 ,  14040167 ,  14040168 ,  14040169 ,  14040170 ,  14040171 ,  14040172 ,  14040173 ,  14040174 ,  14040175 ,  14040176 ,  14040178 ,  14040179 ,  14040180 ,  14040181 ,  14040181 ,  14040181 ,  14040182 ,  14040183 ,  14040184 ,  14040185 ,  14040186 ,  14040187 ,  14040189 ,  14040190 ,  14040191 ,  14040192 ,  14040193 ,  14040194 ,  14040196 ,  14040198 ,  14040199 ,  14040200 ,  14040201 ,  14040202 ,  14040203 ,  14040205 ,  14040206 ,  14040207 ,  14040208 ,  14040209 ,  14040210 ,  14040211 ,  14040212 ,  14040213 ,  14040214 ,  14040215 ,  14040216 ,  14040217 ,  14040219 ,  14040220 ,  14040221 ,  14040223 ,  14040224 ,  14040225 ,  14040226 ,  14040227 ,  14040228 ,  14040229 ,  14040230 ,  14040231 ,  14040232 ,  14040233 ,  14040234 ,  14040235 ,  14040236 ,  14040238 ,  14040239 ,  14040240 ,  14040242 ,  14040243 ,  14040244 ,  14040245 ,  14040246 ,  14040247 ,  14040248 ,  14040249 ,  14040250 ,  14040251 ,  14040252 ,  14040253 ,  14040254 ,  14040255 ,  14040256 ,  14040257 ,  14040258 ,  14040260 ,  14040261 ,  14040262 ,  14040263 ,  14040264 ,  14040265 ,  14040266 ,  14040267 ,  14040269 ,  14040270 ,  14040271 ,  14040272 ,  14040274 ,  14040275 ,  14040276 ,  14040277 ,  14040278 ,  14040279 ,  14040280 ,  14040282 ,  14040283 ,  14040284 ,  14040285 ,  14040286 ,  14040287 ,  14040288 ,  14040289 ,  14040290 ,  14040291 ,  14040292 ,  14040293 ,  14040294 ,  14040295 ,  14040296 ,  14040297 ,  14040298 ,  14040299 ,  14040300 ,  14040301 ,  14040302 ,  14040303 ,  14040304 ,  14040305 ,  14040306 ,  14040307 ,  14040308 ,  14040309 ,  14040310 ,  14040310 ,  14040310 ,  14040311 ,  14040312 ,  14040313 ,  14040314 ,  14040315 ,  14040316 ,  14040317 ,  14040318 ,  14040319 ,  14040320 ,  14040321 ,  14040323 ,  14040324 ,  14040326 ,  14040327 ,  14040328 ,  14040329 ,  14040330 ,  14040331 ,  14040332 ,  14040333 ,  14040334 ,  14040335 ,  14040336 ,  14040337 ,  14040338 ,  14040339 ,  14040340 ,  14040341 ,  14040342 ,  14040343 ,  14040344 ,  14040345 ,  14040347 ,  14040348 ,  14040349 ,  14040350 ,  14040351 ,  14040352 ,  14040353 ,  14040354 ,  14040357 ,  14040359 ,  14040360 ,  14040361 ,  14040362 ,  14040363 ,  14040364 ,  14040365 ,  14040366 ,  14040367 ,  14040368 ,  14040369 ,  14040371 ,  14040372 ,  14040373 ,  14040374 ,  14040375 ,  14040376 ,  14040377 ,  14040379 ,  14040380 ,  14040382 ,  14040383 ,  14040384 ,  14040386 ,  14040387 ,  14040388 ,  14040389 ,  14040390 ,  14040391 ,  14040393 ,  14040394 ,  14040396 ,  14040397 ,  14040398 ,  14040399 ,  14040400 ,  14040401 ,  14040403 ,  14040404 ,  14040405 ,  14040406 ,  14040407 ,  14040408 ,  14040409 ,  14040411 ,  14040414 ,  14040415 ,  14040416 ,  14040417 ,  14040418 ,  14040419 ,  14040420 ,  14040421 ,  14040422 ,  14040423 ,  14040424 ,  14040425 ,  14040426 ,  14040427 ,  14040429 ,  14040430 ,  14040431 ,  14040432 ,  14040433 ,  14040434 ,  14040435 ,  14040437 ,  14040438 ,  14040440 ,  14040441 ,  14040442 ,  14040443 ,  14040444 ,  14040445 ,  14040446 ,  14040447 ,  14040448 ,  14040449 ,  14040453 ,  14040454 ,  14040455 ,  14040457 ,  14040458 ,  14040459 ,  14040460 ,  14040461 ,  14040462 ,  14040463 ,  14040464 ,  14040465 ,  14040466 ,  14040467 ,  14040468 ,  14040469 ,  14040470 ,  14040471 ,  14040472 ,  14040473 ,  14040474 ,  14040475 ,  14040476 ,  14040477 ,  14040478 ,  14040479 ,  14040480 ,  14040481 ,  14040482 ,  14040483 ,  14040484 ,  14040485 ,  14040486 ,  14040487 ,  14040488 ,  14040490 ,  14040491 ,  14040495 ,  14040496 ,  14040497 ,  14040498 ,  14040499 ,  14040500 ,  14040501 ,  14040502 ,  14040503 ,  14040504 ,  14040505 ,  14040506 ,  14040507 ,  14040508 ,  14040509 ,  14040510 ,  14040512 ,  14040513 ,  14040514 ,  14040515 ,  14040516 ,  14040517 ,  14040518 ,  14040519 ,  14040520 ,  14040522 ,  14040523 ,  14040524 ,  14040525 ,  14040526 ,  14040527 ,  14040528 ,  14040529 ,  14040530 ,  14040531 ,  14040532 ,  14040533 ,  14040534 ,  14040535 ,  14040536 ,  14040538 ,  14040539 ,  14040540 ,  14040541 ,  14040542 ,  14040543 ,  14040544 ,  14040545 ,  14040546 ,  14040547 ,  14040548 ,  14040549 ,  14040550 ,  14040551 ,  14040552 ,  14040553 ,  14040554 ,  14040556 ,  14040557 ,  14040558 ,  14040559 ,  14040560 ,  14040561 ,  14040562 ,  14040564 ,  14040565 ,  14040566 ,  14040567 ,  14040568 ,  14040569 ,  14040570 ,  14040571 ,  14040572 ,  14040573 ,  14040574 ,  14040575 ,  14040576 ,  14040577 ,  14040578 ,  14040579 ,  14040580 ,  14040581 ,  14040582 ,  14040583 ,  14040584 ,  14040585 ,  14040586 ,  14040587 ,  14040588 ,  14040589 ,  14040590 ,  14040591 ,  14040592 ,  14040593 ,  14040594 ,  14040595 ,  14040596 ,  14040597 ,  14040598 ,  14040599 ,  14040600 ,  14040601 ,  14040602 ,  14040603 ,  14040604 ,  14040605 ,  14040606 ,  14040607 ,  14040608 ,  14040609 ,  14040610 ,  14040611 ,  14040612 ,  14040613 ,  14040614 ,  14040615 ,  14040616 ,  14040617 ,  14040618 ,  14040620 ,  14040621 ,  14040623 ,  14040624 ,  14040625 ,  14040626 ,  14040627 ,  14040628 ,  14040629 ,  14040630 ,  14040631 ,  14040632 ,  14040633 ,  14040634 ,  14040635 ,  14040636 ,  14040637 ,  14040638 ,  14040639 ,  14040640 ,  14040641 ,  14040643 ,  14040644 ,  14040645 ,  14040646 ,  14040647 ,  14040648 ,  14040649 ,  14040650 ,  14040651 ,  14040652 ,  14040653 ,  14040654 ,  14040655 ,  14040656 ,  14040657 ,  14040658 ,  14040659 ,  14040660 ,  14040661 ,  14040662 ,  14040663 ,  14040664 ,  14040665 ,  14040666 ,  14040667 ,  14040668 ,  14040669 ,  14040670 ,  14040671 ,  14040672 ,  14040673 ,  14040674 ,  14040675 ,  14040676 ,  14040677 ,  14040678 ,  14040679 ,  14040680 ,  14040681 ,  14040682 ,  14040683 ,  14040684 ,  14040685 ,  14040686 ,  14040687 ,  14040688 ,  14040689 ,  14040690 ,  14040691 ,  14040692 ,  14040693 ,  14040694 ,  14040695 ,  14040696 ,  14040697 ,  14040698 ,  14040699 ,  14040700 ,  14040701 ,  14040702 ,  14040703 ,  14040704 ,  14040705 ,  14040706 ,  14040707 ,  14040708 ,  14040709 ,  14040710 ,  14040711 ,  14040712 ,  14040713 ,  14040714 ,  14040715 ,  14040716 ,  14040717 ,  14040718 ,  14040719 ,  14040720 ,  14040721 ,  14040722 ,  14040723 ,  14040724 ,  14040725 ,  14040726 ,  14040727 ,  14040728 ,  14040729 ,  14040730 ,  14040731 ,  14040732 ,  14040733 ,  14040734 ,  14040735 ,  14040736 ,  14040737 ,  14040738 ,  14040739 ,  14040740 ,  14040741 ,  14040742 ,  14040743 ,  14040744 ,  14040745 ,  14040746 ,  14040747 ,  14040748 ,  14040749 ,  14040750 ,  14040751 ,  14040752 ,  14040753 ,  14040755 ,  14040756 ,  14040757 ,  14040758 ,  14040759 ,  14040760 ,  14040762 ,  14040763 ,  14040764 ,  14040765 ,  14040766 ,  14040767 ,  14040768 ,  14040769 ,  14040771 ,  14040772 ,  14040773 ,  14040774 ,  14040775 ,  14040776 ,  14040777 ,  14040778 ,  14040779 ,  14040780 ,  14040781 ,  14040782 ,  14040783 ,  14040784 ,  14040785 ,  14040786 ,  14040787 ,  14040788 ,  14040790 ,  14040791 ,  14040792 ,  14040794 ,  14040797 ,  14040798 ,  14040803 ,  14040805 ,  14040806 ,  14040808 ,  14040810 ,  14040811 ,  14040812 ,  14040814 ,  14040816 ,  14040819 ,  14040820 ,  14040821 ,  14040828 ,  14040830 ,  14040831 ,  14040832 ,  14040834 ,  14040835 ,  14040839 ,  14040841 ,  14040842 ,  14040845 ,  14040847 ,  14040848 ,  14040849 ,  14040852 ,  14040853 ,  14040854 ,  14040855 ,  14040856 ,  14040857 ,  14040858 ,  14040859 ,  14040860 ,  14040861 ,  14040862 ,  14040863 ,  14040864 ,  14040865 ,  14040866 ,  14040868 ,  14040869 ,  14040870 ,  14040871 ,  14040873 ,  14040874 ,  14040875 ,  14040876 ,  14040877 ,  14040878 ,  14040879 ,  14040880 ,  14040881 ,  14040882 ,  14040883 ,  14040884 ,  14040885 ,  14040886 ,  14040887 ,  14040888 ,  14040889 ,  14040890 ,  14040891 ,  14040892 ,  14040893 ,  14040894 ,  14040895 ,  14040896 ,  14040897 ,  14040898 ,  14040899 ,  14040901 ,  14040902 ,  14040903 ,  14040904 ,  14040905 ,  14040906 ,  14040907 ,  14040908 ,  14040909 ,  14040910 ,  14040911 ,  14040913 ,  14040914 ,  14040915 ,  14040916 ,  14040917 ,  14040918 ,  14040919 ,  14040920 ,  14040921 ,  14040923 ,  14040924 ,  14040925 ,  14040926 ,  14040927 ,  14040928 ,  14040929 ,  14040930 ,  14040931 ,  14040933 ,  14040934 ,  14040935 ,  14040936 ,  14040937 ,  14040938 ,  14040939 ,  14040940 ,  14040941 ,  14040942 ,  14040943 ,  14040944 ,  14040945 ,  14040946 ,  14040947 ,  14040948 ,  14040949 ,  14040950 ,  14040951 ,  14040952 ,  14040953 ,  14040955 ,  14040956 ,  14040957 ,  14040958 ,  14040959 ,  14040960 ,  14040961 ,  14040963 ,  14040964 ,  14040965 ,  14040966 ,  14040968 ,  14040969 ,  14040970 ,  14040971 ,  14040972 ,  14040973 ,  14040974 ,  14040975 ,  14040976 ,  14040977 ,  14040978 ,  14040979 ,  14040981 ,  14040982 ,  14040983 ,  14040984 ,  14040985 ,  14040986 ,  14040987 ,  14040988 ,  14040989 ,  14040990 ,  14040991 ,  14040992 ,  14040993 ,  14040994 ,  14040995 ,  14040996 ,  14040997 ,  14040998 ,  14040999 ,  14041000 ,  14041001 ,  14041002 ,  14041003 ,  14041004 ,  14041005 ,  14041006 ,  14041007 ,  14041008 ,  14041009 ,  14041010 ,  14041011 ,  14041012 ,  14041013 ,  14041014 ,  14041015 ,  14041016 ,  14041017 ,  14041018 ,  14041019 ,  14041020 ,  14041022 ,  14041023 ,  14041024 ,  14041025 ,  14041026 ,  14041027 ,  14041028 ,  14041029 ,  14041030 ,  14041031 ,  14041032 ,  14041033 ,  14041034 ,  14041035 ,  14041036 ,  14041037 ,  14041038 ,  14041039 ,  14041040 ,  14041041 ,  14041042 ,  14041043 ,  14041044 ,  14041045 ,  14041046 ,  14041047 ,  14041048 ,  14041049 ,  14041050 ,  14041051 ,  14041052 ,  14041053 ,  14041054 ,  14041055 ,  14041056 ,  14041057 ,  14041058 ,  14041059 ,  14041060 ,  14041061 ,  14041062 ,  14041063 ,  14041064 ,  14041065 ,  14041067 ,  14041068 ,  14041069 ,  14041070 ,  14041071 ,  14041072 ,  14041074 ,  14041075 ,  14041076 ,  14041077 ,  14041078 ,  14041079 ,  14041080 ,  14041081 ,  14041082 ,  14041083 ,  14041084 ,  14041085 ,  14041086 ,  14041088 ,  14041089 ,  14041091 ,  14041092 ,  14041093 ,  14041094 ,  14041095 ,  14041097 ,  14041098 ,  14041099 ,  14041100 ,  14041101 ,  14041102 ,  14041103 ,  14041105 ,  14041107 ,  14041108 ,  14041109 ,  14041110 ,  14041111 ,  14041112 ,  14041113 ,  14041114 ,  14041115 ,  14041116 ,  14041117 ,  14041118 ,  14041119 ,  14041120 ,  14041121 ,  14041123 ,  14041124 ,  14041125 ,  14041126 ,  14041127 ,  14041128 ,  14041129 ,  14041130 ,  14041132 ,  14041133 ,  14041134 ,  14041135 ,  14041136 ,  14041137 ,  14041138 ,  14041141 ,  14041144 ,  14041146 ,  14041148 ,  14041150 ,  14041154 ,  14041155 ,  14041157 ,  14041164 ,  14041167 ,  14041171 ,  14041172 ,  14041174 ,  14041177 ,  14041178 ,  14041179 ,  14041180 ,  14041181 ,  14041182 ,  14041183 ,  14041184 ,  14041186 ,  14041187 ,  14041189 ,  14041190 ,  14041191 ,  14041192 ,  14041193 ,  14041194 ,  14041195 ,  14041196 ,  14041197 ,  14041198 ,  14041199 ,  14041200 ,  14041201 ,  14041203 ,  14041204 ,  14041205 ,  14041206 ,  14041207 ,  14041209 ,  14041211 ,  14041212 ,  14041215 ,  14041216 ,  14041217 ,  14041218 ,  14041220 ,  14041221 ,  14041223 ,  14041224 ,  14041225 ,  14041226 ,  14041227 ,  14041228 ,  14041229 ,  14041230 ,  14041231 ,  14041232 ,  14041233 ,  14041234 ,  14041235 ,  14041236 ,  14041237 ,  14041238 ,  14041239 ,  14041240 ,  14041241 ,  14041242 ,  14041243 ,  14041244 ,  14041245 ,  14041246 ,  14041247 ,  14041248 ,  14041249 ,  14041250 ,  14041251 ,  14041252 ,  14041253 ,  14041254 ,  14041255 ,  14041256 ,  14041257 ,  14041258 ,  14041259 ,  14041260 ,  14041261 ,  14041262 ,  14041263 ,  14041264 ,  14041265 ,  14041266 ,  14041267 ,  14041269 ,  14041270 ,  14041272 ,  14041273 ,  14041274 ,  14041275 ,  14041276 ,  14041277 ,  14041278 ,  14041279 ,  14041280 ,  14041281 ,  14041282 ,  14041283 ,  14041284 ,  14041285 ,  14041286 ,  14041288 ,  14041289 ,  14041290 ,  14041291 ,  14041292 ,  14041293 ,  14041294 ,  14041295 ,  14041296 ,  14041297 ,  14041298 ,  14041299 ,  14041300 ,  14041301 ,  14041302 ,  14041303 ,  14041304 ,  14041305 ,  14041306 ,  14041307 ,  14041308 ,  14041309 ,  14041310 ,  14041311 ,  14041312 ,  14041313 ,  14041314 ,  14041315 ,  14041316 ,  14041317 ,  14041318 ,  14041319 ,  14041320 ,  14041321 ,  14041322 ,  14041324 ,  14041325 ,  14041327 ,  14041328 ,  14041329 ,  14041330 ,  14041331 ,  14041332 ,  14041333 ,  14041334 ,  14041335 ,  14041337 ,  14041338 ,  14041339 ,  14041341 ,  14041342 ,  14041343 ,  14041344 ,  14041345 ,  14041346 ,  14041347 ,  14041348 ,  14041349 ,  14041350 ,  14041351 ,  14041352 ,  14041353 ,  14041354 ,  14041355 ,  14041356 ,  14041357 ,  14041358 ,  14041359 ,  14041360 ,  14041361 ,  14041362 ,  14041363 ,  14041364 ,  14041365 ,  14041366 ,  14041367 ,  14041369 ,  14041370 ,  14041371 ,  14041372 ,  14041373 ,  14041374 ,  14041375 ,  14041376 ,  14041377 ,  14041378 ,  14041379 ,  14041380 ,  14041381 ,  14041382 ,  14041383 ,  14041384 ,  14041385 ,  14041386 ,  14041387 ,  14041388 ,  14041389 ,  14041390 ,  14041391 ,  14041392 ,  14041393 ,  14041394 ,  14041395 ,  14041396 ,  14041397 ,  14041398 ,  14041399 ,  14041400 ,  14041401 ,  14041402 ,  14041403 ,  14041404 ,  14041405 ,  14041406 ,  14041407 ,  14041408 ,  14041409 ,  14041410 ,  14041411 ,  14041412 ,  14041413 ,  14041414 ,  14041415 ,  14041416 ,  14041417 ,  14041418 ,  14041418 ,  14041418 ,  14041419 ,  14041421 ,  14041422 ,  14041423 ,  14041424 ,  14041425 ,  14041426 ,  14041427 ,  14041428 ,  14041429 ,  14041430 ,  14041431 ,  14041432 ,  14041433 ,  14041434 ,  14041435 ,  14041437 ,  14041438 ,  14041439 ,  14041440 ,  14041441 ,  14041442 ,  14041443 ,  14041444 ,  14041445 ,  14041446 ,  14041447 ,  14041448 ,  14041449 ,  14041450 ,  14041451 ,  14041451 ,  14041451 ,  14041452 ,  14041453 ,  14041454 ,  14041455 ,  14041455 ,  14041455 ,  14041456 ,  14041457 ,  14041458 ,  14041459 ,  14041460 ,  14041461 ,  14041462 ,  14041463 ,  14041464 ,  14041466 ,  14041467 ,  14041468 ,  14041469 ,  14041470 ,  14041472 ,  14041473 ,  14041474 ,  14041475 ,  14041476 ,  14041477 ,  14041478 ,  14041479 ,  14041480 ,  14041481 ,  14041482 ,  14041483 ,  14041484 ,  14041485 ,  14041486 ,  14041487 ,  14041488 ,  14041489 ,  14041490 ,  14041491 ,  14041492 ,  14041493 ,  14041494 ,  14041495 ,  14041496 ,  14041497 ,  14041498 ,  14041499 ,  14041500 ,  14041501 ,  14041502 ,  14041503 ,  14041504 ,  14041505 ,  14041506 ,  14041507 ,  14041508 ,  14041509 ,  14041510 ,  14041511 ,  14041512 ,  14041513 ,  14041514 ,  14041515 ,  14041516 ,  14041517 ,  14041518 ,  14041519 ,  14041520 ,  14041521 ,  14041522 ,  14041523 ,  14041525 ,  14041526 ,  14041527 ,  14041528 ,  14041529 ,  14041530 ,  14041531 ,  14041532 ,  14041533 ,  14041534 ,  14041535 ,  14041536 ,  14041537 ,  14041538 ,  14041539 ,  14041540 ,  14041541 ,  14041542 ,  14041544 ,  14041545 ,  14041546 ,  14041547 ,  14041548 ,  14041549 ,  14041550 ,  14041551 ,  14041552 ,  14041553 ,  14041554 ,  14041555 ,  14041556 ,  14041557 ,  14041558 ,  14041559 ,  14041560 ,  14041561 ,  14041562 ,  14041563 ,  14041564 ,  14041565 ,  14041569 ,  14041574 ,  14041576 ,  14041580 ,  14041581 ,  14041583 ,  14041584 ,  14041585 ,  14041586 ,  14041588 ,  14041589 ,  14041590 ,  14041591 ,  14041592 ,  14041593 ,  14041594 ,  14041595 ,  14041596 ,  14041597 ,  14041598 ,  14041599 ,  14041601 ,  14041602 ,  14041603 ,  14041604 ,  14041605 ,  14041606 ,  14041607 ,  14041608 ,  14041609 ,  14041610 ,  14041611 ,  14041612 ,  14041613 ,  14041614 ,  14041615 ,  14041616 ,  14041617 ,  14041618 ,  14041619 ,  14041620 ,  14041621 ,  14041622 ,  14041623 ,  14041624 ,  14041625 ,  14041626 ,  14041627 ,  14041628 ,  14041629 ,  14041630 ,  14041632 ,  14041633 ,  14041634 ,  14041635 ,  14041637 ,  14041638 ,  14041641 ,  14041642 ,  14041643 ,  14041644 ,  14041645 ,  14041646 ,  14041647 ,  14041648 ,  14041650 ,  14041651 ,  14041652 ,  14041653 ,  14041654 ,  14041655 ,  14041656 ,  14041657 ,  14041661 ,  14041662 ,  14041663 ,  14041664 ,  14041665 ,  14041666 ,  14041667 ,  14041668 ,  14041669 ,  14041670 ,  14041672 ,  14041673 ,  14041674 ,  14041675 ,  14041676 ,  14041677 ,  14041679 ,  14041680 ,  14041681 ,  14041682 ,  14041683 ,  14041684 ,  14041685 ,  14041686 ,  14041687 ,  14041688 ,  14041689 ,  14041690 ,  14041691 ,  14041692 ,  14041693 ,  14041694 ,  14041695 ,  14041696 ,  14041698 ,  14041699 ,  14041700 ,  14041701 ,  14041702 ,  14041703 ,  14041704 ,  14041706 ,  14041707 ,  14041708 ,  14041709 ,  14041710 ,  14041711 ,  14041712 ,  14041713 ,  14041714 ,  14041715 ,  14041716 ,  14041717 ,  14041718 ,  14041719 ,  14041720 ,  14041721 ,  14041722 ,  14041723 ,  14041724 ,  14041725 ,  14041726 ,  14041727 ,  14041728 ,  14041729 ,  14041730 ,  14041731 ,  14041732 ,  14041733 ,  14041734 ,  14041735 ,  14041736 ,  14041737 ,  14041738 ,  14041739 ,  14041740 ,  14041741 ,  14041742 ,  14041743 ,  14041744 ,  14041746 ,  14041747 ,  14041748 ,  14041749 ,  14041750 ,  14041752 ,  14041753 ,  14041755 ,  14041756 ,  14041757 ,  14041758 ,  14041759 ,  14041760 ,  14041761 ,  14041762 ,  14041763 ,  14041764 ,  14041765 ,  14041766 ,  14041767 ,  14041768 ,  14041769 ,  14041770 ,  14041771 ,  14041772 ,  14041773 ,  14041774 ,  14041775 ,  14041776 ,  14041777 ,  14041778 ,  14041779 ,  14041780 ,  14041783 ,  14041784 ,  14041785 ,  14041786 ,  14041787 ,  14041788 ,  14041789 ,  14041791 ,  14041792 ,  14041793 ,  14041794 ,  14041795 ,  14041796 ,  14041797 ,  14041800 ,  14041803 ,  14041804 ,  14041805 ,  14041806 ,  14041807 ,  14041808 ,  14041809 ,  14041811 ,  14041814 ,  14041815 ,  14041816 ,  14041817 ,  14041818 ,  14041819 ,  14041820 ,  14041821 ,  14041822 ,  14041823 ,  14041824 ,  14041825 ,  14041826 ,  14041827 ,  14041829 ,  14041830 ,  14041831 ,  14041833 ,  14041834 ,  14041836 ,  14041837 ,  14041838 ,  14041839 ,  14041840 ,  14041842 ,  14041843 ,  14041844 ,  14041845 ,  14041846 ,  14041847 ,  14041848 ,  14041849 ,  14041850 ,  14041851 ,  14041852 ,  14041854 ,  14041856 ,  14041857 ,  14041859 ,  14041860 ,  14041861 ,  14041862 ,  14041863 ,  14041866 ,  14041869 ,  14041872 ,  14041873 ,  14041874 ,  14041875 ,  14041876 ,  14041877 ,  14041878 ,  14041879 ,  14041880 ,  14041881 ,  14041882 ,  14041883 ,  14041884 ,  14041885 ,  14041886 ,  14041887 ,  14041890 ,  14041891 ,  14041893 ,  14041894 ,  14041895 ,  14041896 ,  14041897 ,  14041899 ,  14041900 ,  14041901 ,  14041902 ,  14041904 ,  14041905 ,  14041906 ,  14041907 ,  14041908 ,  14041909 ,  14041910 ,  14041911 ,  14041912 ,  14041913 ,  14041914 ,  14041915 ,  14041916 ,  14041917 ,  14041918 ,  14041919 ,  14041920 ,  14041921 ,  14041922 ,  14041923 ,  14041924 ,  14041926 ,  14041927 ,  14041928 ,  14041929 ,  14041930 ,  14041931 ,  14041932 ,  14041933 ,  14041934 ,  14041935 ,  14041936 ,  14041937 ,  14041938 ,  14041939 ,  14041940 ,  14041941 ,  14041944 ,  14041945 ,  14041946 ,  14041947 ,  14041948 ,  14041949 ,  14041950 ,  14041952 ,  14041953 ,  14041954 ,  14041955 ,  14041956 ,  14041957 ,  14041959 ,  14041960 ,  14041961 ,  14041962 ,  14041963 ,  14041964 ,  14041965 ,  14041966 ,  14041967 ,  14041968 ,  14041969 ,  14041970 ,  14041971 ,  14041972 ,  14041973 ,  14041974 ,  14041975 ,  14041976 ,  14041977 ,  14041978 ,  14041979 ,  14041980 ,  14041981 ,  14041982 ,  14041983 ,  14041984 ,  14041985 ,  14041986 ,  14041987 ,  14041988 ,  14041989 ,  14041990 ,  14041993 ,  14041994 ,  14041995 ,  14041996 ,  14041997 ,  14041998 ,  14041999 ,  14042000 ,  14042001 ,  14042002 ,  14042003 ,  14042004 ,  14042005 ,  14042006 ,  14042007 ,  14042008 ,  14042009 ,  14042010 ,  14042011 ,  14042012 ,  14042013 ,  14042015 ,  14042016 ,  14042017 ,  14042018 ,  14042019 ,  14042021 ,  14042022 ,  14042023 ,  14042024 ,  14042026 ,  14042027 ,  14042028 ,  14042029 ,  14042030 ,  14042031 ,  14042032 ,  14042033 ,  14042034 ,  14042035 ,  14042036 ,  14042037 ,  14042038 ,  14042039 ,  14042040 ,  14042041 ,  14042042 ,  14042043 ,  14042044 ,  14042045 ,  14042046 ,  14042047 ,  14042049 ,  14042052 ,  14042053 ,  14042055 ,  14042056 ,  14042057 ,  14042058 ,  14042059 ,  14042060 ,  14042062 ,  14042063 ,  14042064 ,  14042065 ,  14042066 ,  14042067 ,  14042068 ,  14042069 ,  14042070 ,  14042072 ,  14042073 ,  14042074 ,  14042076 ,  14042077 ,  14042078 ,  14042079 ,  14042081 ,  14042082 ,  14042083 ,  14042084 ,  14042086 ,  14042087 ,  14042088 ,  14042089 ,  14042090 ,  14042091 ,  14042092 ,  14042093 ,  14042094 ,  14042095 ,  14042096 ,  14042097 ,  14042098 ,  14042099 ,  14042100 ,  14042101 ,  14042102 ,  14042104 ,  14042105 ,  14042106 ,  14042107 ,  14042108 ,  14042109 ,  14042110 ,  14042111 ,  14042112 ,  14042113 ,  14042115 ,  14042116 ,  14042117 ,  14042118 ,  14042119 ,  14042120 ,  14042121 ,  14042123 ,  14042124 ,  14042125 ,  14042126 ,  14042127 ,  14042128 ,  14042130 ,  14042132 ,  14042133 ,  14042134 ,  14042135 ,  14042136 ,  14042138 ,  14042139 ,  14042141 ,  14042142 ,  14042143 ,  14042147 ,  14042148 ,  14042149 ,  14042151 ,  14042152 ,  14042154 ,  14042155 ,  14042156 ,  14042157 ,  14042158 ,  14042160 ,  14042161 ,  14042162 ,  14042163 ,  14042164 ,  14042165 ,  14042167 ,  14042171 ,  14042172 ,  14042173 ,  14042174 ,  14042175 ,  14042176 ,  14042177 ,  14042178 ,  14042179 ,  14042180 ,  14042181 ,  14042182 ,  14042183 ,  14042185 ,  14042186 ,  14042187 ,  14042188 ,  14042190 ,  14042192 ,  14042194 ,  14042195 ,  14042196 ,  14042199 ,  14042201 ,  14042202 ,  14042204 ,  14042205 ,  14042208 ,  14042210 ,  14042211 ,  14042212 ,  14042213 ,  14042214 ,  14042215 ,  14042218 ,  14042219 ,  14042220 ,  14042221 ,  14042222 ,  14042224 ,  14042227 ,  14042228 ,  14042230 ,  14042231 ,  14042232 ,  14042233 ,  14042234 ,  14042236 ,  14042237 ,  14042238 ,  14042239 ,  14042243 ,  14042244 ,  14042245 ,  14042246 ,  14042248 ,  14042249 ,  14042250 ,  14042251 ,  14042253 ,  14042254 ,  14042255 ,  14042257 ,  14042260 ,  14042261 ,  14042262 ,  14042263 ,  14042265 ,  14042266 ,  14042267 ,  14042269 ,  14042270 ,  14042273 ,  14042275 ,  14042278 ,  14042280 ,  14042281 ,  14042282 ,  14042284 ,  14042285 ,  14042289 ,  14042293 ,  14042295 ,  14042296 ,  14042298 ,  14042299 ,  14042301 ,  14042309 ,  14042310 ,  14042311 ,  14042312 ,  14042313 ,  14042315 ,  14042316 ,  14042320 ,  14042323 ,  14042324 ,  14042326 ,  14042329 ,  14042331 ,  14042332 ,  14042336 ,  14042340 ,  14042341 ,  14042342 ,  14042346 ,  14042347 ,  14042350 ,  14042353 ,  14042355 ,  14042356 ,  14042358 ,  14042359 ,  14042360 ,  14042361 ,  14042362 ,  14042363 ,  14042364 ,  14042365 ,  14042366 ,  14042367 ,  14042369 ,  14042370 ,  14042371 ,  14042373 ,  14042374 ,  14042375 ,  14042376 ,  14042377 ,  14042378 ,  14042379 ,  14042380 ,  14042381 ,  14042382 ,  14042383 ,  14042384 ,  14042385 ,  14042386 ,  14042387 ,  14042388 ,  14042390 ,  14042391 ,  14042392 ,  14042393 ,  14042394 ,  14042396 ,  14042399 ,  14042400 ,  14042401 ,  14042402 ,  14042403 ,  14042404 ,  14042405 ,  14042406 ,  14042407 ,  14042408 ,  14042410 ,  14042411 ,  14042414 ,  14042415 ,  14042416 ,  14042417 ,  14042420 ,  14042421 ,  14042422 ,  14042424 ,  14042425 ,  14042426 ,  14042428 ,  14042429 ,  14042430 ,  14042432 ,  14042434 ,  14042436 ,  14042437 ,  14042438 ,  14042439 ,  14042442 ,  14042443 ,  14042444 ,  14042446 ,  14042447 ,  14042448 ,  14042450 ,  14042451 ,  14042451 ,  14042452 ,  14042452 ,  14042453 ,  14042454 ,  14042454 ,  14042455 ,  14042455 ,  14042459 ,  14042461 ,  14042462 ,  14042463 ,  14042464 ,  14042465 ,  14042466 ,  14042467 ,  14042468 ,  14042469 ,  14042470 ,  14042471 ,  14042472 ,  14042474 ,  14042475 ,  14042477 ,  14042478 ,  14042479 ,  14042481 ,  14042482 ,  14042483 ,  14042484 ,  14042485 ,  14042486 ,  14042487 ,  14042488 ,  14042490 ,  14042491 ,  14042492 ,  14042493 ,  14042494 ,  14042495 ,  14042496 ,  14042497 ,  14042498 ,  14042499 ,  14042500 ,  14042501 ,  14042502 ,  14042503 ,  14042504 ,  14042505 ,  14042506 ,  14042507 ,  14042508 ,  14042509 ,  14042510 ,  14042511 ,  14042512 ,  14042513 ,  14042514 ,  14042515 ,  14042516 ,  14042517 ,  14042518 ,  14042519 ,  14042520 ,  14042521 ,  14042522 ,  14042523 ,  14042524 ,  14042527 ,  14042528 ,  14042529 ,  14042530 ,  14042531 ,  14042532 ,  14042533 ,  14042534 ,  14042535 ,  14042536 ,  14042537 ,  14042538 ,  14042539 ,  14042540 ,  14042541 ,  14042542 ,  14042543 ,  14042544 ,  14042545 ,  14042546 ,  14042547 ,  14042548 ,  14042549 ,  14042550 ,  14042551 ,  14042552 ,  14042553 ,  14042554 ,  14042555 ,  14042556 ,  14042557 ,  14042558 ,  14042559 ,  14042560 ,  14042561 ,  14042562 ,  14042564 ,  14042565 ,  14042566 ,  14042567 ,  14042568 ,  14042569 ,  14042570 ,  14042571 ,  14042572 ,  14042573 ,  14042574 ,  14042575 ,  14042576 ,  14042577 ,  14042578 ,  14042579 ,  14042580 ,  14042581 ,  14042582 ,  14042583 ,  14042584 ,  14042586 ,  14042587 ,  14042588 ,  14042590 ,  14042591 ,  14042592 ,  14042593 ,  14042594 ,  14042595 ,  14042596 ,  14042597 ,  14042598 ,  14042599 ,  14042600 ,  14042601 ,  14042602 ,  14042603 ,  14042604 ,  14042605 ,  14042606 ,  14042607 ,  14042609 ,  14042610 ,  14042611 ,  14042612 ,  14042613 ,  14042614 ,  14042616 ,  14042617 ,  14042618 ,  14042619 ,  14042620 ,  14042621 ,  14042622 ,  14042623 ,  14042625 ,  14042626 ,  14042627 ,  14042628 ,  14042628 ,  14042628 ,  14042629 ,  14042630 ,  14042631 ,  14042632 ,  14042633 ,  14042634 ,  14042635 ,  14042636 ,  14042637 ,  14042638 ,  14042639 ,  14042640 ,  14042642 ,  14042643 ,  14042644 ,  14042645 ,  14042646 ,  14042647 ,  14042648 ,  14042649 ,  14042650 ,  14042651 ,  14042652 ,  14042653 ,  14042654 ,  14042655 ,  14042656 ,  14042659 ,  14042660 ,  14042661 ,  14042662 ,  14042663 ,  14042664 ,  14042666 ,  14042667 ,  14042668 ,  14042669 ,  14042670 ,  14042671 ,  14042672 ,  14042673 ,  14042674 ,  14042677 ,  14042678 ,  14042679 ,  14042680 ,  14042681 ,  14042682 ,  14042683 ,  14042684 ,  14042685 ,  14042686 ,  14042687 ,  14042688 ,  14042689 ,  14042690 ,  14042691 ,  14042692 ,  14042693 ,  14042694 ,  14042695 ,  14042696 ,  14042697 ,  14042698 ,  14042699 ,  14042700 ,  14042702 ,  14042703 ,  14042704 ,  14042705 ,  14042707 ,  14042708 ,  14042709 ,  14042710 ,  14042711 ,  14042712 ,  14042713 ,  14042714 ,  14042715 ,  14042717 ,  14042718 ,  14042719 ,  14042720 ,  14042722 ,  14042723 ,  14042724 ,  14042725 ,  14042726 ,  14042727 ,  14042728 ,  14042729 ,  14042730 ,  14042731 ,  14042732 ,  14042733 ,  14042734 ,  14042736 ,  14042737 ,  14042739 ,  14042740 ,  14042741 ,  14042742 ,  14042745 ,  14042746 ,  14042747 ,  14042749 ,  14042750 ,  14042751 ,  14042752 ,  14042755 ,  14042757 ,  14042758 ,  14042759 ,  14042760 ,  14042761 ,  14042763 ,  14042764 ,  14042767 ,  14042770 ,  14042771 ,  14042772 ,  14042774 ,  14042775 ,  14042776 ,  14042777 ,  14042780 ,  14042781 ,  14042783 ,  14042784 ,  14042785 ,  14042786 ,  14042788 ,  14042789 ,  14042790 ,  14042793 ,  14042794 ,  14042795 ,  14042798 ,  14042799 ,  14042800 ,  14042802 ,  14042803 ,  14042804 ,  14042805 ,  14042806 ,  14042807 ,  14042808 ,  14042809 ,  14042810 ,  14042811 ,  14042812 ,  14042813 ,  14042814 ,  14042815 ,  14042816 ,  14042817 ,  14042818 ,  14042819 ,  14042820 ,  14042821 ,  14042822 ,  14042823 ,  14042824 ,  14042825 ,  14042826 ,  14042827 ,  14042828 ,  14042829 ,  14042830 ,  14042831 ,  14042832 ,  14042833 ,  14042834 ,  14042835 ,  14042836 ,  14042837 ,  14042838 ,  14042839 ,  14042840 ,  14042842 ,  14042843 ,  14042844 ,  14042845 ,  14042846 ,  14042847 ,  14042848 ,  14042850 ,  14042851 ,  14042852 ,  14042853 ,  14042854 ,  14042855 ,  14042856 ,  14042857 ,  14042858 ,  14042859 ,  14042861 ,  14042862 ,  14042863 ,  14042864 ,  14042865 ,  14042866 ,  14042867 ,  14042868 ,  14042869 ,  14042870 ,  14042871 ,  14042873 ,  14042874 ,  14042875 ,  14042876 ,  14042877 ,  14042878 ,  14042879 ,  14042880 ,  14042881 ,  14042882 ,  14042883 ,  14042884 ,  14042885 ,  14042886 ,  14042887 ,  14042888 ,  14042889 ,  14042890 ,  14042891 ,  14042892 ,  14042893 ,  14042894 ,  14042895 ,  14042896 ,  14042897 ,  14042898 ,  14042899 ,  14042900 ,  14042902 ,  14042903 ,  14042904 ,  14042905 ,  14042906 ,  14042907 ,  14042908 ,  14042909 ,  14042910 ,  14042911 ,  14042912 ,  14042913 ,  14042915 ,  14042918 ,  14042919 ,  14042920 ,  14042921 ,  14042922 ,  14042923 ,  14042924 ,  14042924 ,  14042925 ,  14042927 ,  14042928 ,  14042929 ,  14042929 ,  14042930 ,  14042931 ,  14042932 ,  14042933 ,  14042935 ,  14042936 ,  14042937 ,  14042938 ,  14042939 ,  14042940 ,  14042941 ,  14042942 ,  14042943 ,  14042944 ,  14042945 ,  14042946 ,  14042947 ,  14042948 ,  14042949 ,  14042950 ,  14042951 ,  14042952 ,  14042953 ,  14042954 ,  14042955 ,  14042956 ,  14042957 ,  14042958 ,  14042959 ,  14042960 ,  14042961 ,  14042962 ,  14042963 ,  14042964 ,  14042965 ,  14042966 ,  14042967 ,  14042968 ,  14042969 ,  14042970 ,  14042971 ,  14042972 ,  14042973 ,  14042974 ,  14042975 ,  14042976 ,  14042977 ,  14042979 ,  14042980 ,  14042981 ,  14042982 ,  14042983 ,  14042984 ,  14042985 ,  14042986 ,  14042987 ,  14042988 ,  14042989 ,  14042990 ,  14042990 ,  14042991 ,  14042992 ,  14042993 ,  14042994 ,  14042995 ,  14042996 ,  14042997 ,  14042998 ,  14042999 ,  14043000 ,  14043001 ,  14043002 ,  14043003 ,  14043004 ,  14043005 ,  14043006 ,  14043008 ,  14043009 ,  14043010 ,  14043012 ,  14043013 ,  14043014 ,  14043015 ,  14043016 ,  14043017 ,  14043019 ,  14043020 ,  14043021 ,  14043022 ,  14043023 ,  14043024 ,  14043025 ,  14043026 ,  14043027 ,  14043028 ,  14043029 ,  14043030 ,  14043031 ,  14043032 ,  14043033 ,  14043034 ,  14043035 ,  14043036 ,  14043037 ,  14043038 ,  14043039 ,  14043040 ,  14043041 ,  14043042 ,  14043043 ,  14043044 ,  14043045 ,  14043046 ,  14043047 ,  14043048 ,  14043049 ,  14043050 ,  14043051 ,  14043052 ,  14043053 ,  14043054 ,  14043055 ,  14043056 ,  14043057 ,  14043058 ,  14043059 ,  14043060 ,  14043061 ,  14043062 ,  14043064 ,  14043065 ,  14043066 ,  14043067 ,  14043068 ,  14043069 ,  14043070 ,  14043071 ,  14043072 ,  14043073 ,  14043074 ,  14043075 ,  14043076 ,  14043077 ,  14043078 ,  14043079 ,  14043080 ,  14043081 ,  14043082 ,  14043083 ,  14043084 ,  14043085 ,  14043086 ,  14043087 ,  14043088 ,  14043089 ,  14043090 ,  14043091 ,  14043092 ,  14043093 ,  14043094 ,  14043095 ,  14043096 ,  14043097 ,  14043098 ,  14043101 ,  14043102 ,  14043104 ,  14043105 ,  14043106 ,  14043107 ,  14043108 ,  14043",
      "center_classification_date": "20131114",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands;  and countries of: Costa Rica and Dominican Republic.",
      "state": "IL",
      "product_description": "The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3 Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+3 Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator's manual.  The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-0288-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64263",
      "termination_date": "20161206",
      "more_code_info": "109 ,  14043110 ,  14043111 ,  14043112 ,  14043113 ,  14043116 ,  14043117 ,  14043120 ,  14043122 ,  14043124 ,  14043125 ,  14043126 ,  14043129 ,  14043130 ,  14043132 ,  14043133 ,  14043134 ,  14043135 ,  14043137 ,  14043138 ,  14043139 ,  14043140 ,  14043141 ,  14043142 ,  14043144 ,  14043145 ,  14043146 ,  14043149 ,  14043150 ,  14043154 ,  14043156 ,  14043157 ,  14043158 ,  14043159 ,  14043160 ,  14043161 ,  14043162 ,  14043163 ,  14043165 ,  14043166 ,  14043167 ,  14043168 ,  14043170 ,  14043172 ,  14043173 ,  14043174 ,  14043176 ,  14043180 ,  14043182 ,  14043183 ,  14043184 ,  14043186 ,  14043187 ,  14043188 ,  14043190 ,  14043193 ,  14043195 ,  14043197 ,  14043202 ,  14043203 ,  14043204 ,  14043206 ,  14043207 ,  14043209 ,  14043211 ,  14043214 ,  14043215 ,  14043217 ,  14043218 ,  14043219 ,  14043221 ,  14043222 ,  14043223 ,  14043225 ,  14043226 ,  14043227 ,  14043228 ,  14043229 ,  14043230 ,  14043231 ,  14043233 ,  14043234 ,  14043235 ,  14043236 ,  14043237 ,  14043238 ,  14043240 ,  14043241 ,  14043242 ,  14043243 ,  14043244 ,  14043246 ,  14043247 ,  14043249 ,  14043250 ,  14043251 ,  14043252 ,  14043253 ,  14043254 ,  14043255 ,  14043256 ,  14043257 ,  14043258 ,  14043259 ,  14043260 ,  14043261 ,  14043262 ,  14043263 ,  14043264 ,  14043266 ,  14043267 ,  14043268 ,  14043269 ,  14043271 ,  14043272 ,  14043273 ,  14043275 ,  14043276 ,  14043277 ,  14043277 ,  14043278 ,  14043279 ,  14043280 ,  14043281 ,  14043284 ,  14043285 ,  14043286 ,  14043287 ,  14043288 ,  14043289 ,  14043290 ,  14043291 ,  14043292 ,  14043293 ,  14043294 ,  14043295 ,  14043296 ,  14043297 ,  14043298 ,  14043299 ,  14043300 ,  14043301 ,  14043302 ,  14043303 ,  14043304 ,  14043305 ,  14043306 ,  14043307 ,  14043308 ,  14043309 ,  14043310 ,  14043311 ,  14043312 ,  14043314 ,  14043315 ,  14043316 ,  14043317 ,  14043318 ,  14043319 ,  14043320 ,  14043321 ,  14043322 ,  14043323 ,  14043324 ,  14043325 ,  14043326 ,  14043327 ,  14043328 ,  14043329 ,  14043330 ,  14043331 ,  14043332 ,  14043334 ,  14043335 ,  14043336 ,  14043338 ,  14043340 ,  14043341 ,  14043342 ,  14043343 ,  14043344 ,  14043345 ,  14043346 ,  14043347 ,  14043348 ,  14043349 ,  14043350 ,  14043351 ,  14043352 ,  14043354 ,  14043355 ,  14043356 ,  14043357 ,  14043360 ,  14043361 ,  14043362 ,  14043363 ,  14043366 ,  14043368 ,  14043369 ,  14043370 ,  14043371 ,  14043372 ,  14043374 ,  14043375 ,  14043376 ,  14043377 ,  14043378 ,  14043380 ,  14043381 ,  14043382 ,  14043383 ,  14043385 ,  14043386 ,  14043387 ,  14043389 ,  14043391 ,  14043392 ,  14043393 ,  14043394 ,  14043396 ,  14043397 ,  14043398 ,  14043399 ,  14043402 ,  14043403 ,  14043404 ,  14043405 ,  14043406 ,  14043407 ,  14043408 ,  14043411 ,  14043416 ,  14043417 ,  14043418 ,  14043419 ,  14043420 ,  14043421 ,  14043423 ,  14043424 ,  14043425 ,  14043426 ,  14043427 ,  14043434 ,  14043438 ,  14043440 ,  14043441 ,  14043442 ,  14043443 ,  14043444 ,  14043445 ,  14043446 ,  14043447 ,  14043448 ,  14043450 ,  14043451 ,  14043452 ,  14043453 ,  14043454 ,  14043455 ,  14043458 ,  14043460 ,  14043461 ,  14043462 ,  14043463 ,  14043464 ,  14043465 ,  14043466 ,  14043467 ,  14043468 ,  14043470 ,  14043471 ,  14043472 ,  14043473 ,  14043474 ,  14043475 ,  14043476 ,  14043477 ,  14043478 ,  14043479 ,  14043480 ,  14043481 ,  14043482 ,  14043483 ,  14043484 ,  14043485 ,  14043486 ,  14043487 ,  14043488 ,  14043489 ,  14043490 ,  14043491 ,  14043492 ,  14043493 ,  14043494 ,  14043495 ,  14043496 ,  14043497 ,  14043498 ,  14043499 ,  14043500 ,  14043501 ,  14043502 ,  14043503 ,  14043504 ,  14043505 ,  14043506 ,  14043507 ,  14043508 ,  14043509 ,  14043510 ,  14043511 ,  14043512 ,  14043513 ,  14043514 ,  14043515 ,  14043516 ,  14043517 ,  14043518 ,  14043519 ,  14043520 ,  14043521 ,  14043522 ,  14043523 ,  14043524 ,  14043525 ,  14043526 ,  14043527 ,  14043528 ,  14043529 ,  14043530 ,  14043531 ,  14043533 ,  14043534 ,  14043535 ,  14043536 ,  14043537 ,  14043538 ,  14043539 ,  14043540 ,  14043541 ,  14043542 ,  14043543 ,  14043544 ,  14043545 ,  14043546 ,  14043547 ,  14043548 ,  14043549 ,  14043550 ,  14043551 ,  14043552 ,  14043553 ,  14043554 ,  14043555 ,  14043556 ,  14043557 ,  14043558 ,  14043559 ,  14043560 ,  14043561 ,  14043562 ,  14043563 ,  14043564 ,  14043565 ,  14043566 ,  14043567 ,  14043568 ,  14043569 ,  14043570 ,  14043571 ,  14043572 ,  14043573 ,  14043574 ,  14043575 ,  14043576 ,  14043577 ,  14043578 ,  14043579 ,  14043580 ,  14043581 ,  14043582 ,  14043583 ,  14043584 ,  14043585 ,  14043586 ,  14043587 ,  14043588 ,  14043589 ,  14043590 ,  14043591 ,  14043592 ,  14043593 ,  14043594 ,  14043595 ,  14043596 ,  14043597 ,  14043598 ,  14043599 ,  14043600 ,  14043601 ,  14043602 ,  14043603 ,  14043604 ,  14043605 ,  14043606 ,  14043607 ,  14043608 ,  14043609 ,  14043610 ,  14043611 ,  14043612 ,  14043613 ,  14043614 ,  14043615 ,  14043616 ,  14043617 ,  14043618 ,  14043619 ,  14043620 ,  14043621 ,  14043622 ,  14043623 ,  14043624 ,  14043625 ,  14043626 ,  14043627 ,  14043628 ,  14043629 ,  14043630 ,  14043631 ,  14043633 ,  14043634 ,  14043635 ,  14043636 ,  14043637 ,  14043638 ,  14043639 ,  14043640 ,  14043641 ,  14043642 ,  14043643 ,  14043644 ,  14043645 ,  14043646 ,  14043647 ,  14043648 ,  14043649 ,  14043650 ,  14043651 ,  14043652 ,  14043653 ,  14043654 ,  14043655 ,  14043656 ,  14043657 ,  14043658 ,  14043659 ,  14043660 ,  14043661 ,  14043662 ,  14043663 ,  14043664 ,  14043665 ,  14043666 ,  14043667 ,  14043668 ,  14043669 ,  14043670 ,  14043671 ,  14043672 ,  14043673 ,  14043674 ,  14043675 ,  14043676 ,  14043677 ,  14043678 ,  14043679 ,  14043680 ,  14043681 ,  14043682 ,  14043683 ,  14043684 ,  14043685 ,  14043686 ,  14043687 ,  14043688 ,  14043689 ,  14043690 ,  14043691 ,  14043692 ,  14043693 ,  14043694 ,  14043695 ,  14043696 ,  14043697 ,  14043698 ,  14043699 ,  14043700 ,  14043701 ,  14043702 ,  14043703 ,  14043704 ,  14043705 ,  14043706 ,  14043707 ,  14043708 ,  14043709 ,  14043710 ,  14043711 ,  14043712 ,  14043713 ,  14043714 ,  14043715 ,  14043716 ,  14043717 ,  14043718 ,  14043719 ,  14043720 ,  14043721 ,  14043722 ,  14043723 ,  14043724 ,  14043725 ,  14043726 ,  14043727 ,  14043728 ,  14043729 ,  14043730 ,  14043731 ,  14043732 ,  14043733 ,  14043734 ,  14043735 ,  14043736 ,  14043737 ,  14043738 ,  14043739 ,  14043740 ,  14043741 ,  14043742 ,  14043743 ,  14043744 ,  14043745 ,  14043746 ,  14043747 ,  14043748 ,  14043749 ,  14043750 ,  14043751 ,  14043752 ,  14043753 ,  14043754 ,  14043755 ,  14043756 ,  14043757 ,  14043758 ,  14043759 ,  14043760 ,  14043761 ,  14043762 ,  14043763 ,  14043764 ,  14043765 ,  14043766 ,  14043767 ,  14043768 ,  14043769 ,  14043770 ,  14043771 ,  14043772 ,  14043773 ,  14043774 ,  14043775 ,  14043776 ,  14043777 ,  14043778 ,  14043779 ,  14043780 ,  14043781 ,  14043782 ,  14043783 ,  14043784 ,  14043785 ,  14043786 ,  14043787 ,  14043788 ,  14043789 ,  14043790 ,  14043791 ,  14043792 ,  14043793 ,  14043794 ,  14043795 ,  14043796 ,  14043797 ,  14043798 ,  14043799 ,  14043800 ,  14043801 ,  14043802 ,  14043803 ,  14043804 ,  14043805 ,  14043806 ,  14043807 ,  14043808 ,  14043809 ,  14043810 ,  14043811 ,  14043812 ,  14043813 ,  14043814 ,  14043815 ,  14043816 ,  14043817 ,  14043818 ,  14043819 ,  14043820 ,  14043821 ,  14043822 ,  14043823 ,  14043824 ,  14043825 ,  14043826 ,  14043827 ,  14043828 ,  14043829 ,  14043830 ,  14043831 ,  14043832 ,  14043833 ,  14043834 ,  14043835 ,  14043836 ,  14043837 ,  14043838 ,  14043839 ,  14043840 ,  14043841 ,  14043842 ,  14043843 ,  14043844 ,  14043845 ,  14043846 ,  14043847 ,  14043848 ,  14043849 ,  14043850 ,  14043851 ,  14043852 ,  14043853 ,  14043854 ,  14043855 ,  14043856 ,  14043857 ,  14043858 ,  14043859 ,  14043860 ,  14043861 ,  14043862 ,  14043863 ,  14043864 ,  14043865 ,  14043866 ,  14043867 ,  14043868 ,  14043869 ,  14043870 ,  14043871 ,  14043872 ,  14043873 ,  14043874 ,  14043875 ,  14043876 ,  14043877 ,  14043878 ,  14043879 ,  14043880 ,  14043881 ,  14043882 ,  14043883 ,  14043884 ,  14043885 ,  14043886 ,  14043887 ,  14043888 ,  14043889 ,  14043890 ,  14043891 ,  14043892 ,  14043893 ,  14043894 ,  14043895 ,  14043896 ,  14043897 ,  14043898 ,  14043899 ,  14043900 ,  14043901 ,  14043902 ,  14043903 ,  14043904 ,  14043905 ,  14043906 ,  14043907 ,  14043908 ,  14043909 ,  14043910 ,  14043911 ,  14043912 ,  14043913 ,  14043914 ,  14043915 ,  14043916 ,  14043917 ,  14043918 ,  14043920 ,  14043921 ,  14043922 ,  14043923 ,  14043924 ,  14043925 ,  14043926 ,  14043927 ,  14043928 ,  14043929 ,  14043930 ,  14043931 ,  14043932 ,  14043933 ,  14043934 ,  14043935 ,  14043936 ,  14043937 ,  14043938 ,  14043939 ,  14043940 ,  14043941 ,  14043942 ,  14043943 ,  14043944 ,  14043945 ,  14043946 ,  14043947 ,  14043948 ,  14043949 ,  14043950 ,  14043951 ,  14043952 ,  14043953 ,  14043954 ,  14043956 ,  14043957 ,  14043958 ,  14043959 ,  14043960 ,  14043961 ,  14043962 ,  14043963 ,  14043964 ,  14043965 ,  14043966 ,  14043967 ,  14043968 ,  14043969 ,  14043970 ,  14043971 ,  14043972 ,  14043973 ,  14043974 ,  14043975 ,  14043976 ,  14043977 ,  14043978 ,  14043979 ,  14043980 ,  14043981 ,  14043982 ,  14043983 ,  14043984 ,  14043985 ,  14043986 ,  14043987 ,  14043988 ,  14043989 ,  14043990 ,  14043991 ,  14043992 ,  14043993 ,  14043994 ,  14043995 ,  14043996 ,  14043997 ,  14043998 ,  14043999 ,  14044000 ,  14044001 ,  14044002 ,  14044003 ,  14044004 ,  14044005 ,  14044006 ,  14044007 ,  14044009 ,  14044010 ,  14044011 ,  14044012 ,  14044013 ,  14044014 ,  14044015 ,  14044016 ,  14044017 ,  14044018 ,  14044019 ,  14044021 ,  14044022 ,  14044023 ,  14044024 ,  14044025 ,  14044026 ,  14044028 ,  14044029 ,  14044030 ,  14044031 ,  14044032 ,  14044033 ,  14044034 ,  14044035 ,  14044037 ,  14044038 ,  14044039 ,  14044040 ,  14044041 ,  14044042 ,  14044043 ,  14044044 ,  14044046 ,  14044047 ,  14044048 ,  14044054 ,  14044055 ,  14044056 ,  14044057 ,  14044058 ,  14044059 ,  14044060 ,  14044061 ,  14044062 ,  14044063 ,  14044064 ,  14044065 ,  14044066 ,  14044067 ,  14044068 ,  14044069 ,  14044072 ,  14044073 ,  14044074 ,  14044075 ,  14044076 ,  14044078 ,  14044079 ,  14044080 ,  14044081 ,  14044082 ,  14044083 ,  14044084 ,  14044085 ,  14044086 ,  14044087 ,  14044088 ,  14044089 ,  14044090 ,  14044091 ,  14044092 ,  14044093 ,  14044094 ,  14044095 ,  14044096 ,  14044097 ,  14044098 ,  14044099 ,  14044100 ,  14044101 ,  14044102 ,  14044103 ,  14044104 ,  14044105 ,  14044106 ,  14044107 ,  14044108 ,  14044109 ,  14044110 ,  14044111 ,  14044112 ,  14044113 ,  14044114 ,  14044115 ,  14044116 ,  14044117 ,  14044118 ,  14044119 ,  14044120 ,  14044121 ,  14044122 ,  14044123 ,  14044124 ,  14044125 ,  14044126 ,  14044127 ,  14044128 ,  14044129 ,  14044130 ,  14044131 ,  14044132 ,  14044134 ,  14044135 ,  14044136 ,  14044137 ,  14044138 ,  14044139 ,  14044140 ,  14044141 ,  14044142 ,  14044143 ,  14044144 ,  14044145 ,  14044146 ,  14044147 ,  14044148 ,  14044149 ,  14044150 ,  14044151 ,  14044152 ,  14044153 ,  14044154 ,  14044155 ,  14044156 ,  14044158 ,  14044159 ,  14044160 ,  14044161 ,  14044162 ,  14044163 ,  14044164 ,  14044165 ,  14044166 ,  14044167 ,  14044168 ,  14044169 ,  14044170 ,  14044171 ,  14044173 ,  14044175 ,  14044176 ,  14044177 ,  14044178 ,  14044179 ,  14044180 ,  14044181 ,  14044182 ,  14044183 ,  14044188 ,  14044190 ,  14044192 ,  14044193 ,  14044194 ,  14044195 ,  14044196 ,  14044197 ,  14044202 ,  14044205 ,  14044207 ,  14044208 ,  14044209 ,  14044210 ,  14044211 ,  14044213 ,  14044215 ,  14044216 ,  14044219 ,  14044220 ,  14044221 ,  14044223 ,  14044224 ,  14044225 ,  14044227 ,  14044228 ,  14044230 ,  14044231 ,  14044232 ,  14044236 ,  14044237 ,  14044239 ,  14044241 ,  14044242 ,  14044243 ,  14044443 ,  14044444 ,  14044445 ,  14044446 ,  14044447 ,  14044448 ,  14044449 ,  14044450 ,  14044451 ,  14044452 ,  14044453 ,  14044454 ,  14044455 ,  14044456 ,  14044457 ,  14044458 ,  14044459 ,  14044460 ,  14044461 ,  14044462 ,  14044463 ,  14044464 ,  14044465 ,  14044466 ,  14044467 ,  14044468 ,  14044469 ,  14044470 ,  14044472 ,  14044473 ,  14044474 ,  14044475 ,  14044476 ,  14044478 ,  14044479 ,  14044480 ,  14044481 ,  14044482 ,  14044483 ,  14044484 ,  14044485 ,  14044486 ,  14044487 ,  14044488 ,  14044489 ,  14044490 ,  14044491 ,  14044492 ,  14044493 ,  14044494 ,  14044495 ,  14044496 ,  14044497 ,  14044498 ,  14044499 ,  14044500 ,  14044501 ,  14044502 ,  14044503 ,  14044504 ,  14044506 ,  14044507 ,  14044508 ,  14044509 ,  14044510 ,  14044511 ,  14044512 ,  14044513 ,  14044514 ,  14044515 ,  14044516 ,  14044517 ,  14044518 ,  14044519 ,  14044520 ,  14044521 ,  14044522 ,  14044523 ,  14044524 ,  14044525 ,  14044526 ,  14044527 ,  14044528 ,  14044529 ,  14044530 ,  14044531 ,  14044532 ,  14044533 ,  14044534 ,  14044535 ,  14044536 ,  14044537 ,  14044538 ,  14044539 ,  14044540 ,  14044541 ,  14044542 ,  14044543 ,  14044544 ,  14044545 ,  14044546 ,  14044547 ,  14044548 ,  14044549 ,  14044550 ,  14044551 ,  14044552 ,  14044553 ,  14044554 ,  14044555 ,  14044556 ,  14044557 ,  14044558 ,  14044559 ,  14044560 ,  14044561 ,  14044562 ,  14044563 ,  14044564 ,  14044565 ,  14044566 ,  14044567 ,  14044568 ,  14044569 ,  14044570 ,  14044571 ,  14044572 ,  14044573 ,  14044574 ,  14044575 ,  14044576 ,  14044577 ,  14044578 ,  14044579 ,  14044580 ,  14044581 ,  14044582 ,  14044583 ,  14044584 ,  14044585 ,  14044586 ,  14044587 ,  14044588 ,  14044589 ,  14044590 ,  14044591 ,  14044592 ,  14044593 ,  14044594 ,  14044595 ,  14044596 ,  14044597 ,  14044598 ,  14044599 ,  14044600 ,  14044601 ,  14044602 ,  14044603 ,  14044614 ,  14044615 ,  14044617 ,  14044618 ,  14044619 ,  14044620 ,  14044621 ,  14044622 ,  14044623 ,  14044624 ,  14044625 ,  14044626 ,  14044627 ,  14044628 ,  14044629 ,  14044630 ,  14044631 ,  14044632 ,  14044633 ,  14044634 ,  14044635 ,  14044636 ,  14044637 ,  14044638 ,  14044639 ,  14044640 ,  14044641 ,  14044642 ,  14044643 ,  14044644 ,  14044645 ,  14044646 ,  14044647 ,  14044648 ,  14044649 ,  14044650 ,  14044651 ,  14044652 ,  14044653 ,  14044654 ,  14044655 ,  14044656 ,  14044657 ,  14044658 ,  14044659 ,  14044660 ,  14044661 ,  14044662 ,  14044663 ,  14044664 ,  14044665 ,  14044666 ,  14044667 ,  14044668 ,  14044669 ,  14044670 ,  14044671 ,  14044672 ,  14044673 ,  14044674 ,  14044675 ,  14044676 ,  14044677 ,  14044678 ,  14044679 ,  14044680 ,  14044681 ,  14044682 ,  14044683 ,  14044684 ,  14044685 ,  14044686 ,  14044687 ,  14044688 ,  14044689 ,  14044690 ,  14044691 ,  14044692 ,  14044693 ,  14044694 ,  14044695 ,  14044696 ,  14044697 ,  14044698 ,  14044699 ,  14044700 ,  14044701 ,  14044702 ,  14044703 ,  14044704 ,  14044705 ,  14044706 ,  14044707 ,  14044708 ,  14044709 ,  14044710 ,  14044711 ,  14044712 ,  14044713 ,  14044714 ,  14044715 ,  14044716 ,  14044717 ,  14044718 ,  14044719 ,  14044720 ,  14044721 ,  14044722 ,  14044723 ,  14044724 ,  14044725 ,  14044726 ,  14044727 ,  14044728 ,  14044729 ,  14044730 ,  14044731 ,  14044732 ,  14044733 ,  14044734 ,  14044735 ,  14044736 ,  14044738 ,  14044739 ,  14044740 ,  14044741 ,  14044742 ,  14044744 ,  14044745 ,  14044746 ,  14044747 ,  14044748 ,  14044749 ,  14044750 ,  14044751 ,  14044752 ,  14044753 ,  14044754 ,  14044755 ,  14044756 ,  14044757 ,  14044758 ,  14044759 ,  14044760 ,  14044761 ,  14044762 ,  14044763 ,  14044764 ,  14044765 ,  14044766 ,  14044767 ,  14044768 ,  14044769 ,  14044771 ,  14044773 ,  14044774 ,  14044775 ,  14044776 ,  14044778 ,  14044779 ,  14044780 ,  14044782 ,  14044783 ,  14044784 ,  14044785 ,  14044785 ,  14044786 ,  14044787 ,  14044788 ,  14044789 ,  14044790 ,  14044791 ,  14044792 ,  14044793 ,  14044794 ,  14044795 ,  14044796 ,  14044797 ,  14044798 ,  14044799 ,  14044800 ,  14044801 ,  14044802 ,  14044803 ,  14044804 ,  14044805 ,  14044806 ,  14044807 ,  14044808 ,  14044809 ,  14044810 ,  14044811 ,  14044813 ,  14044814 ,  14044815 ,  14044816 ,  14044818 ,  14044819 ,  14044820 ,  14044821 ,  14044822 ,  14044823 ,  14044824 ,  14044825 ,  14044826 ,  14044827 ,  14044828 ,  14044829 ,  14044830 ,  14044831 ,  14044832 ,  14044833 ,  14044834 ,  14044835 ,  14044836 ,  14044837 ,  14044838 ,  14044839 ,  14044840 ,  14044841 ,  14044842 ,  14044843 ,  14044844 ,  14044845 ,  14044846 ,  14044847 ,  14044848 ,  14044849 ,  14044850 ,  14044851 ,  14044852 ,  14044853 ,  14044854 ,  14044855 ,  14044856 ,  14044857 ,  14044858 ,  14044859 ,  14044860 ,  14044861 ,  14044862 ,  14044863 ,  14044864 ,  14044865 ,  14044866 ,  14044868 ,  14044869 ,  14044870 ,  14044871 ,  14044872 ,  14044873 ,  14044874 ,  14044876 ,  14044877 ,  14044878 ,  14044879 ,  14044880 ,  14044881 ,  14044882 ,  14044883 ,  14044884 ,  14044885 ,  14044886 ,  14044888 ,  14044889 ,  14044890 ,  14044891 ,  14044892 ,  14044893 ,  14044894 ,  14044895 ,  14044896 ,  14044897 ,  14044898 ,  14044899 ,  14044900 ,  14044901 ,  14044902 ,  14044903 ,  14044904 ,  14044905 ,  14044906 ,  14044907 ,  14044908 ,  14044909 ,  14044910 ,  14044911 ,  14044912 ,  14044913 ,  14044914 ,  14044915 ,  14044916 ,  14044917 ,  14044918 ,  14044919 ,  14044920 ,  14044921 ,  14044922 ,  14044924 ,  14044925 ,  14044926 ,  14044927 ,  14044928 ,  14044929 ,  14044930 ,  14044931 ,  14044933 ,  14044934 ,  14044935 ,  14044936 ,  14044937 ,  14044938 ,  14044939 ,  14044940 ,  14044941 ,  14044942 ,  14044943 ,  14044944 ,  14044945 ,  14044946 ,  14044947 ,  14044948 ,  14044949 ,  14044950 ,  14044951 ,  14044952 ,  14044953 ,  14044954 ,  14044955 ,  14044956 ,  14044957 ,  14044958 ,  14044959 ,  14044960 ,  14044961 ,  14044962 ,  14044964 ,  14044965 ,  14044966 ,  14044967 ,  14044968 ,  14044969 ,  14044970 ,  14044971 ,  14044973 ,  14044974 ,  14044975 ,  14044977 ,  14044980 ,  14044981 ,  14044982 ,  14044983 ,  14044984 ,  14044985 ,  14044986 ,  14044987 ,  14044988 ,  14044989 ,  14044990 ,  14044991 ,  14044992 ,  14044993 ,  14044994 ,  14044995 ,  14044996 ,  14044997 ,  14044998 ,  14044999 ,  14045000 ,  14045001 ,  14045002 ,  14045003 ,  14045004 ,  14045005 ,  14045006 ,  14045007 ,  14045008 ,  14045009 ,  14045010 ,  14045011 ,  14045012 ,  14045013 ,  14045014 ,  14045015 ,  14045016 ,  14045017 ,  14045018 ,  14045019 ,  14045020 ,  14045021 ,  14045022 ,  14045023 ,  14045024 ,  14045025 ,  14045025 ,  14045026 ,  14045027 ,  14045027 ,  14045028 ,  14045029 ,  14045030 ,  14045031 ,  14045032 ,  14045033 ,  14045034 ,  14045035 ,  14045036 ,  14045037 ,  14045038 ,  14045039 ,  14045040 ,  14045041 ,  14045042 ,  14045043 ,  14045044 ,  14045045 ,  14045046 ,  14045047 ,  14045048 ,  14045049 ,  14045050 ,  14045051 ,  14045052 ,  14045053 ,  14045054 ,  14045055 ,  14045056 ,  14045057 ,  14045058 ,  14045059 ,  14045060 ,  14045061 ,  14045062 ,  14045063 ,  14045064 ,  14045065 ,  14045066 ,  14045067 ,  14045068 ,  14045069 ,  14045070 ,  14045071 ,  14045073 ,  14045074 ,  14045075 ,  14045076 ,  14045077 ,  14045078 ,  14045079 ,  14045080 ,  14045081 ,  14045082 ,  14045083 ,  14045084 ,  14045085 ,  14045086 ,  14045087 ,  14045088 ,  14045089 ,  14045090 ,  14045091 ,  14045092 ,  14045093 ,  14045094 ,  14045095 ,  14045096 ,  14045097 ,  14045098 ,  14045099 ,  14045100 ,  14045101 ,  14045102 ,  14045103 ,  14045104 ,  14045105 ,  14045106 ,  14045107 ,  14045108 ,  14045109 ,  14045110 ,  14045111 ,  14045112 ,  14045113 ,  14045114 ,  14045115 ,  14045116 ,  14045117 ,  14045117 ,  14045118 ,  14045119 ,  14045120 ,  14045121 ,  14045121 ,  14045122 ,  14045123 ,  14045124 ,  14045125 ,  14045126 ,  14045127 ,  14045128 ,  14045129 ,  14045130 ,  14045131 ,  14045132 ,  14045133 ,  14045134 ,  14045135 ,  14045136 ,  14045137 ,  14045138 ,  14045139 ,  14045140 ,  14045141 ,  14045142 ,  14045143 ,  14045144 ,  14045145 ,  14045146 ,  14045147 ,  14045148 ,  14045150 ,  14045151 ,  14045152 ,  14045153 ,  14045154 ,  14045155 ,  14045156 ,  14045157 ,  14045158 ,  14045159 ,  14045160 ,  14045161 ,  14045162 ,  14045163 ,  14045164 ,  14045165 ,  14045166 ,  14045167 ,  14045168 ,  14045169 ,  14045170 ,  14045171 ,  14045172 ,  14045173 ,  14045174 ,  14045175 ,  14045176 ,  14045177 ,  14045178 ,  14045179 ,  14045180 ,  14045181 ,  14045182 ,  14045183 ,  14045184 ,  14045185 ,  14045186 ,  14045187 ,  14045188 ,  14045189 ,  14045190 ,  14045192 ,  14045193 ,  14045195 ,  14045196 ,  14045197 ,  14045198 ,  14045199 ,  14045200 ,  14045201 ,  14045202 ,  14045203 ,  14045204 ,  14045205 ,  14045206 ,  14045207 ,  14045208 ,  14045209 ,  14045210 ,  14045211 ,  14045212 ,  14045213 ,  14045214 ,  14045215 ,  14045216 ,  14045217 ,  14045218 ,  14045219 ,  14045220 ,  14045221 ,  14045222 ,  14045223 ,  14045224 ,  14045225 ,  14045227 ,  14045228 ,  14045230 ,  14045231 ,  14045232 ,  14045233 ,  14045234 ,  14045235 ,  14045236 ,  14045237 ,  14045238 ,  14045239 ,  14045240 ,  14045241 ,  14045242 ,  14045243 ,  14045245 ,  14045246 ,  14045247 ,  14045248 ,  14045249 ,  14045250 ,  14045251 ,  14045252 ,  14045253 ,  14045254 ,  14045255 ,  14045256 ,  14045257 ,  14045258 ,  14045259 ,  14045260 ,  14045261 ,  14045262 ,  14045263 ,  14045264 ,  14045265 ,  14045266 ,  14045267 ,  14045268 ,  14045269 ,  14045270 ,  14045271 ,  14045272 ,  14045273 ,  14045274 ,  14045275 ,  14045276 ,  14045277 ,  14045278 ,  14045279 ,  14045280 ,  14045281 ,  14045282 ,  14045283 ,  14045284 ,  14045285 ,  14045286 ,  14045287 ,  14045288 ,  14045289 ,  14045290 ,  14045291 ,  14045292 ,  14045293 ,  14045294 ,  14045295 ,  14045296 ,  14045298 ,  14045299 ,  14045300 ,  14045301 ,  14045302 ,  14045303 ,  14045304 ,  14045305 ,  14045306 ,  14045307 ,  14045308 ,  14045309 ,  14045310 ,  14045311 ,  14045312 ,  14045313 ,  14045314 ,  14045315 ,  14045316 ,  14045317 ,  14045318 ,  14045319 ,  14045320 ,  14045321 ,  14045322 ,  14045323 ,  14045324 ,  14045325 ,  14045326 ,  14045327 ,  14045328 ,  14045329 ,  14045330 ,  14045331 ,  14045332 ,  14045333 ,  14045334 ,  14045336 ,  14045337 ,  14045338 ,  14045339 ,  14045340 ,  14045341 ,  14045343 ,  14045344 ,  14045346 ,  14045347 ,  14045348 ,  14045349 ,  14045351 ,  14045352 ,  14045353 ,  14045354 ,  14045355 ,  14045356 ,  14045357 ,  14045358 ,  14045359 ,  14045360 ,  14045361 ,  14045362 ,  14045363 ,  14045364 ,  14045365 ,  14045366 ,  14045367 ,  14045368 ,  14045369 ,  14045370 ,  14045371 ,  14045372 ,  14045373 ,  14045374 ,  14045375 ,  14045376 ,  14045377 ,  14045378 ,  14045379 ,  14045380 ,  14045381 ,  14045382 ,  14045383 ,  14045384 ,  14045385 ,  14045386 ,  14045387 ,  14045388 ,  14045389 ,  14045390 ,  14045391 ,  14045392 ,  14045393 ,  14045394 ,  14045395 ,  14045396 ,  14045397 ,  14045398 ,  14045399 ,  14045400 ,  14045401 ,  14045402 ,  14045403 ,  14045404 ,  14045405 ,  14045406 ,  14045407 ,  14045408 ,  14045409 ,  14045410 ,  14045411 ,  14045412 ,  14045413 ,  14045414 ,  14045415 ,  14045416 ,  14045417 ,  14045418 ,  14045419 ,  14045420 ,  14045421 ,  14045422 ,  14045423 ,  14045424 ,  14045425 ,  14045426 ,  14045427 ,  14045428 ,  14045429 ,  14045430 ,  14045431 ,  14045432 ,  14045433 ,  14045434 ,  14045435 ,  14045436 ,  14045437 ,  14045438 ,  14045439 ,  14045440 ,  14045441 ,  14045442 ,  14045443 ,  14045444 ,  14045445 ,  14045446 ,  14045447 ,  14045448 ,  14045449 ,  14045450 ,  14045451 ,  14045452 ,  14045453 ,  14045454 ,  14045455 ,  14045456 ,  14045457 ,  14045458 ,  14045459 ,  14045460 ,  14045462 ,  14045465 ,  14045466 ,  14045467 ,  14045469 ,  14045470 ,  14045471 ,  14045472 ,  14045473 ,  14045474 ,  14045475 ,  14045480 ,  14045486 ,  14045488 ,  14045489 ,  14045490 ,  14045491 ,  14045492 ,  14045493 ,  14045494 ,  14045497 ,  14045499 ,  14045500 ,  14045501 ,  14045503 ,  14045504 ,  14045506 ,  14045508 ,  14045509 ,  14045510 ,  14045511 ,  14045512 ,  14045514 ,  14045515 ,  14045516 ,  14045517 ,  14045518 ,  14045519 ,  14045520 ,  14045521 ,  14045522 ,  14045523 ,  14045524 ,  14045525 ,  14045526 ,  14045527 ,  14045528 ,  14045529 ,  14045530 ,  14045531 ,  14045532 ,  14045533 ,  14045534 ,  14045535 ,  14045536 ,  14045537 ,  14045538 ,  14045541 ,  14045542 ,  14045543 ,  14045544 ,  14045545 ,  14045546 ,  14045547 ,  14045548 ,  14045549 ,  14045551 ,  14045552 ,  14045553 ,  14045554 ,  14045555 ,  14045556 ,  14045557 ,  14045558 ,  14045559 ,  14045560 ,  14045561 ,  14045562 ,  14045563 ,  14045564 ,  14045565 ,  14045566 ,  14045567 ,  14045568 ,  14045569 ,  14045570 ,  14045571 ,  14045572 ,  14045573 ,  14045574 ,  14045575 ,  14045576 ,  14045577 ,  14045578 ,  14045579 ,  14045580 ,  14045581 ,  14045582 ,  14045583 ,  14045584 ,  14045585 ,  14045586 ,  14045587 ,  14045588 ,  14045589 ,  14045590 ,  14045592 ,  14045593 ,  14045594 ,  14045595 ,  14045596 ,  14045597 ,  14045599 ,  14045600 ,  14045601 ,  14045602 ,  14045603 ,  14045604 ,  14045605 ,  14045606 ,  14045607 ,  14045608 ,  14045609 ,  14045611 ,  14045612 ,  14045613 ,  14045614 ,  14045615 ,  14045616 ,  14045617 ,  14045618 ,  14045619 ,  14045620 ,  14045621 ,  14045622 ,  14045623 ,  14045624 ,  14045625 ,  14045626 ,  14045627 ,  14045628 ,  14045629 ,  14045630 ,  14045631 ,  14045632 ,  14045633 ,  14045634 ,  14045635 ,  14045636 ,  14045637 ,  14045638 ,  14045639 ,  14045640 ,  14045641 ,  14045642 ,  14045643 ,  14045644 ,  14045645 ,  14045646 ,  14045647 ,  14045648 ,  14045649 ,  14045650 ,  14045651 ,  14045652 ,  14045653 ,  14045654 ,  14045655 ,  14045656 ,  14045657 ,  14045658 ,  14045659 ,  14045660 ,  14045661 ,  14045662 ,  14045663 ,  14045664 ,  14045665 ,  14045666 ,  14045667 ,  14045668 ,  14045669 ,  14045670 ,  14045671 ,  14045673 ,  14045674 ,  14045675 ,  14045676 ,  14045677 ,  14045678 ,  14045679 ,  14045680 ,  14045681 ,  14045682 ,  14045683 ,  14045684 ,  14045685 ,  14045686 ,  14045687 ,  14045688 ,  14045689 ,  14045690 ,  14045691 ,  14045692 ,  14045695 ,  14045696 ,  14045697 ,  14045698 ,  14045699 ,  14045700 ,  14045701 ,  14045702 ,  14045703 ,  14045704 ,  14045705 ,  14045706 ,  14045707 ,  14045708 ,  14045710 ,  14045711 ,  14045712 ,  14045713 ,  14045714 ,  14045717 ,  14045718 ,  14045719 ,  14045720 ,  14045721 ,  14045722 ,  14045723 ,  14045724 ,  14045726 ,  14045727 ,  14045728 ,  14045729 ,  14045730 ,  14045731 ,  14045732 ,  14045733 ,  14045734 ,  14045735 ,  14045736 ,  14045737 ,  14045738 ,  14045739 ,  14045740 ,  14045741 ,  14045742 ,  14045743 ,  14045744 ,  14045745 ,  14045746 ,  14045747 ,  14045749 ,  14045750 ,  14045751 ,  14045752 ,  14045753 ,  14045754 ,  14045755 ,  14045756 ,  14045757 ,  14045758 ,  14045759 ,  14045760 ,  14045761 ,  14045762 ,  14045763 ,  14045764 ,  14045765 ,  14045766 ,  14045767 ,  14045768 ,  14045769 ,  14045770 ,  14045771 ,  14045772 ,  14045773 ,  14045774 ,  14045775 ,  14045776 ,  14045777 ,  14045778 ,  14045779 ,  14045780 ,  14045781 ,  14045782 ,  14045783 ,  14045784 ,  14045785 ,  14045786 ,  14045787 ,  14045788 ,  14045789 ,  14045790 ,  14045791 ,  14045792 ,  14045793 ,  14045794 ,  14045795 ,  14045796 ,  14045797 ,  14045798 ,  14045799 ,  14045800 ,  14045801 ,  14045802 ,  14045803 ,  14045804 ,  14045805 ,  14045806 ,  14045807 ,  14045808 ,  14045809 ,  14045810 ,  14045811 ,  14045812 ,  14045813 ,  14045814 ,  14045815 ,  14045816 ,  14045817 ,  14045818 ,  14045819 ,  14045820 ,  14045821 ,  14045822 ,  14045823 ,  14045824 ,  14045825 ,  14045826 ,  14045827 ,  14045828 ,  14045829 ,  14045830 ,  14045831 ,  14045832 ,  14045833 ,  14045834 ,  14045835 ,  14045836 ,  14045837 ,  14045838 ,  14045839 ,  14045840 ,  14045841 ,  14045842 ,  14045843 ,  14045844 ,  14045845 ,  14045846 ,  14045847 ,  14045848 ,  14045849 ,  14045850 ,  14045851 ,  14045852 ,  14045853 ,  14045854 ,  14045855 ,  14045856 ,  14045857 ,  14045858 ,  14045859 ,  14045860 ,  14045861 ,  14045862 ,  14045863 ,  14045864 ,  14045865 ,  14045866 ,  14045867 ,  14045869 ,  14045870 ,  14045871 ,  14045872 ,  14045873 ,  14045874 ,  14045875 ,  14045876 ,  14045877 ,  14045878 ,  14045879 ,  14045880 ,  14045881 ,  14045882 ,  14045883 ,  14045884 ,  14045885 ,  14045886 ,  14045887 ,  14045888 ,  14045889 ,  14045890 ,  14045891 ,  14045892 ,  14045893 ,  14045894 ,  14045895 ,  14045896 ,  14045897 ,  14045898 ,  14045899 ,  14045900 ,  14045901 ,  14045902 ,  14045903 ,  14045904 ,  14045905 ,  14045906 ,  14045907 ,  14045908 ,  14045910 ,  14045911 ,  14045912 ,  14045913 ,  14045914 ,  14045915 ,  14045916 ,  14045918 ,  14045919 ,  14045920 ,  14045921 ,  14045922 ,  14045923 ,  14045924 ,  14045925 ,  14045926 ,  14045927 ,  14045928 ,  14045929 ,  14045936 ,  14045937 ,  14045938 ,  14045939 ,  14045940 ,  14045941 ,  14045942 ,  14045943 ,  14045945 ,  14045946 ,  14045947 ,  14045948 ,  14045949 ,  14045950 ,  14045951 ,  14045952 ,  14045953 ,  14045954 ,  14045955 ,  14045956 ,  14045957 ,  14045958 ,  14045959 ,  14045960 ,  14045961 ,  14045962 ,  14045963 ,  14045964 ,  14045965 ,  14045966 ,  14045967 ,  14045968 ,  14045969 ,  14045970 ,  14045971 ,  14045972 ,  14045973 ,  14045974 ,  14045975 ,  14045976 ,  14045977 ,  14045978 ,  14045979 ,  14045980 ,  14045981 ,  14045982 ,  14045983 ,  14045984 ,  14045985 ,  14045986 ,  14045987 ,  14045988 ,  14045991 ,  14045992 ,  14045994 ,  14045995 ,  14045996 ,  14045997 ,  14045998 ,  14045999 ,  14046000 ,  14046001 ,  14046002 ,  14046003 ,  14046004 ,  14046005 ,  14046006 ,  14046007 ,  14046008 ,  14046009 ,  14046010 ,  14046011 ,  14046012 ,  14046013 ,  14046014 ,  14046016 ,  14046017 ,  14046018 ,  14046019 ,  14046020 ,  14046021 ,  14046023 ,  14046024 ,  14046025 ,  14046026 ,  14046027 ,  14046028 ,  14046029 ,  14046030 ,  14046031 ,  14046032 ,  14046033 ,  14046034 ,  14046035 ,  14046036 ,  14046038 ,  14046039 ,  14046041 ,  14046042 ,  14046043 ,  14046044 ,  14046045 ,  14046046 ,  14046047 ,  14046048 ,  14046049 ,  14046050 ,  14046051 ,  14046052 ,  14046053 ,  14046054 ,  14046055 ,  14046056 ,  14046057 ,  14046058 ,  14046059 ,  14046060 ,  14046061 ,  14046062 ,  14046063 ,  14046064 ,  14046065 ,  14046066 ,  14046067 ,  14046068 ,  14046069 ,  14046070 ,  14046071 ,  14046072 ,  14046073 ,  14046074 ,  14046075 ,  14046076 ,  14046077 ,  14046078 ,  14046079 ,  14046080 ,  14046081 ,  14046082 ,  14046083 ,  14046084 ,  14046085 ,  14046086 ,  14046087 ,  14046088 ,  14046089 ,  14046090 ,  14046091 ,  14046092 ,  14046094 ,  14046095 ,  14046097 ,  14046098 ,  14046099 ,  14046100 ,  14046101 ,  14046102 ,  14046103 ,  14046104 ,  14046106 ,  14046107 ,  14046108 ,  14046109 ,  14046110 ,  14046111 ,  14046112 ,  14046113 ,  14046114 ,  14046116 ,  14046117 ,  14046119 ,  14046120 ,  14046121 ,  14046122 ,  14046123 ,  14046124 ,  14046125 ,  14046126 ,  14046127 ,  14046128 ,  14046129 ,  14046131 ,  14046132 ,  14046133 ,  14046135 ,  14046136 ,  14046137 ,  14046138 ,  14046140 ,  14046142 ,  14046143 ,  14046144 ,  14046145 ,  14046146 ,  14046147 ,  14046148 ,  14046149 ,  14046150 ,  14046151 ,  14046153 ,  14046154 ,  14046157 ,  14046158 ,  14046159 ,  14046165 ,  14046166 ,  14046168 ,  14046170 ,  14046171 ,  14046172 ,  14046173 ,  14046174 ,  14046175 ,  14046176 ,  14046177 ,  14046178 ,  14046180 ,  14046185 ,  14046186 ,  14046187 ,  14046188 ,  14046190 ,  14046207 ,  14046211 ,  14046216 ,  14046221 ,  14046226 ,  14046236 ,  14046238 ,  14046240 ,  14046241 ,  14046242 ,  14046246 ,  14046264 ,  14046267 ,  14046273 ,  14046274 ,  14046282 ,  14046283 ,  14046287 ,  14046288 ,  14046289 ,  14046294 ,  14046301 ,  14046303 ,  14046321 ,  14046325 ,  14046327 ,  14046329 ,  14046332 ,  14046336 ,  14046337 ,  14046342 ,  14046343 ,  14046344 ,  14046345 ,  14046346 ,  14046348 ,  14046349 ,  14046351 ,  14046353 ,  14046354 ,  14046356 ,  14046357 ,  14046358 ,  14046359 ,  14046360 ,  14046361 ,  14046362 ,  14046363 ,  14046364 ,  14046365 ,  14046366 ,  14046367 ,  14046368 ,  14046369 ,  14046370 ,  14046371 ,  14046373 ,  14046374 ,  14046375 ,  14046376 ,  14046377 ,  14046379 ,  14046380 ,  14046381 ,  14046383 ,  14046384 ,  14046385 ,  14046386 ,  14046387 ,  14046388 ,  14046389 ,  14046391 ,  14046393 ,  14046394 ,  14046395 ,  14046397 ,  14046399 ,  14046400 ,  14046401 ,  14046402 ,  14046403 ,  14046405 ,  14046406 ,  14046408 ,  14046409 ,  14046410 ,  14046412 ,  1 4046414 ,  14046415 ,  14046418 ,  14046419 ,  14046420 ,  14046424 ,  14046427 ,  14046428 ,  14046429 ,  14046430 ,  14046431 ,  14046432 ,  14046433 ,  14046434 ,  14046435 ,  14046436 ,  14046438 ,  14046439 ,  14046440 ,  14046442 ,  14046443 ,  14046444 ,  14046445 ,  14046446 ,  14046447 ,  14046448 ,  14046449 ,  14046450 ,  14046452 ,  14046454 ,  14046455 ,  14046456 ,  14046457 ,  14046458 ,  14046459 ,  14046460 ,  14046461 ,  14046462 ,  14046463 ,  14046464 ,  14046466 ,  14046467 ,  14046468 ,  14046469 ,  14046470 ,  14046471 ,  14046472 ,  14046473 ,  14046475 ,  14046476 ,  14046478 ,  14046482 ,  14046483 ,  14046485 ,  14046487 ,  14046488 ,  14046490 ,  14046492 ,  14046493 ,  14046494 ,  14046495 ,  14046496 ,  14046498 ,  14046503 ,  14046515 ,  14046516 ,  14046519 ,  14046520 ,  14046522 ,  14046523 ,  14046524 ,  14046529 ,  14046531 ,  14046533 ,  14046535 ,  14046536 ,  14046542 ,  14046543 ,  14046544 ,  14046545 ,  14046546 ,  14046547 ,  14046549 ,  14046550 ,  14046551 ,  14046552 ,  14046559 ,  14046571 ,  14046580 ,  14046584 ,  14046585 ,  14046586 ,  14046592 ,  14046593 ,  14046595 ,  14046596 ,  14046597 ,  14046598 ,  14046600 ,  14046602 ,  14046604 ,  14046607 ,  14046609 ,  14046610 ,  14046612 ,  14046614 ,  14046615 ,  14046620 ,  14046621 ,  14046626 ,  14046628 ,  14046629 ,  14046632 ,  14046633 ,  14046634 ,  14046640 ,  14046641 ,  14046642 ,  14046643 ,  14046644 ,  14046645 ,  14046647 ,  14046648 ,  14046649 ,  14046650 ,  14046651 ,  14046652 ,  14046654 ,  14046655 ,  14046657 ,  14046659 ,  14046660 ,  14046661 ,  14046662 ,  14046663 ,  14046664 ,  14046665 ,  14046666 ,  14046667 ,  14046668 ,  14046669 ,  14046670 ,  14046671 ,  14046672 ,  14046673 ,  14046675 ,  14046676 ,  14046677 ,  14046678 ,  14046679 ,  14046680 ,  14046681 ,  14046682 ,  14046683 ,  14046684 ,  14046685 ,  14046686 ,  14046687 ,  14046688 ,  14046690 ,  14046692 ,  14046693 ,  14046694 ,  14046695 ,  14046696 ,  14046697 ,  14046698 ,  14046699 ,  14046700 ,  14046702 ,  14046704 ,  14046705 ,  14046707 ,  14046708 ,  14046709 ,  14046710 ,  14046711 ,  14046712 ,  14046713 ,  14046714 ,  14046715 ,  14046716 ,  14046717 ,  14046718 ,  14046722 ,  14046723 ,  14046724 ,  14046725 ,  14046726 ,  14046727 ,  14046728 ,  14046729 ,  14046730 ,  14046731 ,  14046732 ,  14046733 ,  14046734 ,  14046735 ,  14046736 ,  14046737 ,  14046738 ,  14046739 ,  14046740 ,  14046741 ,  14046742 ,  14046743 ,  14046744 ,  14046745 ,  14046746 ,  14046747 ,  14046748 ,  14046749 ,  14046750 ,  14046751 ,  14046752 ,  14046753 ,  14046754 ,  14046755 ,  14046756 ,  14046757 ,  14046758 ,  14046759 ,  14046760 ,  14046761 ,  14046762 ,  14046763 ,  14046764 ,  14046765 ,  14046766 ,  14046767 ,  14046768 ,  14046769 ,  14046770 ,  14046771 ,  14046772 ,  14046773 ,  14046774 ,  14046775 ,  14046776 ,  14046777 ,  14046778 ,  14046779 ,  14046780 ,  14046781 ,  14046782 ,  14046783 ,  14046784 ,  14046785 ,  14046786 ,  14046787 ,  14046788 ,  14046789 ,  14046790 ,  14046792 ,  14046793 ,  14046794 ,  14046795 ,  14046796 ,  14046797 ,  14046798 ,  14046800 ,  14046801 ,  14046802 ,  14046803 ,  14046804 ,  14046805 ,  14046806 ,  14046807 ,  14046808 ,  14046809 ,  14046810 ,  14046811 ,  14046812 ,  14046813 ,  14046814 ,  14046815 ,  14046816 ,  14046817 ,  14046818 ,  14046819 ,  14046820 ,  14046821 ,  14046822 ,  14046823 ,  14046824 ,  14046825 ,  14046826 ,  14046827 ,  14046828 ,  14046829 ,  14046830 ,  14046831 ,  14046832 ,  14046834 ,  14046835 ,  14046836 ,  14046837 ,  14046838 ,  14046839 ,  14046840 ,  14046841 ,  14046842 ,  14046843 ,  14046844 ,  14046845 ,  14046846 ,  14046847 ,  14046848 ,  14046849 ,  14046850 ,  14046851 ,  14046852 ,  14046853 ,  14046855 ,  14046856 ,  14046858 ,  14046859 ,  14046860 ,  14046861 ,  14046863 ,  14046865 ,  14046866 ,  14046867 ,  14046869 ,  14046870 ,  14046871 ,  14046872 ,  14046873 ,  14046875 ,  14046876 ,  14046878 ,  14046879 ,  14046880 ,  14046881 ,  14046882 ,  14046883 ,  14046884 ,  14046885 ,  14046886 ,  14046887 ,  14046888 ,  14046889 ,  14046890 ,  14046891 ,  14046892 ,  14046893 ,  14046894 ,  14046895 ,  14046896 ,  14046897 ,  14046899 ,  14046900 ,  14046901 ,  14046902 ,  14046903 ,  14046904 ,  14046905 ,  14046906 ,  14046907 ,  14046908 ,  14046909 ,  14046910 ,  14046911 ,  14046912 ,  14046913 ,  14046914 ,  14046915 ,  14046916 ,  14046917 ,  14046918 ,  14046919 ,  14046920 ,  14046921 ,  14046922 ,  14046923 ,  14046924 ,  14046925 ,  14046926 ,  14046927 ,  14046928 ,  14046929 ,  14046930 ,  14046931 ,  14046932 ,  14046934 ,  14046935 ,  14046936 ,  14046937 ,  14046938 ,  14046939 ,  14046943 ,  14046945 ,  14046946 ,  14046947 ,  14046948 ,  14046949 ,  14046950 ,  14046951 ,  14046952 ,  14046954 ,  14046955 ,  14046956 ,  14046958 ,  14046959 ,  14046960 ,  14046961 ,  14046962 ,  14046963 ,  14046964 ,  14046965 ,  14046966 ,  14046967 ,  14046968 ,  14046969 ,  14046970 ,  14046971 ,  14046972 ,  14046973 ,  14046974 ,  14046975 ,  14046977 ,  14046978 ,  14046979 ,  14046980 ,  14046981 ,  14046982 ,  14046983 ,  14046984 ,  14046985 ,  14046986 ,  14046987 ,  14046988 ,  14046989 ,  14046990 ,  14046991 ,  14046992 ,  14046993 ,  14046994 ,  14046995 ,  14046996 ,  14046997 ,  14046998 ,  14046999 ,  14047000 ,  14047001 ,  14047002 ,  14047003 ,  14047004 ,  14047005 ,  14047006 ,  14047007 ,  14047008 ,  14047009 ,  14047010 ,  14047011 ,  14047012 ,  14047013 ,  14047014 ,  14047015 ,  14047016 ,  14047018 ,  14047019 ,  14047020 ,  14047021 ,  14047022 ,  14047023 ,  14047024 ,  14047026 ,  14047027 ,  14047028 ,  14047029 ,  14047030 ,  14047031 ,  14047032 ,  14047034 ,  14047035 ,  14047036 ,  14047038 ,  14047039 ,  14047040 ,  14047041 ,  14047042 ,  14047043 ,  14047045 ,  14047046 ,  14047047 ,  14047048 ,  14047049 ,  14047050 ,  14047051 ,  14047052 ,  14047053 ,  14047055 ,  14047056 ,  14047057 ,  14047058 ,  14047060 ,  14047061 ,  14047062 ,  14047063 ,  14047064 ,  14047065 ,  14047066 ,  14047067 ,  14047068 ,  14047069 ,  14047070 ,  14047071 ,  14047072 ,  14047073 ,  14047074 ,  14047075 ,  14047076 ,  14047077 ,  14047078 ,  14047079 ,  14047080 ,  14047081 ,  14047082 ,  14047083 ,  14047084 ,  14047085 ,  14047086 ,  14047087 ,  14047088 ,  14047089 ,  14047090 ,  14047091 ,  14047092 ,  14047093 ,  14047094 ,  14047095 ,  14047096 ,  14047097 ,  14047098 ,  14047099 ,  14047100 ,  14047101 ,  14047102 ,  14047103 ,  14047104 ,  14047105 ,  14047106 ,  14047107 ,  14047108 ,  14047109 ,  14047110 ,  14047111 ,  14047112 ,  14047113 ,  14047114 ,  14047115 ,  14047116 ,  14047117 ,  14047118 ,  14047119 ,  14047120 ,  14047121 ,  14047122 ,  14047123 ,  14047124 ,  14047125 ,  14047126 ,  14047127 ,  14047128 ,  14047129 ,  14047130 ,  14047131 ,  14047133 ,  14047134 ,  14047135 ,  14047136 ,  14047137 ,  14047138 ,  14047140 ,  14047143 ,  14047145 ,  14047147 ,  14047148 ,  14047150 ,  14047151 ,  14047152 ,  14047153 ,  14047154 ,  14047155 ,  14047156 ,  14047158 ,  14047159 ,  14047160 ,  14047163 ,  14047164 ,  14047165 ,  14047166 ,  14047167 ,  14047168 ,  14047169 ,  14047170 ,  14047171 ,  14047172 ,  14047173 ,  14047174 ,  14047175 ,  14047176 ,  14047178 ,  14047179 ,  14047180 ,  14047181 ,  14047182 ,  14047184 ,  14047185 ,  14047186 ,  14047187 ,  14047188 ,  14047189 ,  14047190 ,  14047191 ,  14047192 ,  14047193 ,  14047194 ,  14047195 ,  14047196 ,  14047197 ,  14047198 ,  14047199 ,  14047200 ,  14047201 ,  14047202 ,  14047203 ,  14047204 ,  14047205 ,  14047207 ,  14047208 ,  14047209 ,  14047210 ,  14047211 ,  14047212 ,  14047213 ,  14047214 ,  14047215 ,  14047216 ,  14047217 ,  14047219 ,  14047220 ,  14047221 ,  14047222 ,  14047222 ,  14047223 ,  14047224 ,  14047225 ,  14047226 ,  14047227 ,  14047228 ,  14047229 ,  14047230 ,  14047231 ,  14047233 ,  14047234 ,  14047238 ,  14047239 ,  14047240 ,  14047241 ,  14047242 ,  14047243 ,  14047244 ,  14047245 ,  14047247 ,  14047248 ,  14047249 ,  14047250 ,  14047251 ,  14047252 ,  14047253 ,  14047254 ,  14047255 ,  14047256 ,  14047257 ,  14047258 ,  14047259 ,  14047260 ,  14047263 ,  14047264 ,  14047266 ,  14047267 ,  14047320 ,  14047321 ,  14047322 ,  14047323 ,  14047324 ,  14047325 ,  14047326 ,  14047327 ,  14047328 ,  14047331 ,  14047332 ,  14047333 ,  14047334 ,  14047335 ,  14047336 ,  14047337 ,  14047338 ,  14047339 ,  14047340 ,  14047341 ,  14047342 ,  14047343 ,  14047344 ,  14047345 ,  14047346 ,  14047347 ,  14047348 ,  14047349 ,  14047350 ,  14047351 ,  14047352 ,  14047353 ,  14047355 ,  14047356 ,  14047358 ,  14047360 ,  14047361 ,  14047362 ,  14047363 ,  14047364 ,  14047365 ,  14047366 ,  14047367 ,  14047369 ,  14047370 ,  14047371 ,  14047372 ,  14047373 ,  14047374 ,  14047375 ,  14047380 ,  14047383 ,  14047385 ,  14047386 ,  14047387 ,  14047388 ,  14047389 ,  14047391 ,  14047392 ,  14047393 ,  14047394 ,  14047395 ,  14047396 ,  14047397 ,  14047399 ,  14047400 ,  14047401 ,  14047402 ,  14047403 ,  14047404 ,  14047405 ,  14047407 ,  14047408 ,  14047409 ,  14047410 ,  14047411 ,  14047413 ,  14047414 ,  14047415 ,  14047416 ,  14047417 ,  14047418 ,  14047420 ,  14047421 ,  14047422 ,  14047423 ,  14047424 ,  14047425 ,  14047426 ,  14047427 ,  14047428 ,  14047429 ,  14047430 ,  14047431 ,  14047432 ,  14047433 ,  14047434 ,  14047435 ,  14047436 ,  14047437 ,  14047438 ,  14047439 ,  14047440 ,  14047441 ,  14047442 ,  14047443 ,  14047444 ,  14047445 ,  14047446 ,  14047447 ,  14047448 ,  14047449 ,  14047450 ,  14047451 ,  14047452 ,  14047453 ,  14047454 ,  14047455 ,  14047456 ,  14047457 ,  14047458 ,  14047459 ,  14047460 ,  14047461 ,  14047463 ,  14047464 ,  14047465 ,  14047466 ,  14047467 ,  14047468 ,  14047469 ,  14047470 ,  14047471 ,  14047472 ,  14047473 ,  14047474 ,  14047475 ,  14047476 ,  14047477 ,  14047479 ,  14047480 ,  14047481 ,  14047482 ,  14047483 ,  14047484 ,  14047485 ,  14047486 ,  14047487 ,  14047488 ,  14047489 ,  14047490 ,  14047491 ,  14047492 ,  14047493 ,  14047494 ,  14047495 ,  14047496 ,  14047497 ,  14047498 ,  14047499 ,  14047500 ,  14047501 ,  14047502 ,  14047503 ,  14047504 ,  14047505 ,  14047506 ,  14047507 ,  14047508 ,  14047509 ,  14047510 ,  14047511 ,  14047512 ,  14047514 ,  14047515 ,  14047516 ,  14047517 ,  14047518 ,  14047519 ,  14047520 ,  14047521 ,  14047522 ,  14047523 ,  14047524 ,  14047525 ,  14047526 ,  14047527 ,  14047528 ,  14047529 ,  14047530 ,  14047531 ,  14047532 ,  14047533 ,  14047534 ,  14047535 ,  14047536 ,  14047537 ,  14047538 ,  14047539 ,  14047540 ,  14047541 ,  14047542 ,  14047543 ,  14047544 ,  14047545 ,  14047546 ,  14047547 ,  14047548 ,  14047549 ,  14047550 ,  14047551 ,  14047552 ,  14047553 ,  14047554 ,  14047555 ,  14047556 ,  14047557 ,  14047558 ,  14047559 ,  14047560 ,  14047561 ,  14047562 ,  14047564 ,  14047565 ,  14047566 ,  14047568 ,  14047569 ,  14047570 ,  14047571 ,  14047572 ,  14047573 ,  14047574 ,  14047575 ,  14047576 ,  14047577 ,  14047578 ,  14047579 ,  14047580 ,  14047581 ,  14047582 ,  14047584 ,  14047585 ,  14047586 ,  14047587 ,  14047588 ,  14047589 ,  14047590 ,  14047591 ,  14047592 ,  14047671 ,  14047672 ,  14047673 ,  14047674 ,  14047675 ,  14047676 ,  14047677 ,  14047678 ,  14047679 ,  14047680 ,  14047681 ,  14047682 ,  14047683 ,  14047684 ,  14047685 ,  14047686 ,  14047687 ,  14047688 ,  14047689 ,  14047690 ,  14047691 ,  14047692 ,  14047693 ,  14047694 ,  14047695 ,  14047696 ,  14047697 ,  14047698 ,  14047699 ,  14047700 ,  14047701 ,  14047702 ,  14047704 ,  14047705 ,  14047706 ,  14047707 ,  14047708 ,  14047709 ,  14047710 ,  14047711 ,  14047712 ,  14047713 ,  14047714 ,  14047715 ,  14047716 ,  14047717 ,  14047718 ,  14047719 ,  14047720 ,  14047721 ,  14047723 ,  14047724 ,  14047725 ,  14047727 ,  14047728 ,  14047729 ,  14047730 ,  14047731 ,  14047732 ,  14047733 ,  14047734 ,  14047735 ,  14047736 ,  14047737 ,  14047738 ,  14047739 ,  14047740 ,  14047742 ,  14047743 ,  14047744 ,  14047745 ,  14047746 ,  14047747 ,  14047748 ,  14047749 ,  14047750 ,  14047751 ,  14047752 ,  14047754 ,  14047755 ,  14047756 ,  14047757 ,  14047759 ,  14047760 ,  14047761 ,  14047762 ,  14047763 ,  14047764 ,  14047765 ,  14047766 ,  14047767 ,  14047768 ,  14047769 ,  14047770 ,  14047771 ,  14047772 ,  14047773 ,  14047774 ,  14047775 ,  14047776 ,  14047777 ,  14047778 ,  14047779 ,  14047780 ,  14047781 ,  14047782 ,  14047783 ,  14047784 ,  14047785 ,  14047786 ,  14047787 ,  14047789 ,  14047790 ,  14047791 ,  14047792 ,  14047793 ,  14047794 ,  14047795 ,  14047796 ,  14047797 ,  14047798 ,  14047799 ,  14047800 ,  14047801 ,  14047803 ,  14047804 ,  14047805 ,  14047806 ,  14047807 ,  14047809 ,  14047810 ,  14047812 ,  14047813 ,  14047814 ,  14047815 ,  14047816 ,  14047818 ,  14047820 ,  14047822 ,  14047823 ,  14047825 ,  14047826 ,  14047827 ,  14047829 ,  14047831 ,  14047832 ,  14047833 ,  14047834 ,  14047837 ,  14047838 ,  14047840 ,  14047841 ,  14047842 ,  14047843 ,  14047845 ,  14047846 ,  14047848 ,  14047850 ,  14047851 ,  14047852 ,  14047854 ,  14047855 ,  14047858 ,  14047859 ,  14047860 ,  14047862 ,  14047863 ,  14047864 ,  14047868 ,  14047871 ,  14047881 ,  14047882 ,  14047883 ,  14047885 ,  14047886 ,  14047888 ,  14047891 ,  14047893 ,  14047894 ,  14047895 ,  14047905 ,  14047907 ,  14047908 ,  14047910 ,  14047911 ,  14047913 ,  14047916 ,  14047918 ,  14047922 ,  14047924 ,  14047925 ,  14047926 ,  14047927 ,  14047928 ,  14047929 ,  14047931 ,  14047932 ,  14047933 ,  14047934 ,  14047935 ,  14047938 ,  14047942 ,  14047943 ,  14047944 ,  14047945 ,  14047947 ,  14047948 ,  14047949 ,  14047950 ,  14047952 ,  14047953 ,  14047954 ,  14047958 ,  14047959 ,  14047960 ,  14047961 ,  14047963 ,  14047964 ,  14047968 ,  14047969 ,  14047971 ,  14047972 ,  14047973 ,  14047977 ,  14047978 ,  14047979 ,  14047980 ,  14047982 ,  14047983 ,  14047984 ,  14047985 ,  14047986 ,  14047987 ,  14047988 ,  14047989 ,  14047991 ,  14047992 ,  14047993 ,  14047994 ,  14047995 ,  14047996 ,  16412201 ,  16412202 ,  16412203 ,  16412204 ,  16412205 ,  16412206 ,  16412207 ,  16412208 ,  16412209 ,  16412210 ,  16412211 ,  16412212 ,  16412213 ,  16412214 ,  16412215 ,  16412216 ,  16412217 ,  16412218 ,  16412219 ,  16412220 ,  16412222 ,  16412223 ,  16412224 ,  16412225 ,  16412226 ,  16412227 ,  16412228 ,  16412229 ,  16412230 ,  16412231 ,  16412232 ,  16412233 ,  16412234 ,  16412235 ,  16412236 ,  16412237 ,  16412238 ,  16412239 ,  16412240 ,  16412241 ,  16412242 ,  16412243 ,  16412244 ,  16412245 ,  16412246 ,  16412247 ,  16412248 ,  16412249 ,  16412250 ,  16412251 ,  16412252 ,  16412253 ,  16412254 ,  16412255 ,  16412256 ,  16412257 ,  16412258 ,  16412259 ,  16412260 ,  16412261 ,  16412262 ,  16412263 ,  16412264 ,  16412265 ,  16412266 ,  16412267 ,  16412268 ,  16412269 ,  16412270 ,  16412271 ,  16412272 ,  16412273 ,  16412274 ,  16412275 ,  16412276 ,  16412277 ,  16412278 ,  16412279 ,  16412280 ,  16412281 ,  16412282 ,  16412283 ,  16412285 ,  16412286 ,  16412287 ,  16412288 ,  16412289 ,  16412290 ,  16412291 ,  16412292 ,  16412293 ,  16412294 ,  16412295 ,  16412297 ,  16412298 ,  16412299 ,  16412300 ,  16412301 ,  16412302 ,  16412303 ,  16412304 ,  16412305 ,  16412306 ,  16412307 ,  16412308 ,  16412309 ,  16412310 ,  16412311 ,  16412312 ,  16412313 ,  16412314 ,  16412315 ,  16412316 ,  16412317 ,  16412318 ,  16412319 ,  16412320 ,  16412321 ,  16412322 ,  16412323 ,  16412324 ,  16412325 ,  16412326 ,  16412327 ,  16412328 ,  16412329 ,  16412330 ,  16412331 ,  16412332 ,  16412333 ,  16412334 ,  16412335 ,  16412336 ,  16412337 ,  16412338 ,  16412339 ,  16412340 ,  16412341 ,  16412342 ,  16412343 ,  16412344 ,  16412345 ,  16412346 ,  16412347 ,  16412349 ,  16412350 ,  16412351 ,  16412352 ,  16412353 ,  16412354 ,  16412355 ,  16412356 ,  16412357 ,  16412358 ,  16412359 ,  16412360 ,  16412361 ,  16412362 ,  16412363 ,  16412364 ,  16412365 ,  16412366 ,  16412367 ,  16412368 ,  16412369 ,  16412370 ,  16412371 ,  16412372 ,  16412373 ,  16412374 ,  16412375 ,  16412376 ,  16412377 ,  16412378 ,  16412379 ,  16412380 ,  16412381 ,  16412382 ,  16412383 ,  16412384 ,  16412385 ,  16412386 ,  16412387 ,  16422201 ,  16422202 ,  16422203 ,  16422204 ,  16422205 ,  16422207 ,  16422208 ,  16422209 ,  16422210 ,  16422211 ,  16422212 ,  16422213 ,  16422214 ,  16422215 ,  16422216 ,  16422217 ,  16422218 ,  16422219 ,  16422220 ,  16422221 ,  16422222 ,  16422223 ,  16422224 ,  16422225 ,  16422226 ,  16422227 ,  16422228 ,  16422229 ,  16422230 ,  16422231 ,  16422232 ,  16422233 ,  16422234 ,  16422235 ,  16422236 ,  16422237 ,  16422238 ,  16422239 ,  16422240 ,  16422241 ,  16422242 ,  16422243 ,  16422244 ,  16422245 ,  16422246 ,  16422247 ,  16422248 ,  16422249 ,  16422250 ,  16422251 ,  16422252 ,  16422253 ,  16422254 ,  16422255 ,  16422256 ,  16422257 ,  16422258 ,  16422259 ,  16422260 ,  16422261 ,  16422262 ,  16422263 ,  16422264 ,  16422265 ,  16422266 ,  16422267 ,  16422268 ,  16422269 ,  16422270 ,  16422271 ,  16422272 ,  16422273 ,  16422274 ,  16422275 ,  16422276 ,  16422277 ,  16422278 ,  16422279 ,  16422280 ,  16422281 ,  16422282 ,  16422283 ,  16422284 ,  16422285 ,  16422286 ,  16422287 ,  16422288 ,  16422289 ,  16422290 ,  16422291 ,  16422292 ,  16422293 ,  16422294 ,  16422295 ,  16422296 ,  16422297 ,  16422298 ,  16422299 ,  16422300 ,  16422301 ,  16422302 ,  16422303 ,  16422304 ,  16422305 ,  16422306 ,  16422307 ,  16422308 ,  16422309 ,  16422310 ,  16422311 ,  16422312 ,  16422313 ,  16422314 ,  16422315 ,  16422316 ,  16422317 ,  16422318 ,  16422319 ,  16422321 ,  16422324 ,  16422327 ,  16422331 ,  16422334 ,  16422337 ,  16422341 ,  16422344 ,  16422347 ,  16422351 ,  16422354 ,  16422357 ,  16422361 ,  16422364 ,  16422367 ,  16422371 ,  16422374 ,  16422377 ,  16422381 ,  16422384 ,  16422387 ,  16422391 ,  16422394 ,  16422397 ,  16422401 ,  16422404 ,  16422407 ,  16422411 ,  16422414 ,  16422417 ,  16422421 ,  16422424 ,  16422427 ,  16422431 ,  16422434 ,  16422437 ,  16422441 ,  16422444 ,  16422447 ,  16422451 ,  16422454 ,  16422457 ,  16422461 ,  16422464 ,  16422467 ,  16422471 ,  16422474 ,  16422477 ,  16422481 ,  16422484 ,  16422487 ,  16422491 ,  16422494 ,  16422497 ,  16422501 ,  16422504 ,  16422507 ,  16422511 ,  16422514 ,  16422517 ,  16422521 ,  16422524 ,  16422527 ,  16422531 ,  16422534 ,  16422537 ,  16422541 ,  16422544 ,  16422547 ,  16422551 ,  16422554 ,  16422557 ,  16422561 ,  16422564 ,  16422567 ,  16422571 ,  16422574 ,  16422577 ,  16422581 ,  16422584 ,  16422587 ,  16422591 ,  16422594 ,  16422597 ,  16422601 ,  16422604 ,  16422607 ,  16422611 ,  16422614 ,  16422617 ,  16422621 ,  16422624 ,  16422627 ,  16422631 ,  16422634 ,  16422637 ,  16422641 ,  16422644 ,  16422647 ,  16422651 ,  16422654 ,  16422657 ,  16422661 ,  16422664 ,  16422667 ,  16422671 ,  16422674 ,  16422677 ,  16422681 ,  16422684 ,  16422687 ,  16422691 ,  16422694 ,  16422697 ,  16422701 ,  16794201 ,  16794202 ,  16794204 ,  16794205 ,  16794206 ,  16794207 ,  16794209 ,  16794210 ,  16794211 ,  16794212 ,  16794213 ,  16794214 ,  16794216 ,  16794217 ,  16794218 ,  16794220 ,  16794221 ,  16794222 ,  16794224 ,  16794225 ,  16794226 ,  16794227 ,  16794228 ,  16794230 ,  16794231 ,  16794232 ,  16794233 ,  16794234 ,  16794235 ,  16794236 ,  16794237 ,  16794238 ,  16794239 ,  16794240 ,  16794241 ,  16794242 ,  16794244 ,  16794246 ,  16794248 ,  16794249 ,  16794250 ,  16794252 ,  16794253 ,  16794254 ,  16794255 ,  16794256 ,  16794257 ,  16794259 ,  16794260 ,  16794262 ,  16794263 ,  16794264 ,  16794265 ,  16794266 ,  16794267 ,  16794269 ,  16794270 ,  16794271 ,  16794275 ,  16794276 ,  16794277 ,  16794278 ,  16794279 ,  16794280 ,  16794282 ,  16794284 ,  16794285 ,  16794286 ,  16794287 ,  16794289 ,  16794291 ,  16794292 ,  16794293 ,  16794294 ,  16794295 ,  16794296 ,  16794297 ,  16794299 ,  16794300 ,  16794302 ,  16794303 ,  16794304 ,  16794305 ,  16794306 ,  16794307 ,  16794309 ,  16794311 ,  16794312 ,  16794313 ,  16794316 ,  16794317 ,  16794318 ,  16794319 ,  16794320 ,  16794321 ,  16794322 ,  16794323 ,  16794324 ,  16794325 ,  16794326 ,  16794327 ,  16794328 ,  16794330 ,  16794331 ,  16794332 ,  16794333 ,  16794336 ,  16794337 ,  16794338 ,  16794339 ,  16794342 ,  16794344 ,  16794345 ,  16794346 ,  16794348 ,  16794349 ,  16794350 ,  16794351 ,  16794355 ,  16794356 ,  16794357 ,  16794358 ,  16794360 ,  16794361 ,  16794362 ,  16794363 ,  16794365 ,  16794366 ,  16794367 ,  16794368 ,  16794369 ,  16794370 ,  16794371 ,  16794372 ,  16794373 ,  16794374 ,  16794375 ,  16794377 ,  16794378 ,  16794379 ,  16794380 ,  16794381 ,  16794383 ,  16794385 ,  16794386 ,  16794388 ,  16794389 ,  16794390 ,  16794391 ,  16794392 ,  16794393 ,  16794394 ,  16794395 ,  16794396 ,  16794397 ,  16794398 ,  16794399 ,  16794400 ,  16794401 ,  16794402 ,  16794403 ,  16794404 ,  16794405 ,  16794407 ,  16794408 ,  16794409 ,  16794410 ,  16794411 ,  16794413 ,  16794414 ,  16794415 ,  16794416 ,  16794418 ,  16794419 ,  16794421 ,  16794423 ,  16794424 ,  16794425 ,  16794426 ,  16794427 ,  16794428 ,  16794429 ,  16794430 ,  16794433 ,  16794434 ,  16794436 ,  16794437 ,  16794440 ,  16794441 ,  16794442 ,  16794443 ,  16794444 ,  16794445 ,  16794447 ,  16794448 ,  16794449 ,  16794450 ,  16794451 ,  16794452 ,  16794454 ,  16794455 ,  16794456 ,  16794457 ,  16794458 ,  16794459 ,  16794460 ,  16794461 ,  16794462 ,  16794463 ,  16794464 ,  16794465 ,  16794466 ,  16794467 ,  16794468 ,  16794469 ,  16794470 ,  16794471 ,  16794473 ,  16794474 ,  16794475 ,  16794477 ,  16794478 ,  16794479 ,  16794480 ,  16794482 ,  16794483 ,  16794484 ,  16794485 ,  16794486 ,  16794487 ,  16794488 ,  16794489 ,  16794490 ,  16794491 ,  16794492 ,  16794493 ,  16794495 ,  16794497 ,  16794498 ,  16794499 ,  16794500 ,  16794501 ,  16794502 ,  16794503 ,  16794504 ,  16794505 ,  16794506 ,  16794507 ,  16794509 ,  16794510 ,  16794511 ,  16794512 ,  16794513 ,  16794514 ,  16794515 ,  16794521 ,  16794523 ,  16794525 ,  16794529 ,  16794530 ,  16794532 ,  16794536 ,  16794540 ,  16794541 ,  16794542 ,  16794543 ,  16794544 ,  16794545 ,  16794546 ,  16794552 ,  16794556 ,  16794561 ,  16794566 ,  16794567 ,  16794570 ,  16794573 ,  16794576 ,  16794577 ,  16794586 ,  16794587 ,  16794589 ,  16794590 ,  16794592 ,  16794594 ,  16794596 ,  16794597 ,  16794598 ,  16794604 ,  16794605 ,  16794607 ,  16794608 ,  16794613 ,  16794616 ,  16794628 ,  16794632 ,  16794635 ,  16794637 ,  16794638 ,  16794643 ,  16794644 ,  16794648 ,  16794653 ,  16794656 ,  16794657 ,  16794665 ,  16794668 ,  16794675 ,  16794676 ,  16794679 ,  16794683 ,  16794684 ,  16794688 ,  16794702 ,  16794703 ,  16794704 ,  16794705 ,  16794708 ,  16795201 ,  16795204 ,  16795207 ,  16795211 ,  16795214 ,  16795217 ,  16795221 ,  16795224 ,  16795227 ,  16795231 ,  16795234 ,  16795237 ,  16795241 ,  16795244 ,  16795247 ,  16795251 ,  16795254 ,  16795257 ,  16795261 ,  16795264 ,  16795267 ,  16795271 ,  16795274 ,  16795277 ,  16795281 ,  16795297 ,  16795301 ,  16795304 ,  16795307 ,  16795311 ,  16795314 ,  16795317 ,  16795324 ,  16795327 ,  16795331 ,  16795334 ,  16795341 ,  16795344 ,  16795351 ,  16795354 ,  16795357 ,  16795361 ,  16795364 ,  16795367 ,  16795371 ,  16795377 ,  16795381 ,  16795384 ,  16795387 ,  16795391 ,  16795394 ,  16795397 ,  16795401 ,  16795407 ,  16795411 ,  16795414 ,  16795417 ,  16795421 ,  16795424 ,  16795427 ,  16795434 ,  16795441 ,  16795444 ,  16795451 ,  16795454 ,  16795457 ,  16795461 ,  16795464 ,  16795467 ,  16795471 ,  16795477 ,  16795481 ,  16795484 ,  16795487 ,  16795491 ,  16795501 ,  16795507 ,  16795511 ,  16795521 ,  16838202 ,  16838203 ,  16838204 ,  16838205 ,  16838207 ,  16838208 ,  16838209 ,  16838210 ,  16838211 ,  16838212 ,  16838213 ,  16838214 ,  16838216 ,  16838217 ,  16838218 ,  16838219 ,  16838220 ,  16838221 ,  16838222 ,  16838223 ,  16838224 ,  16838225 ,  16838226 ,  16838227 ,  16838228 ,  16838229 ,  16838230 ,  16838231 ,  16838232 ,  16838233 ,  16838234 ,  16838235 ,  16838236 ,  18627408 ,  18627409 ,  18627410 ,  18627414 ,  18627415 ,  18627417 ,  18627418 ,  18627419 ,  18627420 ,  18627421 ,  18627422 ,  18627423 ,  18627424 ,  18627425 ,  18627426 ,  18627428 ,  18627429 ,  18627430 ,  18627431 ,  18627432 ,  18627435 ,  18627436 ,  18627439 ,  18627440 ,  18627442 ,  18627445 ,  18627447 ,  18627448 ,  18627449 ,  18627450 ,  18627451 ,  18627452 ,  18627453 ,  18627454 ,  18627455 ,  18627456 ,  18627459 ,  18627460 ,  18627461 ,  18627462 ,  18627463 ,  18627464 ,  18627465 ,  18627466 ,  18627467 ,  18627468 ,  18627469 ,  18627470 ,  18627471 ,  18627472 ,  18627474 ,  18627475 ,  18627476 ,  18627477 ,  18627478 ,  18627479 ,  18627481 ,  18627482 ,  18627483 ,  18627484 ,  18627485 ,  18627487 ,  18627488 ,  18627489 ,  18627490 ,  18627491 ,  18627492 ,  18627493 ,  18627494 ,  18627495 ,  18627496 ,  18627497 ,  18627498 ,  18627499 ,  18627500 ,  18627501 ,  18627502 ,  18627505 ,  18627507 ,  18627508 ,  18627509 ,  18627510 ,  18627511 ,  18627512 ,  18627517 ,  18627522 ,  18627525 ,  18627526 ,  18627530 ,  18627532 ,  18627533 ,  18627537 ,  18627542 ,  18627543 ,  18627544 ,  18627547 ,  18627554 ,  18627556 ,  18627558 ,  18627559 ,  18627563 ,  18627566 ,  18627572 ,  18627573 ,  18627581 ,  18627586 ,  18627589 ,  18627599 ,  18627607 ,  18627610 ,  18627612 ,  18627622 ,  18627629 ,  18627643 ,  18627644 ,  18627653 ,  18627661 ,  18627670 ,  18627687 ,  18627694 ,  18627695 ,  18627703 ,  18627704 ,  18627706 ,  18627708 ,  18627709 ,  18627710 ,  18627711 ,  18627713 ,  18627714 ,  18627715 ,  18627717 ,  18627718 ,  18627719 ,  18627720 ,  18627721 ,  18627723 ,  18627725 ,  18627728 ,  18627730 ,  18627731 ,  18627732 ,  18627733 ,  18627736 ,  18627745 ,  18627748 ,  18627749 ,  18627752 ,  18627754 ,  18627757 ,  18627759 ,  18627761 ,  18627762 ,  18627763 ,  18627765 ,  18627766 ,  18627767 ,  18627776 ,  18627777 ,  18627782 ,  18627783 ,  18627784 ,  18627787 ,  18627791 ,  18627801 ,  18627806 ,  18627810 ,  18627818 ,  18627819 ,  18627821 ,  18627822 ,  18627823 ,  18627824 ,  18627825 ,  18627826 ,  18627829 ,  18627830 ,  18627831 ,  18627832 ,  18627834 ,  18627836 ,  18627837 ,  18627840 ,  18627865 ,  18627867 ,  18627870 ,  18627871 ,  18627872 ,  18627874 ,  18627876 ,  18627877 ,  18627878 ,  18627880 ,  18627882 ,  18627884 ,  18627886 ,  18627891 ,  18627892 ,  18627897 ,  18627899 ,  18627901 ,  18627904 ,  18627905 ,  18627906 ,  18627908 ,  18627909 ,  18627910 ,  18627911 ,  18627912 ,  18627915 ,  18627916 ,  18627917 ,  18627918 ,  18627919 ,  18627922 ,  18627926 ,  18627927 ,  18627928 ,  18627930 ,  18627933 ,  18627935 ,  18627938 ,  18627939 ,  18627940 ,  18627941 ,  18627942 ,  18627943 ,  18627944 ,  18627945 ,  18627946 ,  18627947 ,  18627949 ,  18627950 ,  18627951 ,  18627952 ,  18627953 ,  18627954 ,  18627956 ,  18627957 ,  18627959 ,  18627960 ,  18627961 ,  18627962",
      "recall_initiation_date": "20130201",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}